Measurement and Comparison of Output Factors Using Two Detectors for NOVAC7 IntraOperative Radiotherapy Accelerator

Rafiqul Islam, H. Watabe, A. Stefano
{"title":"Measurement and Comparison of Output Factors Using Two Detectors for NOVAC7 IntraOperative Radiotherapy Accelerator","authors":"Rafiqul Islam, H. Watabe, A. Stefano","doi":"10.4236/ijmpcero.2020.92006","DOIUrl":null,"url":null,"abstract":"The aim of this work was to evaluate and compare the performance of comparatively new synthetic PTW 60019 microDiamond with PTW 60017 Diode E detector in measuring the output factors (OF) of IntraOperative Radiation Therapy (IORT) electron beams. For a given electron beam, OFs are defined as the ratio of the dose for any applicator size at the depth of maximum to that for a reference applicator. IORT is an innovative treatment technique that delivers a large single fraction of radiation dose to the tumor bed during surgery. The electron beams considered in this study were generated by the mobile NOVAC7 system. This device produces high-dose-per-pulse electron beams with four different energies in the range from 3 MeV to 9 MeV. We performed measurements for two higher energies, namely 7MeV and 9 MeV. The beam collimation was performed through Perspex (PMMA) cylindrical applicators with different diameters. The accurate dose delivery of IORT tightly depends on the precision of measured dose by reference applicator and the output factors of clinical applicators. The output factors were measured using microDiamond and Diode E detectors. The microDiamond detector performance was compared with a Diode E detector. Determined output factors of two detectors were in good agreement. The maximum deviations of output factors for microDiamond were found 2.74%, and 2.17% for 7 MeV and 9 MeV, respectively with respect to the PTW Diode E. The microDiamond detector was shown to exhibit excellent properties for output factor measurements and could be considered as a suitable tool for electron beam dosimetry.","PeriodicalId":14028,"journal":{"name":"International Journal of Medical Physics, Clinical Engineering and Radiation Oncology","volume":"76 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2020-03-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Medical Physics, Clinical Engineering and Radiation Oncology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4236/ijmpcero.2020.92006","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

The aim of this work was to evaluate and compare the performance of comparatively new synthetic PTW 60019 microDiamond with PTW 60017 Diode E detector in measuring the output factors (OF) of IntraOperative Radiation Therapy (IORT) electron beams. For a given electron beam, OFs are defined as the ratio of the dose for any applicator size at the depth of maximum to that for a reference applicator. IORT is an innovative treatment technique that delivers a large single fraction of radiation dose to the tumor bed during surgery. The electron beams considered in this study were generated by the mobile NOVAC7 system. This device produces high-dose-per-pulse electron beams with four different energies in the range from 3 MeV to 9 MeV. We performed measurements for two higher energies, namely 7MeV and 9 MeV. The beam collimation was performed through Perspex (PMMA) cylindrical applicators with different diameters. The accurate dose delivery of IORT tightly depends on the precision of measured dose by reference applicator and the output factors of clinical applicators. The output factors were measured using microDiamond and Diode E detectors. The microDiamond detector performance was compared with a Diode E detector. Determined output factors of two detectors were in good agreement. The maximum deviations of output factors for microDiamond were found 2.74%, and 2.17% for 7 MeV and 9 MeV, respectively with respect to the PTW Diode E. The microDiamond detector was shown to exhibit excellent properties for output factor measurements and could be considered as a suitable tool for electron beam dosimetry.
NOVAC7术中放疗加速器两种探测器输出因子的测量与比较
本文的目的是评价和比较较新的合成PTW 60019微金刚石与PTW 60017二极管E探测器在测量术中放射治疗(IORT)电子束输出因子(of)方面的性能。对于给定的电子束,OFs定义为任何尺寸的施加器在最大深度处的剂量与参考施加器的剂量之比。IORT是一种创新的治疗技术,在手术过程中向肿瘤床提供大剂量的单一辐射剂量。本研究中考虑的电子束是由移动NOVAC7系统产生的。该装置产生高剂量脉冲电子束,具有4种不同的能量,范围从3兆电子伏到9兆电子伏。我们进行了两个更高能量的测量,即7MeV和9mev。光束准直是通过不同直径的有机玻璃(PMMA)圆柱形施加器进行的。IORT的准确给药与参考施药器测量剂量的精度和临床施药器的输出系数密切相关。用微金刚石和二极管E探测器测量了输出因子。比较了微金刚石探测器与二极管E探测器的性能。两种检测器的输出系数测定结果一致。在7 MeV和9 MeV时,微金刚石的输出因子与PTW二极管e的最大偏差分别为2.74%和2.17%。结果表明,微金刚石探测器具有良好的输出因子测量性能,可作为电子束剂量测定的合适工具。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信