M. Popovic, J. Beathe, E. Gbaje, Marla Sharp, S. Memtsoudis
{"title":"Effect of portable negative pressure units on expelled aerosols in the operating room environment","authors":"M. Popovic, J. Beathe, E. Gbaje, Marla Sharp, S. Memtsoudis","doi":"10.1136/rapm-2022-103489","DOIUrl":null,"url":null,"abstract":"Introduction Spontaneously breathing patients undergoing procedures under regional anesthesia can expose operating room personnel to infectious agents. The use of localized negative pressure within proximity of a patient’s airway is expected to reduce the amount of bioaerosols dispersed particularly for anesthesia staff who are frequently near the patient’s airway. Methods In the experiment, aerosols were produced using a polydisperse aerosol generator with nebulized saline. A portable negative pressure unit was set up at set distances of 10 cm and 30 cm with the aim of reducing aerosol particle counts detected by a laser-based particle counter. Results Without the portable negative pressure unit, the median concentration of 0.5 µm aerosols detected was 3128 (1533, 22832) particles/ft3/min. With the portable negative pressure unit 10 cm and 30 cm from the site of aerosol emittance, the median concentration compared with background concentration was −0.5 (−8, 8) particles/ft3/min and 398 (89, 1749) particles/ft3/min, respectively. Conclusions For particle concentrations of 0.5 µm, 0.7 µm, and 1.0 µm a significant amount of aerosol reduction was observed (p<0.001). Further experiments are warranted to assess the safety of staff when encountering a potentially infectious patient in the operating room.","PeriodicalId":21046,"journal":{"name":"Regional Anesthesia & Pain Medicine","volume":"8 1","pages":"426 - 429"},"PeriodicalIF":0.0000,"publicationDate":"2022-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"7","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Regional Anesthesia & Pain Medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1136/rapm-2022-103489","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 7
Abstract
Introduction Spontaneously breathing patients undergoing procedures under regional anesthesia can expose operating room personnel to infectious agents. The use of localized negative pressure within proximity of a patient’s airway is expected to reduce the amount of bioaerosols dispersed particularly for anesthesia staff who are frequently near the patient’s airway. Methods In the experiment, aerosols were produced using a polydisperse aerosol generator with nebulized saline. A portable negative pressure unit was set up at set distances of 10 cm and 30 cm with the aim of reducing aerosol particle counts detected by a laser-based particle counter. Results Without the portable negative pressure unit, the median concentration of 0.5 µm aerosols detected was 3128 (1533, 22832) particles/ft3/min. With the portable negative pressure unit 10 cm and 30 cm from the site of aerosol emittance, the median concentration compared with background concentration was −0.5 (−8, 8) particles/ft3/min and 398 (89, 1749) particles/ft3/min, respectively. Conclusions For particle concentrations of 0.5 µm, 0.7 µm, and 1.0 µm a significant amount of aerosol reduction was observed (p<0.001). Further experiments are warranted to assess the safety of staff when encountering a potentially infectious patient in the operating room.