Linearity Test for Harshaw TLD (Type: TLD-100H) Base on Individual Calibration Method

Luay Abdulsahib Rasool, Naashat Raheem Al-Ataby, Alaa Fadil Hashim
{"title":"Linearity Test for Harshaw TLD (Type: TLD-100H) Base on Individual Calibration Method","authors":"Luay Abdulsahib Rasool, Naashat Raheem Al-Ataby, Alaa Fadil Hashim","doi":"10.11648/J.NS.20190401.11","DOIUrl":null,"url":null,"abstract":"The testing of the individual monitoring instruments is important to demonstrate the performance of the instruments to give accurate measurements in workplace environment. In this research, 18 Thermoluminescence dosimetry (TLD) units were calibrated individually at surface water phantom and exposed with 60Co source at block 32 in Malaysia Nuclear Agency. The TLD were exposed at 5.00 meter distance from the source. The exposed TLD in terms of Personal Dose Equivalent at 10mm depth tissue, (Hp (10)) equal to 2.00mSv. The exposed TLD then be measured using winRems software from Harshaw TLD reader 6600 plus for defining the calibration factor in term of mSv/nC. After that all the 18 unit TLD were tested using linearity testing method and 18 TLD units were exposed with different dose that were 1mSv, 5mSv, 7mSv, 10mSv, 15mSv, and 20mSv. The research is conducted to satisfy two main objectives which was to obtain linear regression coefficient R2 ~ 1 and to show that the ratio of measured value over standard values are within ICRP trumpet acceptance limit curve, which are within (-33% to +50%).","PeriodicalId":88069,"journal":{"name":"Nuclear science abstracts","volume":"24 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2019-06-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nuclear science abstracts","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.11648/J.NS.20190401.11","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 2

Abstract

The testing of the individual monitoring instruments is important to demonstrate the performance of the instruments to give accurate measurements in workplace environment. In this research, 18 Thermoluminescence dosimetry (TLD) units were calibrated individually at surface water phantom and exposed with 60Co source at block 32 in Malaysia Nuclear Agency. The TLD were exposed at 5.00 meter distance from the source. The exposed TLD in terms of Personal Dose Equivalent at 10mm depth tissue, (Hp (10)) equal to 2.00mSv. The exposed TLD then be measured using winRems software from Harshaw TLD reader 6600 plus for defining the calibration factor in term of mSv/nC. After that all the 18 unit TLD were tested using linearity testing method and 18 TLD units were exposed with different dose that were 1mSv, 5mSv, 7mSv, 10mSv, 15mSv, and 20mSv. The research is conducted to satisfy two main objectives which was to obtain linear regression coefficient R2 ~ 1 and to show that the ratio of measured value over standard values are within ICRP trumpet acceptance limit curve, which are within (-33% to +50%).
Harshaw TLD(型号:TLD- 100h)基于单独校准法的线性测试
单个监测仪器的测试是重要的,以证明仪器的性能,以提供准确的测量工作环境。在这项研究中,18个热释光剂量计(TLD)单元在马来西亚核机构32区块的地表水幻影中单独校准,并暴露于60Co源。TLD暴露在距离辐射源5.00米的地方。以个人剂量当量(Hp(10))计算,暴露于10mm深度组织的TLD等于2.00mSv。然后使用Harshaw TLD阅读器6600 plus的winRems软件测量暴露的TLD,以mSv/nC定义校准因子。然后对18个TLD单元进行线性检测,分别以1mSv、5mSv、7mSv、10mSv、15mSv、20mSv的不同剂量照射18个TLD单元。本研究的主要目的是获得线性回归系数R2 ~ 1,并表明实测值与标准值的比值在ICRP小号接受极限曲线范围内(-33% ~ +50%)。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信