DS18B20 sensor calibration compared with fluke hart scientific standard sensor

Asep Nata Permana, I. M. S. Wibawa, I. K. Putra
{"title":"DS18B20 sensor calibration compared with fluke hart scientific standard sensor","authors":"Asep Nata Permana, I. M. S. Wibawa, I. K. Putra","doi":"10.31295/IJPM.V4N1.1225","DOIUrl":null,"url":null,"abstract":"--The DS18B20 sensor calibration has been conducted using a fluke hart with a temperature chamber at the center for meteorology, climatology, and geophysics region III Denpasar. The calibration process is done by stabilizing the temperature in the chamber, then the temperature on the DS18B20 sensor and the standard thermometer in the chamber are read and compared. The setpoints used were 10 C, 20 C, 30 C and 40 C each set point was read four times. The results of the calculation of the standard deviation at each setpoint are ± 0,39 C, ± 0,12 C, ± 0,00 C, ± 0,00 C and the results of the uncertainty for each setpoint are 0,65 C, ± 0,20 C, ± 0,18 C, ± 0,18 C. The results of the DS18B20 sensor calibration using a fluke hart with a temperature chamber obtained the highest standard deviation value of ±0,39 C and the highest uncertainty value was 0,65 C. This value indicates that the DS18B20 sensor calibrated with a standard tool at the 10 C set point has not met the standard value applied by WMO, which is ±0.3 C. Keywords---calibration, DS18B20 sensor, fluke hart, temperature chamber.","PeriodicalId":22540,"journal":{"name":"The International Journal of Physics","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2021-03-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"The International Journal of Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.31295/IJPM.V4N1.1225","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 2

Abstract

--The DS18B20 sensor calibration has been conducted using a fluke hart with a temperature chamber at the center for meteorology, climatology, and geophysics region III Denpasar. The calibration process is done by stabilizing the temperature in the chamber, then the temperature on the DS18B20 sensor and the standard thermometer in the chamber are read and compared. The setpoints used were 10 C, 20 C, 30 C and 40 C each set point was read four times. The results of the calculation of the standard deviation at each setpoint are ± 0,39 C, ± 0,12 C, ± 0,00 C, ± 0,00 C and the results of the uncertainty for each setpoint are 0,65 C, ± 0,20 C, ± 0,18 C, ± 0,18 C. The results of the DS18B20 sensor calibration using a fluke hart with a temperature chamber obtained the highest standard deviation value of ±0,39 C and the highest uncertainty value was 0,65 C. This value indicates that the DS18B20 sensor calibrated with a standard tool at the 10 C set point has not met the standard value applied by WMO, which is ±0.3 C. Keywords---calibration, DS18B20 sensor, fluke hart, temperature chamber.
DS18B20传感器与fluke hart科学标准传感器的标定比较
DS18B20传感器校准已在气象、气候学和地球物理III区登巴萨中心使用带有温度室的吸虫hart进行。校准过程是通过稳定实验室内的温度来完成的,然后读取DS18B20传感器上的温度和实验室内标准温度计上的温度并进行比较。使用的设定值为10℃、20℃、30℃和40℃,每个设定值读取四次。每个设定值的标准差计算结果为±0.39℃,±0.12℃,±0.000℃,±0.000℃,每个设定值的不确定度计算结果为0.65℃,±0.20℃,±0.18℃,采用带温度室的吸片哈特校准DS18B20传感器的结果得到的最高标准偏差值为±0.39℃,最高不确定度值为0.65℃,这表明在10℃设定点用标准工具校准的DS18B20传感器未达到WMO应用的±0.3℃标准值。关键词:校准,DS18B20传感器,吸片哈特,温度室。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
文献相关原料
公司名称 产品信息 采购帮参考价格
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信