Dosimetric Characteristics of 6 MV Medical Linac at BAEC

Sudeb Roy, P. K. Das, R. Khatun, M. A. Rahman, S. Akter, Tushar Kumar, M. M. Ahasan
{"title":"Dosimetric Characteristics of 6 MV Medical Linac at BAEC","authors":"Sudeb Roy, P. K. Das, R. Khatun, M. A. Rahman, S. Akter, Tushar Kumar, M. M. Ahasan","doi":"10.4236/IJMPCERO.2021.101004","DOIUrl":null,"url":null,"abstract":"Dosimetric characteristic is one of the essential parameters of a medical linear accelerator (LINAC), which must be obtained before clinical use. The dosimetric characteristics for 6 MV photon beam were measured and compared with the corresponding published data. The study was done using a Varian linear accelerator (Model Clinac-iX) at the Institute of Nuclear Medical Physics (INMP), AERE, Savar, Dhaka, under the Bangladesh Atomic Energy Commission (BAEC). The data is taken for 10 field sizes (2 × 2, 3 × 3, 5 × 5, 7 × 7, 10 × 10, 15 × 15, 20 × 20, 25 × 25, 30 × 30 and 40 × 40 cm2) at same conditions. The measured Percent Depth Dose (PDD) curves were obtained for 6 MV photon beams with the field as mentioned above and compared with the calculated PDD curves. The measured depth dose (Dmax) for reference field size (FS) 10 × 10 cm2 is 15.99 mm, and the PDD at 10 cm depth (D10) is 66.87% for 6 MV photon energies that are found to be compatible with the published report BJR supplement 25. The measured PDD curves for photon energies show a good agreement with the standard PDD curves. The photon beam dosimetry data found in the current study are compatible and all the tolerances are within the clinically acceptable tolerance limit.","PeriodicalId":14028,"journal":{"name":"International Journal of Medical Physics, Clinical Engineering and Radiation Oncology","volume":"5 1","pages":"38-46"},"PeriodicalIF":0.0000,"publicationDate":"2021-01-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Medical Physics, Clinical Engineering and Radiation Oncology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4236/IJMPCERO.2021.101004","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Dosimetric characteristic is one of the essential parameters of a medical linear accelerator (LINAC), which must be obtained before clinical use. The dosimetric characteristics for 6 MV photon beam were measured and compared with the corresponding published data. The study was done using a Varian linear accelerator (Model Clinac-iX) at the Institute of Nuclear Medical Physics (INMP), AERE, Savar, Dhaka, under the Bangladesh Atomic Energy Commission (BAEC). The data is taken for 10 field sizes (2 × 2, 3 × 3, 5 × 5, 7 × 7, 10 × 10, 15 × 15, 20 × 20, 25 × 25, 30 × 30 and 40 × 40 cm2) at same conditions. The measured Percent Depth Dose (PDD) curves were obtained for 6 MV photon beams with the field as mentioned above and compared with the calculated PDD curves. The measured depth dose (Dmax) for reference field size (FS) 10 × 10 cm2 is 15.99 mm, and the PDD at 10 cm depth (D10) is 66.87% for 6 MV photon energies that are found to be compatible with the published report BJR supplement 25. The measured PDD curves for photon energies show a good agreement with the standard PDD curves. The photon beam dosimetry data found in the current study are compatible and all the tolerances are within the clinically acceptable tolerance limit.
6 MV医用直线剂在BAEC的剂量学特性
剂量学特性是医用直线加速器的重要参数之一,在临床使用前必须掌握剂量学特性。测量了6 MV光子束的剂量学特性,并与已发表的相关数据进行了比较。这项研究是在孟加拉国原子能委员会(BAEC)下属的达卡Savar AERE核医学物理研究所(INMP)使用瓦里安直线加速器(Clinac-iX模型)完成的。在相同条件下,选取了10个面积(2 × 2、3 × 3、5 × 5、7 × 7、10 × 10、15 × 15、20 × 20、25 × 25、30 × 30和40 × 40 cm2)的数据。在上述场作用下,得到了6 MV光子束流的实测百分比深度剂量曲线,并与计算的百分比深度剂量曲线进行了比较。参考场尺寸(FS)为10 × 10 cm2时的测量深度剂量(Dmax)为15.99 mm, 6 MV光子能量在10 cm深度处的PDD (D10)为66.87%,与已发表的报告BJR supplement 25一致。实测的光子能量PDD曲线与标准PDD曲线吻合较好。本研究中发现的光子束剂量学数据是兼容的,所有的耐受性都在临床可接受的耐受性范围内。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信