Treatment of Thrombus Formation Associated With the MicroMed DeBakey VAD Using Recombinant Tissue Plasminogen Activator

M. Rothenburger, M. Wilhelm, D. Hammel, C. Schmidt, T. D. Tjan, D. Böcker, H. Scheld, Ç. Schmid
{"title":"Treatment of Thrombus Formation Associated With the MicroMed DeBakey VAD Using Recombinant Tissue Plasminogen Activator","authors":"M. Rothenburger, M. Wilhelm, D. Hammel, C. Schmidt, T. D. Tjan, D. Böcker, H. Scheld, Ç. Schmid","doi":"10.1161/01.CIR.0000032913.33237.F5","DOIUrl":null,"url":null,"abstract":"BackgroundThe latest generation of left ventricular assist devices consists of nonpulsatile impeller pumps. In these small pumps, thrombus formation inside the device does not lead to thromboembolic end-organ dysfunction but may dramatically impair pump flow. We report on our experience with thrombus-related pump dysfunctions of the MicroMed DeBakey left ventricular assist device and its treatment. MethodsEight of 22 patients with a MicroMed DeBakey VAD presented with a critically reduced pump flow. In 7 cases, an increased power demand indicative of progressive thrombus formation associated with the device was evident, whereas 1 case presented with thrombus formation within the inflow conduit associated with a very low power demand. Brief spontaneously resolving pump stops had been noted in 6 patients. All 8 patients were treated with 100 mg of recombinant tissue plasminogen activator (rt-PA), administered via an IV line. ResultsRt-PA lysis led to an increase of pump flow along with a reduction of power demand within a short time in all patients. No severe bleeding complications occurred. However, 4 patients experienced transient epistaxis. All patients could be discharged from intensive care immediately after discontinuation of thrombolytic therapy. ConclusionRt-PA lysis is a very effective tool for thrombus-related pump dysfunction in patients with impeller pumps, which renders emergency surgical exchange unnecessary in most cases.","PeriodicalId":10194,"journal":{"name":"Circulation: Journal of the American Heart Association","volume":"18 1","pages":"I-189-I-192"},"PeriodicalIF":0.0000,"publicationDate":"2002-09-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"88","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Circulation: Journal of the American Heart Association","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1161/01.CIR.0000032913.33237.F5","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 88

Abstract

BackgroundThe latest generation of left ventricular assist devices consists of nonpulsatile impeller pumps. In these small pumps, thrombus formation inside the device does not lead to thromboembolic end-organ dysfunction but may dramatically impair pump flow. We report on our experience with thrombus-related pump dysfunctions of the MicroMed DeBakey left ventricular assist device and its treatment. MethodsEight of 22 patients with a MicroMed DeBakey VAD presented with a critically reduced pump flow. In 7 cases, an increased power demand indicative of progressive thrombus formation associated with the device was evident, whereas 1 case presented with thrombus formation within the inflow conduit associated with a very low power demand. Brief spontaneously resolving pump stops had been noted in 6 patients. All 8 patients were treated with 100 mg of recombinant tissue plasminogen activator (rt-PA), administered via an IV line. ResultsRt-PA lysis led to an increase of pump flow along with a reduction of power demand within a short time in all patients. No severe bleeding complications occurred. However, 4 patients experienced transient epistaxis. All patients could be discharged from intensive care immediately after discontinuation of thrombolytic therapy. ConclusionRt-PA lysis is a very effective tool for thrombus-related pump dysfunction in patients with impeller pumps, which renders emergency surgical exchange unnecessary in most cases.
重组组织型纤溶酶原激活剂治疗MicroMed DeBakey VAD相关血栓形成
最新一代的左心室辅助装置由非脉动叶轮泵组成。在这些小泵中,装置内的血栓形成不会导致血栓栓塞性终末器官功能障碍,但可能严重损害泵的流动。我们报告了MicroMed DeBakey左心室辅助装置的血栓相关泵功能障碍及其治疗的经验。方法22例MicroMed DeBakey VAD患者中8例出现泵流量严重减少。在7例患者中,明显出现了与该装置相关的进行性血栓形成的电力需求增加,而1例患者出现了与非常低的电力需求相关的流入导管内血栓形成。6例患者出现短暂的自行缓解的泵停。所有8例患者均接受100 mg重组组织型纤溶酶原激活剂(rt-PA),通过静脉滴注。结果rt - pa溶解导致所有患者在短时间内泵流量增加,功率需求降低。无严重出血并发症发生。然而,4例患者出现短暂性鼻出血。所有患者均可在停止溶栓治疗后立即出院。结论对于叶轮泵患者血栓相关性泵功能障碍,rt - pa溶解是一种非常有效的治疗手段,多数情况下无需紧急手术置换。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信