Effect of Age Related Breathing Parameters on the Performance of a New Nebulizer System: An In-Vitro Study

M. Knoch, E. Wunderlich
{"title":"Effect of Age Related Breathing Parameters on the Performance of a New Nebulizer System: An In-Vitro Study","authors":"M. Knoch, E. Wunderlich","doi":"10.1089/JAM.1995.8.285","DOIUrl":null,"url":null,"abstract":"ABSTRACT A new continuously operated nebulizer system (PARI LC) has been designed for use in patients from one month of age to adulthood. Two different configurations are available, one with a son silicon face mask for use in babies and infants (three different mask sizes), and one with a mouthpiece with integrated expiratory valve. The nebulizer is based on an open-vent technology in combination with a valve system which enables the entrained air to pass through the nebulizing chamber, thus increasing aerosol output during inspiration. In this study the effect of tidal volume and breathing frequency on the nebulizer efficiency has been investigated by using a breath simulator at fixed 50% inspiration and 50% expiration cycles. Sodium fluoride was used as a tracer and analysed from filters simulating the aerosol to be inhaled by a patient (aerosol delivered) and the amount exhausted to the environment (aerosol wasted). Each configuration of the nebulizer was tested by simulating the relevant breathing con...","PeriodicalId":14879,"journal":{"name":"Journal of Aerosol Medicine-deposition Clearance and Effects in The Lung","volume":"139 1","pages":"285-288"},"PeriodicalIF":0.0000,"publicationDate":"1995-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"8","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Aerosol Medicine-deposition Clearance and Effects in The Lung","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1089/JAM.1995.8.285","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 8

Abstract

ABSTRACT A new continuously operated nebulizer system (PARI LC) has been designed for use in patients from one month of age to adulthood. Two different configurations are available, one with a son silicon face mask for use in babies and infants (three different mask sizes), and one with a mouthpiece with integrated expiratory valve. The nebulizer is based on an open-vent technology in combination with a valve system which enables the entrained air to pass through the nebulizing chamber, thus increasing aerosol output during inspiration. In this study the effect of tidal volume and breathing frequency on the nebulizer efficiency has been investigated by using a breath simulator at fixed 50% inspiration and 50% expiration cycles. Sodium fluoride was used as a tracer and analysed from filters simulating the aerosol to be inhaled by a patient (aerosol delivered) and the amount exhausted to the environment (aerosol wasted). Each configuration of the nebulizer was tested by simulating the relevant breathing con...
年龄相关呼吸参数对新型雾化器系统性能的影响:体外研究
一种新的连续操作雾化器系统(PARI LC)已经设计用于从一个月大到成年的患者。有两种不同的配置可供选择,一种是用于婴儿和婴儿的硅胶面罩(三种不同的面罩尺寸),另一种是带有集成呼气阀的吸口。该喷雾器基于开放式排气技术,并结合阀门系统,使夹带的空气能够通过雾化室,从而增加吸入过程中的气溶胶输出。在固定的50%吸气和50%呼气周期下,利用呼吸模拟器研究了潮气量和呼吸频率对雾化器效率的影响。使用氟化钠作为示踪剂,并从模拟患者吸入的气雾剂(输送的气雾剂)和排放到环境中的量(浪费的气雾剂)的过滤器进行分析。通过模拟相关的呼吸条件,对雾化器的每种配置进行了测试。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信