Clinical Commissioning and Dosimetric Verification of the Raystation Treatment Planning System

T. Tuğrul
{"title":"Clinical Commissioning and Dosimetric Verification of the Raystation Treatment Planning System","authors":"T. Tuğrul","doi":"10.26565/2312-4334-2021-4-13","DOIUrl":null,"url":null,"abstract":"Background: The software used by treatment planning systems (TPS) plays an important role for treatments using radiation. The accuracy of the calculated dose in radiation treatments depends on the assumptions made by the TPS. In this study, we summarize our methods and results regarding clinical commissioning of the basic functions needed for photon therapy. Materials and Method: Measurements were obtained for the 6 and 15 MV photon energies obtained from the Siemens Artiste linear accelerator device. Important data such as percent deep dose, profile and output measurements were taken in the water phantom and transferred to the RayStation Treatment Planning System. Results: When the absolute dose values calculated by the RayStation TPS are compared with the water phantom data, the differences obtained are less than 3%. When the 2-dimensional quality control of asymmetrical areas and patients with IMRT plan was controlled by gamma analysis method, the gamma rate was more than 95%. Conclusion: One of the most important quality control tests is TPS acceptance tests, which must be performed before clinical use. In this study, in which we checked the basic dose measurement and patient planning, it was seen that the RayStation TPS can be used in patient treatment for clinical use. The doses calculated by the RayStation TPS were found to be reliable and within the expected accuracy range. These results are sufficient for the application of 3-dimensional conformal radiotherapy (3D-CRT) and IMRT technique.","PeriodicalId":76806,"journal":{"name":"Vital and health statistics. Ser. 4, Documents and committee reports","volume":"32 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2021-12-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Vital and health statistics. Ser. 4, Documents and committee reports","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.26565/2312-4334-2021-4-13","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background: The software used by treatment planning systems (TPS) plays an important role for treatments using radiation. The accuracy of the calculated dose in radiation treatments depends on the assumptions made by the TPS. In this study, we summarize our methods and results regarding clinical commissioning of the basic functions needed for photon therapy. Materials and Method: Measurements were obtained for the 6 and 15 MV photon energies obtained from the Siemens Artiste linear accelerator device. Important data such as percent deep dose, profile and output measurements were taken in the water phantom and transferred to the RayStation Treatment Planning System. Results: When the absolute dose values calculated by the RayStation TPS are compared with the water phantom data, the differences obtained are less than 3%. When the 2-dimensional quality control of asymmetrical areas and patients with IMRT plan was controlled by gamma analysis method, the gamma rate was more than 95%. Conclusion: One of the most important quality control tests is TPS acceptance tests, which must be performed before clinical use. In this study, in which we checked the basic dose measurement and patient planning, it was seen that the RayStation TPS can be used in patient treatment for clinical use. The doses calculated by the RayStation TPS were found to be reliable and within the expected accuracy range. These results are sufficient for the application of 3-dimensional conformal radiotherapy (3D-CRT) and IMRT technique.
放射站治疗计划系统的临床调试和剂量学验证
背景:放射治疗计划系统(TPS)软件在放射治疗中起着重要作用。放射治疗中计算剂量的准确性取决于TPS所作的假设。在本研究中,我们总结了关于光子治疗所需基本功能的临床调试的方法和结果。材料和方法:测量了西门子Artiste直线加速器装置获得的6 MV和15 MV光子能量。重要数据,如深度剂量百分比、剖面和输出测量值,在水模体中进行,并传输到RayStation治疗计划系统。结果:将RayStation TPS计算的绝对剂量值与水影数据进行比较,差异小于3%。当采用伽玛分析法控制非对称区域和IMRT计划患者的二维质量控制时,伽玛率均大于95%。结论:TPS验收是最重要的质量控制检查之一,必须在临床使用前进行。在本研究中,我们检查了基本剂量测量和患者计划,发现RayStation TPS可以用于患者治疗,可供临床使用。RayStation TPS计算的剂量是可靠的,在预期的精度范围内。这些结果足以为三维适形放疗(3D-CRT)和IMRT技术的应用提供依据。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书