Radiobiological assessment of dose-to-medium or dose-to-water with Acuros XB algorithm compared with Anisotropic Analytical Algorithm for lung cancer radiotherapy- What should we know to manage the transition?

A. Chaikh, C. Khamphan, A. Delbaere, J. Ojala, Robin García, J. Balosso
{"title":"Radiobiological assessment of dose-to-medium or dose-to-water with Acuros XB algorithm compared with Anisotropic Analytical Algorithm for lung cancer radiotherapy- What should we know to manage the transition?","authors":"A. Chaikh, C. Khamphan, A. Delbaere, J. Ojala, Robin García, J. Balosso","doi":"10.14319/IJCTO.51.18","DOIUrl":null,"url":null,"abstract":"Purpose: To track the dosimetric changes for similar dose prescriptions, when dose calculation algorithms are upgraded in the treatment planning system (TPS). Clinically significant representations of the treatment outcomes are used to provide interpretable data for radiation oncologists, as the equivalent uniform dose (EUD), the tumor control probability (TCP), the late toxicity as normal tissue complication probability (NTCP) and the uncomplicated tumor control probability (UTCP) scores. Results are presented and discussed in a clinical perspective. Methods: Ten lung cancer patients were included in this study. For each patient, five treatment plans were generated. The doses were calculated using Anisotropic Analytical Algorithm (AAA) and both Acuros XB (AXB) dose reporting modes: dose-to-medium AXB D(m,m) and dose-to-water AXB D(w,m). In plans 1, 2 and 3, the doses were calculated respectively with AAA, AXB D(m,m) and AXB D(w,m) using exactly the same prescription dose and beam set-up. The doses in plans 4 and 5 were calculated using both AXB dose reporting modes using, as input, the same number of monitor units (MUs) as yielded by AAA, with the same beam set-up. The EUD, TCP and NTCP were computed using the assumed radiobiological parameters from literature. The Wilcoxon paired test was used to calculate p-values. Results: Using the same prescription dose, TCP values were higher with AXB than with AAA, and corresponding UTCP scores were 1-2% better with p < 0.05. In addition, absolute NTCP values were slightly increased with AXB. Both AXB dose reporting modes yielded comparable lower TCP and NTCP values (again in the order of 1-2%) than with AAA, when using same MU numbers as with AAA. Conclusion: Compared to AAA, taken as reference, both AXB dose reporting modes yielded better results. AAA showed very close values to AXB D(w,m), but it is difficult to give recommendation between D(w,m) and D(m,m) yet, due to the lack of recommended radiobiological parameters associated with these dose reporting modes. We suggest doing experimental and modelling studies to determine the real radiobiological effects in both targets and organs at risks. Should the differences be substantial in some conditions and relevant to clinical practice, discussions regarding dose prescription and optimization of the tolerance doses to OAR should be undertaken between medical physicists and radiation oncologists.","PeriodicalId":13931,"journal":{"name":"International Journal of Cancer Therapy and Oncology","volume":"103 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2017-12-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Cancer Therapy and Oncology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.14319/IJCTO.51.18","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

Purpose: To track the dosimetric changes for similar dose prescriptions, when dose calculation algorithms are upgraded in the treatment planning system (TPS). Clinically significant representations of the treatment outcomes are used to provide interpretable data for radiation oncologists, as the equivalent uniform dose (EUD), the tumor control probability (TCP), the late toxicity as normal tissue complication probability (NTCP) and the uncomplicated tumor control probability (UTCP) scores. Results are presented and discussed in a clinical perspective. Methods: Ten lung cancer patients were included in this study. For each patient, five treatment plans were generated. The doses were calculated using Anisotropic Analytical Algorithm (AAA) and both Acuros XB (AXB) dose reporting modes: dose-to-medium AXB D(m,m) and dose-to-water AXB D(w,m). In plans 1, 2 and 3, the doses were calculated respectively with AAA, AXB D(m,m) and AXB D(w,m) using exactly the same prescription dose and beam set-up. The doses in plans 4 and 5 were calculated using both AXB dose reporting modes using, as input, the same number of monitor units (MUs) as yielded by AAA, with the same beam set-up. The EUD, TCP and NTCP were computed using the assumed radiobiological parameters from literature. The Wilcoxon paired test was used to calculate p-values. Results: Using the same prescription dose, TCP values were higher with AXB than with AAA, and corresponding UTCP scores were 1-2% better with p < 0.05. In addition, absolute NTCP values were slightly increased with AXB. Both AXB dose reporting modes yielded comparable lower TCP and NTCP values (again in the order of 1-2%) than with AAA, when using same MU numbers as with AAA. Conclusion: Compared to AAA, taken as reference, both AXB dose reporting modes yielded better results. AAA showed very close values to AXB D(w,m), but it is difficult to give recommendation between D(w,m) and D(m,m) yet, due to the lack of recommended radiobiological parameters associated with these dose reporting modes. We suggest doing experimental and modelling studies to determine the real radiobiological effects in both targets and organs at risks. Should the differences be substantial in some conditions and relevant to clinical practice, discussions regarding dose prescription and optimization of the tolerance doses to OAR should be undertaken between medical physicists and radiation oncologists.
用acros XB算法与各向异性分析算法比较剂量对介质或剂量对水的放射生物学评估肺癌放疗-我们应该知道什么来管理过渡?
目的:跟踪治疗计划系统(TPS)中剂量计算算法升级后类似剂量处方的剂量学变化。治疗结果的临床意义表征用于为放射肿瘤学家提供可解释的数据,如等效均匀剂量(EUD)、肿瘤控制概率(TCP)、晚期毒性作为正常组织并发症概率(NTCP)和无并发症肿瘤控制概率(UTCP)评分。结果提出并讨论在临床角度。方法:选取10例肺癌患者作为研究对象。针对每位患者,生成了5个治疗方案。采用各向异性分析算法(AAA)和acros XB (AXB)两种剂量报告模式计算剂量:剂量对介质AXB D(m,m)和剂量对水AXB D(w,m)。在方案1、方案2和方案3中,使用完全相同的处方剂量和光束设置,分别用AAA、AXB D(m,m)和AXB D(w,m)计算剂量。方案4和方案5中的剂量是使用两种AXB剂量报告模式计算的,使用与AAA产生的相同数量的监测单位(MUs)作为输入,具有相同的光束设置。EUD、TCP和NTCP使用文献中假设的放射生物学参数计算。采用Wilcoxon配对检验计算p值。结果:在处方剂量相同的情况下,AXB组的TCP值高于AAA组,其UTCP评分高1 ~ 2%,差异有统计学意义(p < 0.05)。此外,绝对NTCP值在AXB的作用下略有增加。当使用与AAA相同的MU数时,两种AXB剂量报告模式的TCP和NTCP值均比AAA低(同样为1-2%)。结论:以AAA为参照,两种AXB剂量报告模式的结果均优于AAA。AAA值与AXB D(w,m)非常接近,但由于缺乏与这些剂量报告模式相关的推荐放射生物学参数,因此很难在D(w,m)和D(m,m)之间给出推荐值。我们建议进行实验和建模研究,以确定目标和危险器官的真实放射生物学效应。如果在某些情况下差异很大并且与临床实践有关,则应在医学物理学家和放射肿瘤学家之间就剂量处方和OAR耐受剂量的优化进行讨论。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书