Uji Biokompabilitas pada Implan Orthopedi Antara Implan Impor, Implan Lokal dari Material Impor, dan Prototipe Stainless Steel 316L dari Material Lokal
{"title":"Uji Biokompabilitas pada Implan Orthopedi Antara Implan Impor, Implan Lokal dari Material Impor, dan Prototipe Stainless Steel 316L dari Material Lokal","authors":"Ferdiansyah Mahyudin, Heppy Chandra Waskita, Dwikora Novembri Utomo, Heri Suroto, Triwahyudi Martanto","doi":"10.30651/JQM.V3I1.1612","DOIUrl":null,"url":null,"abstract":"ABSTRACT Demands of orthopedic implants are increasing in Indonesia, but its production are limited and still imported. Stainless steel, especially 316L is one of metal that is used commonly. This research evaluates biocompatibility of stainless steel 316L implant prototype, local implant manufactured using imported stainless steel, and imported stainless steel implant. Cytotoxicity test used mesenchymal human cell with MTT assay. Irritation test, acute toxicity/pyrogenicity test, and implantation test used 24 Wistar rats respectively which were assigned to 4 groups. We evaluate Draize score, weight and rectal temperature changes, and fibrous capsule ring thickness. Cytotoxicity test shows viable cell percentage in stainless steel 316L implant prototype, local implant, and imported implant are 83.7%, 87.5%, and 83.7% respectively. Comparison test using One Way ANOVA shows no significant difference between 3 groups (p>0.05). Iritation test results in 3 groups are grade 0 (no edema and no erythema). Acute toxicity/pyrogenicity test shows no statistically significant difference (p>0.05) among all groups. There is also no significant difference (p>0.05) of fibrous capsule ring thickness on 1 and 4 weeks after implantation. Based on ISO for medical test guide, stainless steel 316L implant prototype is safe and has similar biocompatibility test result compared to local and imported implant.Keywords : stainless steel 316L, biocompatibility, orthopaedic implantCorrespondence to : ferdiansyah@fk.unair.ac.id ABSTRAKKebutuhan implan orthopedi terus meningkat di Indonesia, namun produksinya terbatas dan sebagian besar masih impor dari luar negeri. Stainless steel, utama 316L adalah salah satu metal yang sering digunakan. Penelitian ini mengevaluasi biokompatibilitas dari prototipe implan stainless steel 316L, implan lokal yang diproduksi menggunakan stainless steel impor, dan implan stainless steel impor. Uji sitotoksisitas menggunakan sel mesenkim human dengan MTT assay. Uji iritasi, uji pirogenositas, dan uji implantasi menggunakan 24 tikus Wistar yang terbagi dalam 4 grup. Kami mengevaluasi skor Draize, perubahan berat dan temperatur rektal, dan ketebalan fibrous capsule ring. Uji sitotoksisitas menunjukkan persentase viabilitas sel pada prototipe stainless steel 316L, implan lokal, dan implan impor berturut-turut adalah 83,7%; 87,5%; dan 83,7%. Uji komparasi dengan One Way Anova tidak menunjukkan perbedaan signifikan diantara 3 grup (p>0,05). Hasil uji iritasi 3 grup menunjukkan grade 0 (tidak ada edema dan eritema). Uji pirogenositas tidak menunjukkan perbedaan signifikan pada semua grup (p>0,05). Tidak ada perbedaan signifikan (p>0,05) pula pada ketebalan fibrous capsule ring di minggu ke 1 dan 4 setelah implantasi. Berdasarkan ISO untuk pedoman uji di bidang medis, prototipe stainless steel 316L aman dan memiliki kesamaan hasil uji biokompatibilitas bila dibandingkan dengan implan lokal dan impor.Kata kunci : stainless steel 316L, uji biokompatibilitas, implan orthopaediKorespondensi : ferdiansyah@fk.unair.ac.id","PeriodicalId":31682,"journal":{"name":"Qanun Medika Jurnal Kedokteran Fakultas Kedokteran Universitas Muhammadiyah Surabaya","volume":"52 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2019-01-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Qanun Medika Jurnal Kedokteran Fakultas Kedokteran Universitas Muhammadiyah Surabaya","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.30651/JQM.V3I1.1612","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1
Abstract
ABSTRACT Demands of orthopedic implants are increasing in Indonesia, but its production are limited and still imported. Stainless steel, especially 316L is one of metal that is used commonly. This research evaluates biocompatibility of stainless steel 316L implant prototype, local implant manufactured using imported stainless steel, and imported stainless steel implant. Cytotoxicity test used mesenchymal human cell with MTT assay. Irritation test, acute toxicity/pyrogenicity test, and implantation test used 24 Wistar rats respectively which were assigned to 4 groups. We evaluate Draize score, weight and rectal temperature changes, and fibrous capsule ring thickness. Cytotoxicity test shows viable cell percentage in stainless steel 316L implant prototype, local implant, and imported implant are 83.7%, 87.5%, and 83.7% respectively. Comparison test using One Way ANOVA shows no significant difference between 3 groups (p>0.05). Iritation test results in 3 groups are grade 0 (no edema and no erythema). Acute toxicity/pyrogenicity test shows no statistically significant difference (p>0.05) among all groups. There is also no significant difference (p>0.05) of fibrous capsule ring thickness on 1 and 4 weeks after implantation. Based on ISO for medical test guide, stainless steel 316L implant prototype is safe and has similar biocompatibility test result compared to local and imported implant.Keywords : stainless steel 316L, biocompatibility, orthopaedic implantCorrespondence to : ferdiansyah@fk.unair.ac.id ABSTRAKKebutuhan implan orthopedi terus meningkat di Indonesia, namun produksinya terbatas dan sebagian besar masih impor dari luar negeri. Stainless steel, utama 316L adalah salah satu metal yang sering digunakan. Penelitian ini mengevaluasi biokompatibilitas dari prototipe implan stainless steel 316L, implan lokal yang diproduksi menggunakan stainless steel impor, dan implan stainless steel impor. Uji sitotoksisitas menggunakan sel mesenkim human dengan MTT assay. Uji iritasi, uji pirogenositas, dan uji implantasi menggunakan 24 tikus Wistar yang terbagi dalam 4 grup. Kami mengevaluasi skor Draize, perubahan berat dan temperatur rektal, dan ketebalan fibrous capsule ring. Uji sitotoksisitas menunjukkan persentase viabilitas sel pada prototipe stainless steel 316L, implan lokal, dan implan impor berturut-turut adalah 83,7%; 87,5%; dan 83,7%. Uji komparasi dengan One Way Anova tidak menunjukkan perbedaan signifikan diantara 3 grup (p>0,05). Hasil uji iritasi 3 grup menunjukkan grade 0 (tidak ada edema dan eritema). Uji pirogenositas tidak menunjukkan perbedaan signifikan pada semua grup (p>0,05). Tidak ada perbedaan signifikan (p>0,05) pula pada ketebalan fibrous capsule ring di minggu ke 1 dan 4 setelah implantasi. Berdasarkan ISO untuk pedoman uji di bidang medis, prototipe stainless steel 316L aman dan memiliki kesamaan hasil uji biokompatibilitas bila dibandingkan dengan implan lokal dan impor.Kata kunci : stainless steel 316L, uji biokompatibilitas, implan orthopaediKorespondensi : ferdiansyah@fk.unair.ac.id
印度尼西亚对骨科植入物的需求正在增加,但其生产有限,仍需进口。不锈钢,尤其是316L是最常用的金属之一。本研究对不锈钢316L种植体原型、国产进口不锈钢种植体和进口不锈钢种植体的生物相容性进行了评价。细胞毒性试验采用人间充质细胞MTT法。24只Wistar大鼠分为4组,分别进行刺激试验、急性毒性/热原性试验和着床试验。我们评估Draize评分、体重和直肠温度变化以及纤维囊环厚度。细胞毒试验结果表明,316L不锈钢种植体原型、国产种植体和进口种植体的活细胞率分别为83.7%、87.5%和83.7%。单因素方差分析(One - Way ANOVA)显示,3组间差异无统计学意义(p>0.05)。3组患者的刺激试验结果均为0级(无水肿、无红斑)。急性毒性/热原性试验各组间差异无统计学意义(p>0.05)。植入后1周和4周纤维囊环厚度差异无统计学意义(p>0.05)。根据ISO医学试验指南,不锈钢316L种植体原型是安全的,与本地种植体和进口种植体相比,生物相容性测试结果相似。关键词:316L不锈钢,生物相容性,骨科植入物,链接:ferdiansyah@fk.unair.ac.id摘要:印度尼西亚kebutuhan植入物,namun产品,sinya terbatas, sebagian besar, masih,进口,进口。不锈钢,utama 316L adalah salah satu金属阳服务digunakan。Penelitian ini mengevaluasi biokompatibilitas dilian不锈钢316L原型,impan本地yang diproduksi menggunakan不锈钢进口,dan impan不锈钢进口。Uji sitotoksisitas menggunakan sel mesenkim人登革热MTT测定。黄芪,黄芪,黄芪,黄芪,黄芪,黄芪,黄芪,黄芪,黄芪。Kami mengevaluasi skor Draize, perubahan berat dan温度rektal, dan ketebalan纤维囊环。Uji sitotoksisitas menunjukkan代表的可行性是卖帕原型不锈钢316L, implan local,但implan import berut -turut adalah 83,7%;87年,5%;丹83年7%。单因素方差分析(One - Way Anova)与3组比较,差异有统计学意义(p> 0.05)。Hasil uji iritasi 3组menunjukkan 0级(水肿和水肿)。前列腺癌组与前列腺癌组比较差异有统计学意义(p> 0.05)。有些ada perbedaan signifikan (p > 0 05)普拉篇ketebalan纤维囊环di minggu柯1丹4 setelah implantasi。Berdasarkan ISO untuk pedoman uji di bidang介质,原型不锈钢316L mandan memiliki kesamaan hasil uji biokompatiilias bila dibandingkan dengan implan本地丹进口。Kata kunci:不锈钢316L, uji biokompatibilitas, implan orthopaediKorespondensi: ferdiansyah@fk.unair.ac.id