Method to quantify the decentration of the multifocal intraocular lenses in relation to the visual axis

F. Polit
{"title":"Method to quantify the decentration of the multifocal intraocular lenses in relation to the visual axis","authors":"F. Polit","doi":"10.15234/VPA.V17I4.513","DOIUrl":null,"url":null,"abstract":"Since position and centration have shown to affect the effectiveness of multifocal intraocular lenses (MFIOLs), some authors have proposed using the \"coaxially sighted IOL light reflex\" as a landmark to allow consistent centration of MFIOLs. Postoperatively, assessing the location of the intraocular lens with respect to the visual axis is also important. Centration of the MFIOLs can be identified postoperatively by slit lamp biomicroscopy, but it is still a qualitative method. The Nidek OPD Scan III is used. The instrument obtains retro illumination images, which allows the observation of the diffractive rings of the multifocal intraocular lenses implanted following phacoemulsification. The retro illumination image describes the location of the visual axis (white cross with blue borders +), the center of the photopic (pink cross) and mesopic (turquoise cross) pupil, quantifies the distance between the center of the pupil and the visual axis, and places the visual axis in degrees. It is attempted to match the rings of the lens with that of the acetate grid. Once the best possible equidistance between the lens and grid rings has been achieved. This method can be useful to be able to have reference of a real value of decentration of the lenses and thus have a surgical plan and try to center the lenses. We are also creating a technique to properly center the lenses, in addition to performing an analysis of the results in terms of postoperative visual quality in relation to the lens's decentration and thus see how this really affects.","PeriodicalId":53032,"journal":{"name":"Vision PanAmerica","volume":"58 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2018-12-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Vision PanAmerica","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.15234/VPA.V17I4.513","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Since position and centration have shown to affect the effectiveness of multifocal intraocular lenses (MFIOLs), some authors have proposed using the "coaxially sighted IOL light reflex" as a landmark to allow consistent centration of MFIOLs. Postoperatively, assessing the location of the intraocular lens with respect to the visual axis is also important. Centration of the MFIOLs can be identified postoperatively by slit lamp biomicroscopy, but it is still a qualitative method. The Nidek OPD Scan III is used. The instrument obtains retro illumination images, which allows the observation of the diffractive rings of the multifocal intraocular lenses implanted following phacoemulsification. The retro illumination image describes the location of the visual axis (white cross with blue borders +), the center of the photopic (pink cross) and mesopic (turquoise cross) pupil, quantifies the distance between the center of the pupil and the visual axis, and places the visual axis in degrees. It is attempted to match the rings of the lens with that of the acetate grid. Once the best possible equidistance between the lens and grid rings has been achieved. This method can be useful to be able to have reference of a real value of decentration of the lenses and thus have a surgical plan and try to center the lenses. We are also creating a technique to properly center the lenses, in addition to performing an analysis of the results in terms of postoperative visual quality in relation to the lens's decentration and thus see how this really affects.
方法量化多焦人工晶状体相对于视轴的离体
由于位置和集中度会影响多焦人工晶状体(MFIOLs)的有效性,一些作者建议使用“同轴视人工晶状体光反射”作为一个标志,以使MFIOLs的集中度保持一致。术后,评估人工晶状体相对于视轴的位置也很重要。术后可通过裂隙灯生物显微镜检测MFIOLs的浓度,但仍是定性方法。使用Nidek OPD Scan III。该仪器获得复古照明图像,允许观察超声乳化术后植入的多焦人工晶状体的衍射环。复古照明图像描述了视轴(白色十字带蓝色边框+)、感光瞳孔中心(粉色十字)和介观瞳孔中心(绿松石十字)的位置,量化了瞳孔中心与视轴之间的距离,并将视轴以度表示。它试图将镜头的环与醋酸网格的环相匹配。一旦镜头和栅格环之间达到了最佳的等距。这种方法可以为晶状体离心的实际值提供参考,从而制定手术计划并尝试使晶状体居中。我们也在创造一种技术来正确地将晶状体居中,除了对术后视觉质量与晶状体分散的关系进行分析,从而了解这是如何影响的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
审稿时长
4 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信