{"title":"Flow-injection spectrophotometric determination of yttrium with Arsenazo III","authors":"Kate Grudpan, Wiboon Praditweangkum, Ponlayuth Sooksamiti, Robert Edwards","doi":"10.1002/(SICI)1098-2728(1999)11:5<279::AID-LRA6>3.0.CO;2-Y","DOIUrl":null,"url":null,"abstract":"We proposed a simple flow-injection spectrophotometric procedure for the determination of yttrium using Arsenazo III and optimized various conditions (pH for color formation and sample, buffer systems, Arsenazo III concentration, injection volume, flow-rate, and mixing coils). The calibration was linear at least up to 1 mgY/L, with a detection limit (3σ) of 0.02 mgY/L, RSD of 0.8% (n = 11 at 0.40 mgY/L). A throughput of 60 injections h−1 can be achieved. The procedure was applied to ore leachates of reference materials and samples. © 1999 John Wiley & Sons, Inc. Lab Robotics and Automation 11: 279–283, 1999","PeriodicalId":100863,"journal":{"name":"Laboratory Robotics and Automation","volume":"11 5","pages":"279-283"},"PeriodicalIF":0.0000,"publicationDate":"1999-10-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1002/(SICI)1098-2728(1999)11:5<279::AID-LRA6>3.0.CO;2-Y","citationCount":"3","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Laboratory Robotics and Automation","FirstCategoryId":"1085","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/%28SICI%291098-2728%281999%2911%3A5%3C279%3A%3AAID-LRA6%3E3.0.CO%3B2-Y","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 3
偶氮胂流动注射分光光度法测定钇
我们提出了一种简单的流动注射分光光度法,用偶氮胂III测定钇,并优化了各种条件(显色和样品的pH值、缓冲系统、偶氮胂III浓度、进样量、流速和混合线圈)。该方法在1 mgY/L以下呈线性,检出限(3σ)为0.02 mgY/L, RSD为0.8%(在0.40 mgY/L时n = 11)。可实现60次注入h−1的通量。该程序适用于标准物质和样品的矿石渗滤液。©1999 John Wiley &儿子,Inc。机械工程学报,2009,31 (2):379 - 383
本文章由计算机程序翻译,如有差异,请以英文原文为准。