Performance Evaluation of a Linear Pulse Motor-Driven Intra-Aortic Balloon Pump

K. Kobayashi, K. Suzuki, H. Nakagawa, H. Yamada
{"title":"Performance Evaluation of a Linear Pulse Motor-Driven Intra-Aortic Balloon Pump","authors":"K. Kobayashi, K. Suzuki, H. Nakagawa, H. Yamada","doi":"10.3379/JMSJMAG.18.525","DOIUrl":null,"url":null,"abstract":"An intra-aortic balloon pump (IABP) is an effective device for supporting circulation in the event of acute heart failure. Small IABP devices are in urgent demand in medical fields. The authors have investigated an IABP device which uses a linear pulse motor (LPM) in order to reduce the device size. This paper discusses the characteristics of the LPM developed by the authors to drive the IABP, and describes the results of evaluation of an LPM-driven IABP. Our investigation revealed the following. (1) The trial LPM was 1.3 L in volume and had a mass of 6.3 kg. Its use will reduce the weight of IABP devices. (2) The LPM has a high thrust/volume ratio of 0.88 N/ml. (3) The driving characteristic of the LPM-driven IABP was evaluated using a mock testing device. The maximum pumping rate of this IABP device is approximately 120 beats/min. As a result, an LPM-driven IABP will be capable of sufficiently high performance to enable clinical use.","PeriodicalId":100647,"journal":{"name":"IEEE Translation Journal on Magnetics in Japan","volume":"-3 1","pages":"98-103"},"PeriodicalIF":0.0000,"publicationDate":"1994-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE Translation Journal on Magnetics in Japan","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3379/JMSJMAG.18.525","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 2

Abstract

An intra-aortic balloon pump (IABP) is an effective device for supporting circulation in the event of acute heart failure. Small IABP devices are in urgent demand in medical fields. The authors have investigated an IABP device which uses a linear pulse motor (LPM) in order to reduce the device size. This paper discusses the characteristics of the LPM developed by the authors to drive the IABP, and describes the results of evaluation of an LPM-driven IABP. Our investigation revealed the following. (1) The trial LPM was 1.3 L in volume and had a mass of 6.3 kg. Its use will reduce the weight of IABP devices. (2) The LPM has a high thrust/volume ratio of 0.88 N/ml. (3) The driving characteristic of the LPM-driven IABP was evaluated using a mock testing device. The maximum pumping rate of this IABP device is approximately 120 beats/min. As a result, an LPM-driven IABP will be capable of sufficiently high performance to enable clinical use.
线性脉冲马达驱动主动脉内球囊泵的性能评价
主动脉内球囊泵(IABP)是急性心力衰竭时支持血液循环的有效装置。小型IABP设备在医疗领域的需求非常迫切。为了减小器件尺寸,作者研究了一种使用线性脉冲电机(LPM)的IABP器件。本文讨论了LPM驱动IABP的特点,并描述了LPM驱动IABP的评价结果。我们的调查结果如下。(1)试验LPM体积为1.3 L,质量为6.3 kg。它的使用将减轻IABP设备的重量。(2) LPM具有较高的推力/体积比0.88 N/ml。(3)利用模拟试验装置对lpm驱动的IABP的驱动特性进行了评价。该IABP装置的最大泵送速率约为120次/分钟。因此,lpm驱动的IABP将具有足够高的性能,可以用于临床。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信