The Effects of Defect Type and Depth, and Measurement Direction on the Implant Stability Quotient Value.

Seung-Yun Shin, S. Shin, Seung-Beom Kye, Jongrak Hong, J. Paeng, Seok-Woo Chang, Seung-Min Yang
{"title":"The Effects of Defect Type and Depth, and Measurement Direction on the Implant Stability Quotient Value.","authors":"Seung-Yun Shin, S. Shin, Seung-Beom Kye, Jongrak Hong, J. Paeng, Seok-Woo Chang, Seung-Min Yang","doi":"10.1563/AAID-JOI-D-13-0031","DOIUrl":null,"url":null,"abstract":"The purpose of this study was to evaluate the effect of defect type and depth as well as measurement direction on implant stability in an ex vivo peri-implant bovine rib bone model. Six kinds of defects (3-wall 2.5 mm, 3-wall 5 mm, 1-wall 2.5 mm, 1-wall 5 mm, circumferential 2.5 mm, circumferential 5 mm), and control (no defect) were prepared in 14 bovine rib bones. A total of 84 defects and 14 controls were created. The same type and size of implants (4 × 10 mm) were placed in each group. The thickness of cortical bone and the insertion torque were measured for each defect, and the implant stability quotient (ISQ) value was measured 3 times from 4 different directions. The thickness of cortical bone ranged from 2.71-3.18 mm. Insertion torque decreased as the defect size increased. As the defect size for the same defect depth increased, the ISQ value decreased (P < .001). There were significant differences between the ISQ values obtained with different measurement directions only between the control and 3-wall 5 mm defect (P < .0001). The ISQ value opposite to the defect direction was higher than that in the defect direction in all 3 directions of the 3-wall and 1-wall 5 mm defects. ISQ values were influenced by defect type and depth. Loss of cortical bone reduced the stability of implants and reduced the ISQ value. Measurement direction also influenced ISQ values.","PeriodicalId":77634,"journal":{"name":"The International journal of oral implantology : implantologist","volume":"1 1","pages":"652-6"},"PeriodicalIF":0.0000,"publicationDate":"2015-12-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"12","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"The International journal of oral implantology : implantologist","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1563/AAID-JOI-D-13-0031","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 12

Abstract

The purpose of this study was to evaluate the effect of defect type and depth as well as measurement direction on implant stability in an ex vivo peri-implant bovine rib bone model. Six kinds of defects (3-wall 2.5 mm, 3-wall 5 mm, 1-wall 2.5 mm, 1-wall 5 mm, circumferential 2.5 mm, circumferential 5 mm), and control (no defect) were prepared in 14 bovine rib bones. A total of 84 defects and 14 controls were created. The same type and size of implants (4 × 10 mm) were placed in each group. The thickness of cortical bone and the insertion torque were measured for each defect, and the implant stability quotient (ISQ) value was measured 3 times from 4 different directions. The thickness of cortical bone ranged from 2.71-3.18 mm. Insertion torque decreased as the defect size increased. As the defect size for the same defect depth increased, the ISQ value decreased (P < .001). There were significant differences between the ISQ values obtained with different measurement directions only between the control and 3-wall 5 mm defect (P < .0001). The ISQ value opposite to the defect direction was higher than that in the defect direction in all 3 directions of the 3-wall and 1-wall 5 mm defects. ISQ values were influenced by defect type and depth. Loss of cortical bone reduced the stability of implants and reduced the ISQ value. Measurement direction also influenced ISQ values.
缺陷类型、深度及测量方向对种植体稳定商值的影响。
本研究旨在探讨缺损类型、缺损深度及测量方向对牛肋骨离体模型种植体稳定性的影响。在14块牛肋骨上制备6种缺损(3- 2.5 mm、3- 5mm、1- 2.5 mm、1- 5mm、圆周2.5 mm、圆周5mm)和对照组(无缺损)。总共创建了84个缺陷和14个控制。各组植入同种类型、同种尺寸的种植体(4 × 10 mm)。测量每个缺损的皮质骨厚度和插入扭矩,并从4个不同方向测量种植体稳定商(ISQ)值3次。皮质骨厚度为2.71 ~ 3.18 mm。插入力矩随着缺陷尺寸的增大而减小。随着相同缺陷深度的缺陷尺寸的增加,ISQ值降低(P < 0.001)。不同测量方向下的ISQ值仅在对照组与3壁5 mm缺陷间存在显著差异(P < 0.0001)。在3壁和1壁5mm缺陷的3个方向上,缺陷方向相反的ISQ值均高于缺陷方向。ISQ值受缺陷类型和深度的影响。皮质骨的丢失降低了种植体的稳定性,降低了ISQ值。测量方向对ISQ值也有影响。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书