DEVELOPMENT AND VALIDATION OF NOVEL UV-VISIBLE SPECTROPHOTOMETRIC METHOD FOR ESTIMATION OF TEPOTINIB IN BULK AND IN PHARMACEUTICAL FORMULATION

Monika Shirwar, Shubhangi Birajdar, S. Garad, S. Kumbhar
{"title":"DEVELOPMENT AND VALIDATION OF NOVEL UV-VISIBLE SPECTROPHOTOMETRIC METHOD FOR ESTIMATION OF TEPOTINIB IN BULK AND IN PHARMACEUTICAL FORMULATION","authors":"Monika Shirwar, Shubhangi Birajdar, S. Garad, S. Kumbhar","doi":"10.22159/ijpps.2023v15i9.48431","DOIUrl":null,"url":null,"abstract":"Objective: The objective of the study is to develop a new, simple, rapid, accurate, and economical UV-spectrophotometric method and validate the same for the estimation of Tepotinib in bulk and in pharmaceutical formulation as per ICH guidelines.\nMethods: The present work was carried out using the UV-Visible double-beam spectrophotometer model Systronics 2201. Tepotinib was found soluble in water, methanol, methanol so, analytical-grade methanol was used as a solvent for conducting the work. The λmax of the tepotinib was determined by dissolving pure drug in methanol scanned in the range of 200-800 nm. The present method was validated for the linearity, accuracy, precision, Limit of Detection and Limit of Quantification.\nResults: The maximum absorbance of tepotinib obtained at a wavelength of 272 nm. The method was found linear in the range of 3-15 µg/ml with the regression coefficient of 0.996 and the equation y = 0.0699x+0.0335. The accuracy was found to be in the range of 96.8-98.5%, the intra-day and inter-day precision % RSD value was 0.262 and 0.69, respectively and the LOD and LOQ were 0.0925 µg/ml and 0.28 µg/ml respectively.\nConclusion: The method demonstrated good reproducibility and recovery so, proposed method can be successfully applied for the routine analysis of tepotinib in bulk and pharmaceutical dosage form.","PeriodicalId":14188,"journal":{"name":"International Journal of Pharmacy and Pharmaceutical Sciences","volume":"72 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2023-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Pharmacy and Pharmaceutical Sciences","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.22159/ijpps.2023v15i9.48431","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: The objective of the study is to develop a new, simple, rapid, accurate, and economical UV-spectrophotometric method and validate the same for the estimation of Tepotinib in bulk and in pharmaceutical formulation as per ICH guidelines. Methods: The present work was carried out using the UV-Visible double-beam spectrophotometer model Systronics 2201. Tepotinib was found soluble in water, methanol, methanol so, analytical-grade methanol was used as a solvent for conducting the work. The λmax of the tepotinib was determined by dissolving pure drug in methanol scanned in the range of 200-800 nm. The present method was validated for the linearity, accuracy, precision, Limit of Detection and Limit of Quantification. Results: The maximum absorbance of tepotinib obtained at a wavelength of 272 nm. The method was found linear in the range of 3-15 µg/ml with the regression coefficient of 0.996 and the equation y = 0.0699x+0.0335. The accuracy was found to be in the range of 96.8-98.5%, the intra-day and inter-day precision % RSD value was 0.262 and 0.69, respectively and the LOD and LOQ were 0.0925 µg/ml and 0.28 µg/ml respectively. Conclusion: The method demonstrated good reproducibility and recovery so, proposed method can be successfully applied for the routine analysis of tepotinib in bulk and pharmaceutical dosage form.
新型紫外可见分光光度法测定原料药和制剂中替波替尼的含量
目的:本研究的目的是建立一种新的、简单、快速、准确、经济的紫外分光光度法,并根据ICH指南验证原料药和制剂中替波替尼的测定方法。方法:采用Systronics 2201型紫外-可见双光束分光光度计。发现替波替尼可溶于水、甲醇、甲醇,因此,分析级甲醇作为溶剂进行工作。通过在200 ~ 800 nm扫描范围内将药物溶解于甲醇中来测定替波替尼的λmax。对方法的线性度、准确度、精密度、检出限和定量限进行了验证。结果:替波替尼在272 nm波长处吸光度最大。该方法在3 ~ 15µg/ml范围内线性良好,回归系数为0.996,方程为y = 0.0699x+0.0335。准确度在96.8 ~ 98.5%之间,日内、日间精密度% RSD分别为0.262和0.69,检出限和定量限分别为0.0925µg/ml和0.28µg/ml。结论:该方法重现性好,回收率高,可用于替泊替尼原料药和制剂制剂的常规分析。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信