{"title":"An automatic system for stepwise treatment of solid samples and application to pollution evaluation by measuring ion lixiviation rates in lichens","authors":"Mohamed Arhoun, José L. Martín-Herrera, Eva Barreno, Guillermo Ramis-Ramos","doi":"10.1002/(SICI)1098-2728(1999)11:2<121::AID-LRA8>3.0.CO;2-W","DOIUrl":null,"url":null,"abstract":"<p>An automatic system for the stepwise treatment of solid samples is described. The system is controlled by a PC computer that is connected to a burette equipped with a three-way valve, to a peristaltic pump, and to a robot arm through two serial interfaces. In a cycle, the burette introduces a solvent volume in the sample cell, sonication is applied, and, after a fixed time, the pump carries the solvent to the vial selected by the robot arm. Measurements can be performed both in-line and off-line by analyzing the vial series. The program was written in Visual Basic 3. Lixiviation rates of several cations from the lichen Ramalina farinacea were measured by capillary zone electrophoresis using the vial series obtained in several experimental conditions. The automatic system gave better reproducibility than the manual procedure. Initial lixiviation rates were higher in the presence of ultrasounds. © 1999 John Wiley & Sons, Inc. Lab Robotics and Automation 11: 121–126, 1999</p>","PeriodicalId":100863,"journal":{"name":"Laboratory Robotics and Automation","volume":"11 2","pages":"121-126"},"PeriodicalIF":0.0000,"publicationDate":"1999-04-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1002/(SICI)1098-2728(1999)11:2<121::AID-LRA8>3.0.CO;2-W","citationCount":"4","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Laboratory Robotics and Automation","FirstCategoryId":"1085","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/%28SICI%291098-2728%281999%2911%3A2%3C121%3A%3AAID-LRA8%3E3.0.CO%3B2-W","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 4
Abstract
An automatic system for the stepwise treatment of solid samples is described. The system is controlled by a PC computer that is connected to a burette equipped with a three-way valve, to a peristaltic pump, and to a robot arm through two serial interfaces. In a cycle, the burette introduces a solvent volume in the sample cell, sonication is applied, and, after a fixed time, the pump carries the solvent to the vial selected by the robot arm. Measurements can be performed both in-line and off-line by analyzing the vial series. The program was written in Visual Basic 3. Lixiviation rates of several cations from the lichen Ramalina farinacea were measured by capillary zone electrophoresis using the vial series obtained in several experimental conditions. The automatic system gave better reproducibility than the manual procedure. Initial lixiviation rates were higher in the presence of ultrasounds. © 1999 John Wiley & Sons, Inc. Lab Robotics and Automation 11: 121–126, 1999
一种用于固体样品分步处理及地衣中离子浸出率污染评价的自动系统
描述了一种用于固体样品逐步处理的自动系统。该系统由一台PC电脑控制,该电脑通过两个串行接口连接到装有三通阀的滴管、蠕动泵和机械臂上。在一个循环中,滴管将溶剂体积引入样品池,应用超声,并在固定时间后,泵将溶剂输送到机械臂选择的小瓶中。通过分析小瓶系列,可以进行在线和离线测量。该程序是用Visual Basic 3编写的。采用毛细管区带电泳法测定了在不同实验条件下得到的小瓶系列产品中几种阳离子的浸出率。自动系统的重现性比人工程序好。超声在场时,初始浸出率较高。©1999 John Wiley &儿子,Inc。实验室机器人与自动化,1999
本文章由计算机程序翻译,如有差异,请以英文原文为准。