The NBS Two-Pressure Humidity Generator, Mark 2

S. Hasegawa, J. W. Little
{"title":"The NBS Two-Pressure Humidity Generator, Mark 2","authors":"S. Hasegawa, J. W. Little","doi":"10.6028/jres.081A.010","DOIUrl":null,"url":null,"abstract":"A new humidity calibration facility which uses the two-pressure principle for generating gas of known humidity has been developed at NBS for calibrating and testing hygrometers. The relative humidity range of the two-pressure humidity generator is 3 to 98 percent for ambient temperatures −60 ° to 80 °C and test chamber pressures 5 to 200 kPa (absolute). This is equivalent to a nominal dew/frost point range of −80 ° to 80 °C. Intercomparison tests were made with the NBS standard gravimetric hygrometer over a portion of the generator’s operating range. The estimated maximum uncertainty (three standard deviations) is 0.2 percent RH for temperatures 0 ° to 80 °C which in units of dew point corresponds to an estimated maximum uncertainty of 0.04 °C for dew points −35 ° to 80 °C.","PeriodicalId":94340,"journal":{"name":"Journal of research of the National Bureau of Standards. Section A, Physics and chemistry","volume":"22 1","pages":"81 - 88"},"PeriodicalIF":0.0000,"publicationDate":"1977-02-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"51","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of research of the National Bureau of Standards. Section A, Physics and chemistry","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6028/jres.081A.010","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 51

Abstract

A new humidity calibration facility which uses the two-pressure principle for generating gas of known humidity has been developed at NBS for calibrating and testing hygrometers. The relative humidity range of the two-pressure humidity generator is 3 to 98 percent for ambient temperatures −60 ° to 80 °C and test chamber pressures 5 to 200 kPa (absolute). This is equivalent to a nominal dew/frost point range of −80 ° to 80 °C. Intercomparison tests were made with the NBS standard gravimetric hygrometer over a portion of the generator’s operating range. The estimated maximum uncertainty (three standard deviations) is 0.2 percent RH for temperatures 0 ° to 80 °C which in units of dew point corresponds to an estimated maximum uncertainty of 0.04 °C for dew points −35 ° to 80 °C.
NBS双压湿度发生器,Mark 2
国家测量局已开发了一套新的湿度校正设备,该设备使用双压力原理产生已知湿度的气体,用于校准和测试湿度计。当环境温度为- 60°C至80°C,测试室压力为5至200kpa(绝对)时,双压湿度发生器的相对湿度范围为3%至98%。这相当于名义上的露点/霜点范围为- 80°到80°C。在发电机的一部分工作范围内,与NBS标准重力湿度计进行了比对试验。对于温度0°至80°C,估计最大不确定度(三个标准差)为0.2% RH,以露点为单位,对应于露点- 35°至80°C的估计最大不确定度为0.04°C。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信