Application of DQE for quantitative assessment of detectors to estimate AEC efficiency in digital mammography

IF 0.7 Q4 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
E. Fabiszewska, A. Wysocka-Rabin, M. Dobrzyńska, W. Skrzyński, K. Pasicz
{"title":"Application of DQE for quantitative assessment of detectors to estimate AEC efficiency in digital mammography","authors":"E. Fabiszewska, A. Wysocka-Rabin, M. Dobrzyńska, W. Skrzyński, K. Pasicz","doi":"10.2478/pjmpe-2021-0007","DOIUrl":null,"url":null,"abstract":"Abstract Optimisation of the detector’s exposure parameters settings for image quality and patient dose is an important task in digital mammography. Assessment of a digital detector’s performance can be done objectively and without operator bias by determining the Detective Quantum Efficiency (DQE). The authors of this article aim to prove that the performance of the AEC system can be objectively portrayed through DQE. The results were examined for influence of KAD changes on DQE values and to determine if it was possible to obtain similar DQE values for different exposures. While analysing the effect of the operation of the AEC system described with DQE, the doses received by women during mammography examinations were considered, as well. The AEC system’s exposure control mechanism cannot guarantee the same DQE value for different object thicknesses. When the object thickness increases, the AEC system should increase the KAD value to obtain the same DQE value. The result of increasing KAD would be the increase of mean glandular dose for some women. However, assuming that DQE is a good indicator of image quality, introducing the proposed changes to the AEC system’s operation would result in the same image quality for all breast thicknesses. This approach to DQE use for AEC system evaluation is independent of the image processing procedure and can be the basis for changes to system calibration done by the manufacturer’s technical support team.","PeriodicalId":53955,"journal":{"name":"Polish Journal of Medical Physics and Engineering","volume":"7 1","pages":"51 - 56"},"PeriodicalIF":0.7000,"publicationDate":"2021-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Polish Journal of Medical Physics and Engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.2478/pjmpe-2021-0007","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 1

Abstract

Abstract Optimisation of the detector’s exposure parameters settings for image quality and patient dose is an important task in digital mammography. Assessment of a digital detector’s performance can be done objectively and without operator bias by determining the Detective Quantum Efficiency (DQE). The authors of this article aim to prove that the performance of the AEC system can be objectively portrayed through DQE. The results were examined for influence of KAD changes on DQE values and to determine if it was possible to obtain similar DQE values for different exposures. While analysing the effect of the operation of the AEC system described with DQE, the doses received by women during mammography examinations were considered, as well. The AEC system’s exposure control mechanism cannot guarantee the same DQE value for different object thicknesses. When the object thickness increases, the AEC system should increase the KAD value to obtain the same DQE value. The result of increasing KAD would be the increase of mean glandular dose for some women. However, assuming that DQE is a good indicator of image quality, introducing the proposed changes to the AEC system’s operation would result in the same image quality for all breast thicknesses. This approach to DQE use for AEC system evaluation is independent of the image processing procedure and can be the basis for changes to system calibration done by the manufacturer’s technical support team.
应用DQE对检测器进行定量评估,以估计数字乳房x线摄影中AEC的效率
在数字化乳房x线摄影中,优化探测器的曝光参数设置以适应图像质量和患者剂量是一项重要任务。通过确定探测量子效率(DQE),可以客观、无算子偏差地评估数字探测器的性能。本文的作者旨在证明通过DQE可以客观地描述AEC系统的性能。检查了KAD变化对DQE值的影响,并确定是否有可能在不同的暴露下获得相似的DQE值。在分析用DQE描述的AEC系统操作的影响时,也考虑了妇女在乳房x光检查期间接受的剂量。AEC系统的曝光控制机制不能保证不同物体厚度的DQE值相同。当物体厚度增加时,AEC系统应增加KAD值,以获得相同的DQE值。KAD增加的结果可能是某些女性平均腺体剂量增加。然而,假设DQE是一个很好的图像质量指标,在AEC系统的操作中引入提议的更改将导致所有乳房厚度的图像质量相同。这种用于AEC系统评估的DQE方法独立于图像处理程序,可以作为制造商技术支持团队更改系统校准的基础。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Polish Journal of Medical Physics and Engineering
Polish Journal of Medical Physics and Engineering RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
1.30
自引率
0.00%
发文量
19
期刊介绍: Polish Journal of Medical Physics and Engineering (PJMPE) (Online ISSN: 1898-0309; Print ISSN: 1425-4689) is an official publication of the Polish Society of Medical Physics. It is a peer-reviewed, open access scientific journal with no publication fees. The issues are published quarterly online. The Journal publishes original contribution in medical physics and biomedical engineering.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信