Highly Portable and Sensitive Filter-Based Antibody-or DNA Aptamer-Enzyme-Linked Fluorescence Detection of Potential Bacterial Pathogens Proximal to Agricultural Fields

J. Bruno
{"title":"Highly Portable and Sensitive Filter-Based Antibody-or DNA Aptamer-Enzyme-Linked Fluorescence Detection of Potential Bacterial Pathogens Proximal to Agricultural Fields","authors":"J. Bruno","doi":"10.1166/asem.2020.2627","DOIUrl":null,"url":null,"abstract":"A novel, sensitive, rapid and highly portable system consisting of a membrane filtration unit connected to a large syringe with tubing was used to detect Escherchia coli and Salmonella enterica bacteria in a mobile laboratory setting. These simple fluorescent enzyme-linked\n molecular assays consisted of biotinylated DNA aptamers or antibodies which were allowed to bind the target bacteria and be deposited on 0.45 μm pore size membrane filters by suction, followed by binding of streptavidin-peroxidase conjugate, wash steps and development in Amplex® UltraRed\n (AUR; resazurin derivative) with hydrogen peroxide activation. Fluorescence values of developed AUR from the filter surface were assessed with a commercially-available handheld fluorometer. Preliminary data indicate detection limits in the range of 50–110 E. coli and Salmonella\n bacterial cells per sample with good specificity and total processing times under one hour.","PeriodicalId":7213,"journal":{"name":"Advanced Science, Engineering and Medicine","volume":"13 1","pages":"909-913"},"PeriodicalIF":0.0000,"publicationDate":"2020-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Advanced Science, Engineering and Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1166/asem.2020.2627","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

A novel, sensitive, rapid and highly portable system consisting of a membrane filtration unit connected to a large syringe with tubing was used to detect Escherchia coli and Salmonella enterica bacteria in a mobile laboratory setting. These simple fluorescent enzyme-linked molecular assays consisted of biotinylated DNA aptamers or antibodies which were allowed to bind the target bacteria and be deposited on 0.45 μm pore size membrane filters by suction, followed by binding of streptavidin-peroxidase conjugate, wash steps and development in Amplex® UltraRed (AUR; resazurin derivative) with hydrogen peroxide activation. Fluorescence values of developed AUR from the filter surface were assessed with a commercially-available handheld fluorometer. Preliminary data indicate detection limits in the range of 50–110 E. coli and Salmonella bacterial cells per sample with good specificity and total processing times under one hour.
高度便携和敏感的基于过滤器的抗体或DNA适配体-酶联荧光检测近地农田潜在细菌病原体
一种新型、灵敏、快速和高度便携的系统,由一个膜过滤单元连接到一个带管的大型注射器组成,用于在移动实验室环境中检测大肠杆菌和肠沙门氏菌。这些简单的荧光酶联分子检测包括生物素化的DNA适体或抗体,这些适体或抗体被允许与目标细菌结合,并通过吸力沉积在0.45 μm孔径的膜过滤器上,然后结合链亲和素-过氧化物酶偶联物,清洗步骤和Amplex®UltraRed (AUR;reazurin衍生物)过氧化氢活化。利用市售的手持式荧光计对滤光片表面形成的AUR的荧光值进行评估。初步数据表明,每个样品的检出限为50-110个大肠杆菌和沙门氏菌细胞,具有良好的特异性,总处理时间低于1小时。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信