{"title":"The NBS Two-Pressure Humidity Generator, Mark 2","authors":"S. Hasegawa, J. W. Little","doi":"10.6028/jres.081A.010","DOIUrl":null,"url":null,"abstract":"A new humidity calibration facility which uses the two-pressure principle for generating gas of known humidity has been developed at NBS for calibrating and testing hygrometers. The relative humidity range of the two-pressure humidity generator is 3 to 98 percent for ambient temperatures −60 ° to 80 °C and test chamber pressures 5 to 200 kPa (absolute). This is equivalent to a nominal dew/frost point range of −80 ° to 80 °C. Intercomparison tests were made with the NBS standard gravimetric hygrometer over a portion of the generator’s operating range. The estimated maximum uncertainty (three standard deviations) is 0.2 percent RH for temperatures 0 ° to 80 °C which in units of dew point corresponds to an estimated maximum uncertainty of 0.04 °C for dew points −35 ° to 80 °C.","PeriodicalId":94340,"journal":{"name":"Journal of research of the National Bureau of Standards. Section A, Physics and chemistry","volume":"22 1","pages":"81 - 88"},"PeriodicalIF":0.0000,"publicationDate":"1977-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"51","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of research of the National Bureau of Standards. Section A, Physics and chemistry","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6028/jres.081A.010","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 51
Abstract
A new humidity calibration facility which uses the two-pressure principle for generating gas of known humidity has been developed at NBS for calibrating and testing hygrometers. The relative humidity range of the two-pressure humidity generator is 3 to 98 percent for ambient temperatures −60 ° to 80 °C and test chamber pressures 5 to 200 kPa (absolute). This is equivalent to a nominal dew/frost point range of −80 ° to 80 °C. Intercomparison tests were made with the NBS standard gravimetric hygrometer over a portion of the generator’s operating range. The estimated maximum uncertainty (three standard deviations) is 0.2 percent RH for temperatures 0 ° to 80 °C which in units of dew point corresponds to an estimated maximum uncertainty of 0.04 °C for dew points −35 ° to 80 °C.