Recalibration of the NBS Glass Standards of Spectral Transmittance1

H. J. Keegan, J. C. Schleter, M. Belknap
{"title":"Recalibration of the NBS Glass Standards of Spectral Transmittance1","authors":"H. J. Keegan, J. C. Schleter, M. Belknap","doi":"10.6028/jres.067A.056","DOIUrl":null,"url":null,"abstract":"In 1934, Gibson, Walker, and Brown developed sets of four colored glass filters to serve as working standards of spectral transmittance for checking the reliability of spectrophotometers. Several sets of these glasses were measured carefully and reserved and designated as future reference standards. Duplicate standards evaluated by comparison with the reference standards are available by purchase to the public. The current set of reference standards was established in the years 1945 to 1947, and one of these reference standards (selenium-red) was recalibrated in 1952. This paper reports a recalibration, made in 1961 and 1962, of all four glasses (selenium-red, carbon-yellow, copper-green, cobalt-blue) on three spectrophotometers (Cary 14, Beckman DU, König-Martens). Except for the cobalt-blue standard, the values of spectral transmittance found differ from those previously assigned by amounts differing at some wavelengths by as much as or slightly more than the uncertainties estimated for the present values, though not by amounts exceeding the combined uncertainties of the present and previous determinations. The indicated changes for these three standards are fairly regular, however, and support the view that the selenium-red and carbon-yellow standards are changing chiefly by formation of a reflectance-reducing film on the surfaces. The indicated rate of upward drift is slow, and suggests that it takes about 10 years for the drift to exceed the assigned uncertainity.","PeriodicalId":94340,"journal":{"name":"Journal of research of the National Bureau of Standards. Section A, Physics and chemistry","volume":"31 1","pages":"577 - 584"},"PeriodicalIF":0.0000,"publicationDate":"1963-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"5","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of research of the National Bureau of Standards. Section A, Physics and chemistry","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6028/jres.067A.056","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 5

Abstract

In 1934, Gibson, Walker, and Brown developed sets of four colored glass filters to serve as working standards of spectral transmittance for checking the reliability of spectrophotometers. Several sets of these glasses were measured carefully and reserved and designated as future reference standards. Duplicate standards evaluated by comparison with the reference standards are available by purchase to the public. The current set of reference standards was established in the years 1945 to 1947, and one of these reference standards (selenium-red) was recalibrated in 1952. This paper reports a recalibration, made in 1961 and 1962, of all four glasses (selenium-red, carbon-yellow, copper-green, cobalt-blue) on three spectrophotometers (Cary 14, Beckman DU, König-Martens). Except for the cobalt-blue standard, the values of spectral transmittance found differ from those previously assigned by amounts differing at some wavelengths by as much as or slightly more than the uncertainties estimated for the present values, though not by amounts exceeding the combined uncertainties of the present and previous determinations. The indicated changes for these three standards are fairly regular, however, and support the view that the selenium-red and carbon-yellow standards are changing chiefly by formation of a reflectance-reducing film on the surfaces. The indicated rate of upward drift is slow, and suggests that it takes about 10 years for the drift to exceed the assigned uncertainity.
NBS玻璃光谱透过率标准的再标定
1934年,Gibson, Walker和Brown开发了一套四种彩色玻璃滤光片,作为检查分光光度计可靠性的光谱透射率的工作标准。有几套这样的玻璃杯被仔细地测量过,并保留下来作为将来的参考标准。通过与参考标准比较评估的重复标准可向公众购买。目前的一套参考标准是在1945年至1947年建立的,其中一个参考标准(硒红)于1952年重新校准。本文报告了1961年和1962年在三个分光光度计(Cary 14, Beckman DU, König-Martens)上对所有四种玻璃(硒红,碳黄,铜绿,钴蓝)进行的重新校准。除了钴蓝标准外,所发现的光谱透过率值与先前的值不同,在某些波长上的差异与当前值估计的不确定度相差相同或略大,但不超过当前和以前测定的不确定度总和。然而,这三种标准的变化是相当有规律的,这支持了硒红和碳黄标准的变化主要是由于表面上形成了一层减少反射的薄膜。所指示的向上漂移的速率是缓慢的,并且表明漂移需要大约10年的时间才能超过指定的不确定度。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信