{"title":"Identification of contagious SARS-CoV-2 infected individuals by Roche's Rapid Antigen Test.","authors":"Harald H Kessler, Florian Prüller, Melina Hardt, Evelyn Stelzl, Esther Föderl-Höbenreich, Sabine Pailer, Andreas Lueger, Philipp Kreuzer, Kurt Zatloukal, Markus Herrmann","doi":"10.1515/cclm-2021-1276","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>Rapid antigen tests (RAT) can provide valuable information on the presence or absence SARS-CoV-2 within 15 min without the need of a laboratory. The analytical and diagnostic characteristics of available RATs has led to the question whether they can safely distinguish between infectious and non-infectious patients in an acute care setting.</p><p><strong>Methods: </strong>Three nasopharyngeal swabs for the analysis by RAT, reverse transcriptase real time polymerase chain reaction (RT-qPCR), and a cell culture based infection assay were collected from 67 patients that presented to the emergency department of the University Hospital of Graz (Austria). The first swab was used for on-site RAT testing in the emergency department using the Roche SARS-CoV-2 RAT. The second swab was sent to the central laboratory of the hospital for RT-qPCR with two independent methods (Cepheid Xpert<sup>®</sup> Xpress SARS-CoV-2 assay and Roche Cobas SARS-CoV-2 Test) and repeat RAT testing using the same commercial test. With the third swab a cell culture-based infection assay was performed.</p><p><strong>Results: </strong>The RATs performed from independent samples showed substantial agreement (Cohen's-kappa: 0.73, p<0.001). All patients with a positive RAT had positive RT-qPCR with cycle threshold (ct) values <25. Fifteen out of 55 RAT-negative samples were RT-qPCR positive with ct values between 25 and 40. The inoculation of cell cultures with RT-qPCR negative swabs and RT-qPCR positive swabs with ct values >25 did not induce cytopathic effects that were related to SARS-CoV-2. The infection assays from four RAT-negative patients showed cytopathic effects that were induced by other pathogens.</p><p><strong>Conclusions: </strong>The SARS-CoV-2 RAT from Roche Diagnostics is a valuable tool for managing symptomatic patients. RAT-negative patients may be regarded as non-contagious.</p>","PeriodicalId":10388,"journal":{"name":"Clinical Chemistry and Laboratory Medicine (CCLM)","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2022-03-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"7","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Clinical Chemistry and Laboratory Medicine (CCLM)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1515/cclm-2021-1276","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2022/4/26 0:00:00","PubModel":"Print","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 7
Abstract
Objectives: Rapid antigen tests (RAT) can provide valuable information on the presence or absence SARS-CoV-2 within 15 min without the need of a laboratory. The analytical and diagnostic characteristics of available RATs has led to the question whether they can safely distinguish between infectious and non-infectious patients in an acute care setting.
Methods: Three nasopharyngeal swabs for the analysis by RAT, reverse transcriptase real time polymerase chain reaction (RT-qPCR), and a cell culture based infection assay were collected from 67 patients that presented to the emergency department of the University Hospital of Graz (Austria). The first swab was used for on-site RAT testing in the emergency department using the Roche SARS-CoV-2 RAT. The second swab was sent to the central laboratory of the hospital for RT-qPCR with two independent methods (Cepheid Xpert® Xpress SARS-CoV-2 assay and Roche Cobas SARS-CoV-2 Test) and repeat RAT testing using the same commercial test. With the third swab a cell culture-based infection assay was performed.
Results: The RATs performed from independent samples showed substantial agreement (Cohen's-kappa: 0.73, p<0.001). All patients with a positive RAT had positive RT-qPCR with cycle threshold (ct) values <25. Fifteen out of 55 RAT-negative samples were RT-qPCR positive with ct values between 25 and 40. The inoculation of cell cultures with RT-qPCR negative swabs and RT-qPCR positive swabs with ct values >25 did not induce cytopathic effects that were related to SARS-CoV-2. The infection assays from four RAT-negative patients showed cytopathic effects that were induced by other pathogens.
Conclusions: The SARS-CoV-2 RAT from Roche Diagnostics is a valuable tool for managing symptomatic patients. RAT-negative patients may be regarded as non-contagious.