B.W. Sedlmaier , A. Blödow , S. Jovanovic , L. Nagli , H.G. Eberle
{"title":"Laser Myringotomy with Different IR Laser Systems","authors":"B.W. Sedlmaier , A. Blödow , S. Jovanovic , L. Nagli , H.G. Eberle","doi":"10.1016/S0938-765X(98)80020-5","DOIUrl":null,"url":null,"abstract":"<div><p>Lasers suitable for myringotomy are the erbium:YAG laser (2 940 nm) and the CO<sub>2</sub>laser (10 600 nm). The study examines the laser-tissue interaction with tympanic membranes of guineapigs, horses and formalin-fixed human tympanic membranes and the effects demonstrated by light-microscopy and scanning-electron-microscopy. The minimum energy densities for a perforation with the erbium:YAG laser in guinea-pig eardrums and formalin-fixed human tympanic membranes are 8 J/cm<sup>2</sup> and 16 J/cm<sup>2</sup> respectively. There are no thermic side effects. With the CO<sub>2</sub>laser thermic side effects only occur with energy transmission via silver halide polycrystalline fiber. The minimum power density for perforation is 400 W/cm<sup>2</sup> (pulse duration 50 ms). With the microslad 719 micromanipulator (Sharplan, Israel, Tel Aviv), the minimum power densities for perforation of guineapig and horse eardrums and for formalin-fixed human tympanic membranes are 150 W/cm<sup>2</sup>, 300 W/cm<sup>2</sup> and 600 W/cm<sup>2</sup> (pulse duration: 50 ms) respectively. The minimum power density to achieve a perforation with the SwiftLase™ 757 scanner (Sharplan, Israel, Tel Aviv), is 250 W/cm<sup>2</sup> in guinea-pig eardrums (pulse duration: 100 ms). A prototype of a hand-held CO<sub>2</sub>laser otoscope is suitable for performing laser myringotomies in formalin-fixed human tympanic membranes.</p></div><div><p>Die geeigneten Laser für die Lasermyringotomie sind der Erbium: YAG-Laser (2940 nm) und der CO<sub>2</sub>-Laser (10600 nm). Die vorliegende Arbeit untersucht die Laser-Gewebeinteraktion der oben genannten Systeme mit Meerschweinchen- und Pferdetrommelfellen sowie mit formalinfixierten menschlichen Trommelfellen. Die Effekte werden lichtmikroskopisch und rasterelektronenmikroskopisch analysiert. Die minimale Energiedichte für die Perforation der Membrana tympani liegen für den Erbium:YAG-Laser beim Meerschweinchen bei 8 J/cm<sup>2</sup> und beim formalinfixierten humanen Trommelfell bei 16 J/cm<sup>2</sup>. Lichtmikroskopisch und rasterelektronenmikroskopisch sind keine thermischen Nebeneffekte zu sehen.</p><p>Mit dem CO<sub>2</sub>- Laser können ausgedehntere thermische Effekte im umliegenden Gewebe nur bei der Lasermyringotomie mit der polykristallinen Silberhalydfaser erkannt werden. Hier liegt die minimale Leistungsdichte für eine Perforation bei 400 W/cm<sup>2</sup> (Pulsdauer: 50 ms). Mit dem Mikromanipulator Microslad 719 (Sharplan, Israel, Tel Aviv), betragen die minimalen Leistungsdichten für eine Myringotomieperforation beim Meerschweinchentrommelfell 150 W/cm<sup>2</sup>, beim Pferdetrommelfell 300 W/cm<sup>2</sup> und beim formalinfixierten humanen Trommelfell 600 W/cm<sup>2</sup> (Pulsdauer: 50 ms). Die minimale Leistungsdichte, urn eine Perforation mit dem SwiftLase™ 757 Scanner (Sharplan, Israel, Tel Aviv) zu erzielen, beläuft sich beim Meerschweinchentrommelfell auf 250 W/cm<sup>2</sup> (Pulsdauer: 100 ms). Ein Prototyp eines handgeführten CO<sub>2</sub>-Laserotoskops ist geeignet Lasermyringotomien bei formalinfixierten, humanen Trommelfellen durchzuführen.</p></div>","PeriodicalId":100866,"journal":{"name":"Laser-Medizin: eine interdisziplin?re Zeitschrift ; Praxis, Klinik, Forschung","volume":"14 1","pages":"Pages 41-45"},"PeriodicalIF":0.0000,"publicationDate":"1998-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/S0938-765X(98)80020-5","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Laser-Medizin: eine interdisziplin?re Zeitschrift ; Praxis, Klinik, Forschung","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0938765X98800205","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2012/3/15 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Lasers suitable for myringotomy are the erbium:YAG laser (2 940 nm) and the CO2laser (10 600 nm). The study examines the laser-tissue interaction with tympanic membranes of guineapigs, horses and formalin-fixed human tympanic membranes and the effects demonstrated by light-microscopy and scanning-electron-microscopy. The minimum energy densities for a perforation with the erbium:YAG laser in guinea-pig eardrums and formalin-fixed human tympanic membranes are 8 J/cm2 and 16 J/cm2 respectively. There are no thermic side effects. With the CO2laser thermic side effects only occur with energy transmission via silver halide polycrystalline fiber. The minimum power density for perforation is 400 W/cm2 (pulse duration 50 ms). With the microslad 719 micromanipulator (Sharplan, Israel, Tel Aviv), the minimum power densities for perforation of guineapig and horse eardrums and for formalin-fixed human tympanic membranes are 150 W/cm2, 300 W/cm2 and 600 W/cm2 (pulse duration: 50 ms) respectively. The minimum power density to achieve a perforation with the SwiftLase™ 757 scanner (Sharplan, Israel, Tel Aviv), is 250 W/cm2 in guinea-pig eardrums (pulse duration: 100 ms). A prototype of a hand-held CO2laser otoscope is suitable for performing laser myringotomies in formalin-fixed human tympanic membranes.
Die geeigneten Laser für die Lasermyringotomie sind der Erbium: YAG-Laser (2940 nm) und der CO2-Laser (10600 nm). Die vorliegende Arbeit untersucht die Laser-Gewebeinteraktion der oben genannten Systeme mit Meerschweinchen- und Pferdetrommelfellen sowie mit formalinfixierten menschlichen Trommelfellen. Die Effekte werden lichtmikroskopisch und rasterelektronenmikroskopisch analysiert. Die minimale Energiedichte für die Perforation der Membrana tympani liegen für den Erbium:YAG-Laser beim Meerschweinchen bei 8 J/cm2 und beim formalinfixierten humanen Trommelfell bei 16 J/cm2. Lichtmikroskopisch und rasterelektronenmikroskopisch sind keine thermischen Nebeneffekte zu sehen.
Mit dem CO2- Laser können ausgedehntere thermische Effekte im umliegenden Gewebe nur bei der Lasermyringotomie mit der polykristallinen Silberhalydfaser erkannt werden. Hier liegt die minimale Leistungsdichte für eine Perforation bei 400 W/cm2 (Pulsdauer: 50 ms). Mit dem Mikromanipulator Microslad 719 (Sharplan, Israel, Tel Aviv), betragen die minimalen Leistungsdichten für eine Myringotomieperforation beim Meerschweinchentrommelfell 150 W/cm2, beim Pferdetrommelfell 300 W/cm2 und beim formalinfixierten humanen Trommelfell 600 W/cm2 (Pulsdauer: 50 ms). Die minimale Leistungsdichte, urn eine Perforation mit dem SwiftLase™ 757 Scanner (Sharplan, Israel, Tel Aviv) zu erzielen, beläuft sich beim Meerschweinchentrommelfell auf 250 W/cm2 (Pulsdauer: 100 ms). Ein Prototyp eines handgeführten CO2-Laserotoskops ist geeignet Lasermyringotomien bei formalinfixierten, humanen Trommelfellen durchzuführen.