Assessment of an Aggregate-Level Hand Hygiene Monitoring Technology for Measuring Hand Hygiene Performance Among Healthcare Personnel

H. Limper, L. Slawsky, Sylvia Garcia-Houchins, S. Mehta, R. Hershow, E. Landon
{"title":"Assessment of an Aggregate-Level Hand Hygiene Monitoring Technology for Measuring Hand Hygiene Performance Among Healthcare Personnel","authors":"H. Limper, L. Slawsky, Sylvia Garcia-Houchins, S. Mehta, R. Hershow, E. Landon","doi":"10.1017/ice.2016.298","DOIUrl":null,"url":null,"abstract":"BACKGROUND Despite significant advances in technological methods for hand hygiene surveillance, a lack of evidence prohibits comparison of systems to one another or against the current gold standard of direct observation. OBJECTIVE To validate a hand hygiene monitoring technology (HHMT) designed to capture hand hygiene behaviors aggregated at the hospital-unit level (GOJO Industries, Akron, OH). METHODS Our team followed a rigorous validation approach to assess the sensitivity and positive predictive value (PPV) of an HHMT. A planned path was first used to measure the accuracy of the system when purposefully activated by investigators. Next, behavioral validation was used to quantify accuracy of the system in capturing real-world behaviors. RESULTS During the planned path phase, investigators performed 4,872 unique events across 3 distinct hospital buildings varying in size and age since construction. Overall sensitivity across the medical center was 88.7% with a PPV of 99.2%. During the behavioral validation phase, trained direct observers recorded 5,539 unique events across 3 distinct hospital buildings. Overall sensitivity across the medical center was 92.7% and PPV was 84.4%. CONCLUSION Objective measures of sensitivity and PPV indicate the promise of the benefit of this and other HHMTs to capture basic behaviors associated with hand hygiene. Infect Control Hosp Epidemiol 2017;38:348–352","PeriodicalId":13655,"journal":{"name":"Infection Control & Hospital Epidemiology","volume":"23 1","pages":"348 - 352"},"PeriodicalIF":0.0000,"publicationDate":"2016-12-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"20","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Infection Control & Hospital Epidemiology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1017/ice.2016.298","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 20

Abstract

BACKGROUND Despite significant advances in technological methods for hand hygiene surveillance, a lack of evidence prohibits comparison of systems to one another or against the current gold standard of direct observation. OBJECTIVE To validate a hand hygiene monitoring technology (HHMT) designed to capture hand hygiene behaviors aggregated at the hospital-unit level (GOJO Industries, Akron, OH). METHODS Our team followed a rigorous validation approach to assess the sensitivity and positive predictive value (PPV) of an HHMT. A planned path was first used to measure the accuracy of the system when purposefully activated by investigators. Next, behavioral validation was used to quantify accuracy of the system in capturing real-world behaviors. RESULTS During the planned path phase, investigators performed 4,872 unique events across 3 distinct hospital buildings varying in size and age since construction. Overall sensitivity across the medical center was 88.7% with a PPV of 99.2%. During the behavioral validation phase, trained direct observers recorded 5,539 unique events across 3 distinct hospital buildings. Overall sensitivity across the medical center was 92.7% and PPV was 84.4%. CONCLUSION Objective measures of sensitivity and PPV indicate the promise of the benefit of this and other HHMTs to capture basic behaviors associated with hand hygiene. Infect Control Hosp Epidemiol 2017;38:348–352
一种测量医护人员手卫生表现的整体水平手卫生监测技术的评估
背景:尽管手部卫生监测的技术方法取得了重大进展,但缺乏证据阻碍了系统之间的比较或与当前直接观察的黄金标准的比较。目的验证一种手卫生监测技术(HHMT),该技术旨在捕获医院-单位层面的手卫生行为(GOJO Industries, Akron, OH)。方法采用严格的验证方法评估HHMT的敏感性和阳性预测值(PPV)。当调查员有目的地激活时,首先使用计划路径来测量系统的准确性。接下来,使用行为验证来量化系统在捕获现实世界行为方面的准确性。结果:在规划路径阶段,研究人员在3座不同的医院建筑中进行了4872次独特的事件,这些建筑自建成以来的规模和年龄都有所不同。整个医疗中心的总体敏感性为88.7%,PPV为99.2%。在行为验证阶段,训练有素的直接观察员在3个不同的医院建筑物中记录了5,539个独特事件。整个医疗中心的总体敏感性为92.7%,PPV为84.4%。结论:敏感性和PPV的客观测量表明,该方法和其他HHMTs有望捕获与手卫生相关的基本行为。中华流行病学杂志,2017;38 (4):349 - 349
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信