{"title":"A new calibration concept of the Pulsed Electro Acoustic method allowing to improve charge density measurement","authors":"M. Arnaout, F. Baudoin, L. Berquez, D. Payan","doi":"10.1109/CEIDP.2011.6232643","DOIUrl":null,"url":null,"abstract":"This work is focused on the improvement of the calibration technique in order to obtain a well-conditioned transfer matrix in a Pulsed Electro-Acoustic -PEA- cell. A numerical electro-acoustic model has been developed with software Comsol. Using this model, we can establish the output voltage of the piezoelectric sensor and acoustic pressure at each point of the calculation domain. Our approach consists to recover charge distribution within the sample using deconvolution method and transfer matrix obtained with the new calibration technique. This new method has been tested with theoretical and experimental signals. The main conclusion is that an improvement of the spatial resolution is demonstrated, compared to the standard method usually used.","PeriodicalId":6317,"journal":{"name":"2011 Annual Report Conference on Electrical Insulation and Dielectric Phenomena","volume":"28 1","pages":"247-250"},"PeriodicalIF":0.0000,"publicationDate":"2011-10-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"3","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2011 Annual Report Conference on Electrical Insulation and Dielectric Phenomena","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/CEIDP.2011.6232643","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 3
Abstract
This work is focused on the improvement of the calibration technique in order to obtain a well-conditioned transfer matrix in a Pulsed Electro-Acoustic -PEA- cell. A numerical electro-acoustic model has been developed with software Comsol. Using this model, we can establish the output voltage of the piezoelectric sensor and acoustic pressure at each point of the calculation domain. Our approach consists to recover charge distribution within the sample using deconvolution method and transfer matrix obtained with the new calibration technique. This new method has been tested with theoretical and experimental signals. The main conclusion is that an improvement of the spatial resolution is demonstrated, compared to the standard method usually used.
本文的工作重点是改进校准技术,以便在脉冲电声- pea -细胞中获得条件良好的传递矩阵。利用Comsol软件建立了一个数值电声模型。利用该模型,可以建立计算域内各点的压电传感器输出电压和声压。我们的方法包括利用反褶积方法和利用新校准技术获得的传递矩阵来恢复样品内的电荷分布。这种新方法已经用理论和实验信号进行了验证。主要结论是,与通常使用的标准方法相比,证明了空间分辨率的提高。