{"title":"Using Signal Processing Techniques to Enhance the TemperatureMeasurements Using Nanohole Array Sensors","authors":"Masoud Modaresifar, G. Kowalski, D. Larson","doi":"10.4172/2153-2435.1000545","DOIUrl":null,"url":null,"abstract":"A signal processing approach to reducing noise in the observed transmission through nanohole array sensors used in a microscale calorimeter is presented. The results demonstrate that using rectification and moving average processes, a well-defined Extraordinary Optical Transmission (EOT)-temperature calibration curve can be developed in the presence of a moving light fringe pattern.","PeriodicalId":19833,"journal":{"name":"Pharmaceutica Analytica Acta","volume":"26 1","pages":"1-2"},"PeriodicalIF":0.0000,"publicationDate":"2017-04-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pharmaceutica Analytica Acta","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4172/2153-2435.1000545","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1
Abstract
A signal processing approach to reducing noise in the observed transmission through nanohole array sensors used in a microscale calorimeter is presented. The results demonstrate that using rectification and moving average processes, a well-defined Extraordinary Optical Transmission (EOT)-temperature calibration curve can be developed in the presence of a moving light fringe pattern.