Mechanized lymphatic drainage in acute decompensated heart failure. A study on a hydrodynamic test bench

A. Buchnev, G. Itkin, A. Drobyshev, A. P. Kuleshov, O. Y. Esipova, A. I. Syrbu
{"title":"Mechanized lymphatic drainage in acute decompensated heart failure. A study on a hydrodynamic test bench","authors":"A. Buchnev, G. Itkin, A. Drobyshev, A. P. Kuleshov, O. Y. Esipova, A. I. Syrbu","doi":"10.15825/1995-1191-2022-4-54-59","DOIUrl":null,"url":null,"abstract":"Objective: to investigate the effectiveness of a new mechanized lymphatic drainage method in acute decompensated heart failure (ADHF) modeling through local reduction in venous pressure in the site of lymphatic drainage from the thoracic duct.Materials and methods. Main components of the device are a catheter with built-in inlet and outlet mechanical valves designed for insertion into the left brachiocephalic vein through the left internal jugular vein. It comes with an extracorporeal drive system made as a valveless pulsator pump with a 10 ml shock discharge and a controller ensuring preset frequency and pressure/rarefaction duty cycle. The operating principle of the device is based on local reduction of venous pressure in the site of lymphatic drainage from the thoracic duct (in the junction of the left internal jugular and subclavian veins).Results. When modeling hydrodynamics under ADHF conditions on a hydrodynamic test bench, the upper venous flow through the left brachiocephalic vein was 0.4 l/min, the pressure in the site of lymphatic drainage from the thoracic duct, was decreased from 20–25 mmHg to 0–5 mmHg due to operation of the mechanized drainage device with suction/injection phase duration ratio 0.2/0.8 and pulsator pump operating frequency from 30 to 60 beats/min.","PeriodicalId":21400,"journal":{"name":"Russian Journal of Transplantology and Artificial Organs","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2022-10-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Russian Journal of Transplantology and Artificial Organs","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.15825/1995-1191-2022-4-54-59","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: to investigate the effectiveness of a new mechanized lymphatic drainage method in acute decompensated heart failure (ADHF) modeling through local reduction in venous pressure in the site of lymphatic drainage from the thoracic duct.Materials and methods. Main components of the device are a catheter with built-in inlet and outlet mechanical valves designed for insertion into the left brachiocephalic vein through the left internal jugular vein. It comes with an extracorporeal drive system made as a valveless pulsator pump with a 10 ml shock discharge and a controller ensuring preset frequency and pressure/rarefaction duty cycle. The operating principle of the device is based on local reduction of venous pressure in the site of lymphatic drainage from the thoracic duct (in the junction of the left internal jugular and subclavian veins).Results. When modeling hydrodynamics under ADHF conditions on a hydrodynamic test bench, the upper venous flow through the left brachiocephalic vein was 0.4 l/min, the pressure in the site of lymphatic drainage from the thoracic duct, was decreased from 20–25 mmHg to 0–5 mmHg due to operation of the mechanized drainage device with suction/injection phase duration ratio 0.2/0.8 and pulsator pump operating frequency from 30 to 60 beats/min.
急性失代偿性心力衰竭的机械淋巴引流。水动力试验台的研究
目的:探讨一种新的机械化淋巴引流方法通过胸导管淋巴引流部位局部降低静脉压在急性失代偿性心力衰竭(ADHF)模型中的应用效果。材料和方法。该装置的主要部件是一根内置进出机械阀的导管,设计用于通过左颈内静脉插入左头臂静脉。它配备了一个体外驱动系统,作为一个无阀脉动泵,具有10毫升冲击放电和一个控制器,确保预设频率和压力/稀薄占空比。该装置的工作原理是在胸导管淋巴引流部位(左颈内静脉和锁骨下静脉交界处)局部降低静脉压。在水动力试验台模拟ADHF条件下的水动力学时,通过左头臂静脉的上静脉流量为0.4 l/min,胸导管淋巴引流部位的压力由20-25 mmHg降至0-5 mmHg,原因是使用了机械化引流装置,吸入/注射相时比为0.2/0.8,脉动泵工作频率为30 ~ 60次/min。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信