Use of a 2-Dimensional Ionization Chamber Array to Measure Head Leakage of a Varian Truebeam® Linear Accelerator

S. Taneja, J. Teruel, Lei Hu, J. Xue, D. Barbee
{"title":"Use of a 2-Dimensional Ionization Chamber Array to Measure Head Leakage of a Varian Truebeam® Linear Accelerator","authors":"S. Taneja, J. Teruel, Lei Hu, J. Xue, D. Barbee","doi":"10.4236/ijmpcero.2020.93009","DOIUrl":null,"url":null,"abstract":"The purpose of this work is to evaluate the use of a two-dimensional (2D) planar ion chamber array to characterize leakage radiation from the head of the linear accelerator. Ion chamber arrays provide a benefit over a singular ion chamber measurement as they allow for the measurement of a larger area in order to isolate the point of maximum leakage dose and the small size of each individual ion chamber minimizes volume-averaging effects. A Varian Truebeam® undergoing acceptance testing was used for all measurements. The gantry was wrapped in Portal Pack for Localization (PPL) radiographic film in order to isolate the location of maximum leakage. A calibration curve was developed and used to determine dose-to-film. An Ion Chamber Profiler (IC Profiler™) manufactured by Sun Nuclear Corporation was used to confirm measurements by the PPL film. All measurements were normalized to leakage at 100 cm from the target relative to the central axis. Three points were investigated with the IC Profiler, including the top of the gantry, the Varian logo, and the side of the gantry. For the three locations, respectively, the PPL film and the IC profiler were measured 0.142% and 0.131%, 0.036% and 0.030%, and 0.014% and 0.019%. The good agreement between the PPL film and the IC Profiler provides confidence in the use of a more efficient and accurate ion chamber array for head leakage measurements.","PeriodicalId":14028,"journal":{"name":"International Journal of Medical Physics, Clinical Engineering and Radiation Oncology","volume":"75 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2020-06-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Medical Physics, Clinical Engineering and Radiation Oncology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4236/ijmpcero.2020.93009","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 2

Abstract

The purpose of this work is to evaluate the use of a two-dimensional (2D) planar ion chamber array to characterize leakage radiation from the head of the linear accelerator. Ion chamber arrays provide a benefit over a singular ion chamber measurement as they allow for the measurement of a larger area in order to isolate the point of maximum leakage dose and the small size of each individual ion chamber minimizes volume-averaging effects. A Varian Truebeam® undergoing acceptance testing was used for all measurements. The gantry was wrapped in Portal Pack for Localization (PPL) radiographic film in order to isolate the location of maximum leakage. A calibration curve was developed and used to determine dose-to-film. An Ion Chamber Profiler (IC Profiler™) manufactured by Sun Nuclear Corporation was used to confirm measurements by the PPL film. All measurements were normalized to leakage at 100 cm from the target relative to the central axis. Three points were investigated with the IC Profiler, including the top of the gantry, the Varian logo, and the side of the gantry. For the three locations, respectively, the PPL film and the IC profiler were measured 0.142% and 0.131%, 0.036% and 0.030%, and 0.014% and 0.019%. The good agreement between the PPL film and the IC Profiler provides confidence in the use of a more efficient and accurate ion chamber array for head leakage measurements.
使用二维电离室阵列测量瓦里安Truebeam线性加速器的封头泄漏
这项工作的目的是评估使用二维平面离子室阵列来表征线性加速器头部的泄漏辐射。离子室阵列提供了比单一离子室测量更大的好处,因为它们允许测量更大的区域,以隔离最大泄漏剂量的点,并且每个单独离子室的小尺寸最大限度地减少了体积平均效应。所有测量均使用经过验收测试的瓦里安Truebeam®。为了隔离最大泄漏的位置,门框被包裹在定位门户包(PPL)射线摄影胶片中。建立了一条校准曲线,并用于测定剂量-膜比。使用太阳核公司生产的离子室分析器(IC Profiler™)来确认PPL膜的测量。所有测量归一化为距离目标相对于中心轴100厘米处的泄漏。三点被调查与IC分析器,包括龙门架的顶部,瓦里安标志,和龙门架的一面。在三个位置,PPL薄膜和IC轮廓仪的测量值分别为0.142%和0.131%,0.036%和0.030%,0.014%和0.019%。PPL薄膜和IC分析器之间的良好一致性为使用更有效和准确的离子室阵列进行封头泄漏测量提供了信心。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信