Optimal set-up for ultrasound guided punctures using new scanner applications: an in-vitro study

Steen Karstrup, Jens Brøns, Lone Morsel, Niels Juul, Peter von der Recke
{"title":"Optimal set-up for ultrasound guided punctures using new scanner applications: an in-vitro study","authors":"Steen Karstrup,&nbsp;Jens Brøns,&nbsp;Lone Morsel,&nbsp;Niels Juul,&nbsp;Peter von der Recke","doi":"10.1016/S0929-8266(01)00170-7","DOIUrl":null,"url":null,"abstract":"<div><p><em>Objective:</em> To investigate if US-visualisation of the needle tip echo during US-guided punctures could be improved by use of new technological applications. <em>Method</em>: an US-guided 18 G Trocar needle was inserted into targets of a puncture phantom. The punctures were performed in series of 10 using different settings of the US-scanner (GE Logic 700 Expert). At 7-cm of puncture depth the quality of the echo was tested using four different settings; normal (N), N+automatic tissue optimising (ATO), coded harmonic imaging (CHI), CHI+ATO and at 13-cm of puncture depths six different settings; N, N+ATO, coded excitation (CE), CE+ATO, CHI, CHI+ATO. In total 100 (40+60) images were randomly numbered and read independently by three radiologists with regard to scoring of the quality of the echoes from the needle tip, needle shaft and the target. <em>Results:</em> US visualisation of the needle tip was significantly (<em>P</em>&lt;0.005) improved as compared to normal settings (N) when the settings of ATO and CE were used. CHI resulted in the lowest score. A high agreement between observers was registered. Similar results were registered with regard to scorings from the needle shaft and target. <em>Conclusion</em>: Not only changes of needle designs and puncture techniques can enhance echoes from the needle but also changes in the settings of the US-scanner with the use of new technological applications can improve visualisation of the needle echo.</p></div>","PeriodicalId":79592,"journal":{"name":"European journal of ultrasound : official journal of the European Federation of Societies for Ultrasound in Medicine and Biology","volume":"15 1","pages":"Pages 77-84"},"PeriodicalIF":0.0000,"publicationDate":"2002-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/S0929-8266(01)00170-7","citationCount":"14","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European journal of ultrasound : official journal of the European Federation of Societies for Ultrasound in Medicine and Biology","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0929826601001707","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 14

Abstract

Objective: To investigate if US-visualisation of the needle tip echo during US-guided punctures could be improved by use of new technological applications. Method: an US-guided 18 G Trocar needle was inserted into targets of a puncture phantom. The punctures were performed in series of 10 using different settings of the US-scanner (GE Logic 700 Expert). At 7-cm of puncture depth the quality of the echo was tested using four different settings; normal (N), N+automatic tissue optimising (ATO), coded harmonic imaging (CHI), CHI+ATO and at 13-cm of puncture depths six different settings; N, N+ATO, coded excitation (CE), CE+ATO, CHI, CHI+ATO. In total 100 (40+60) images were randomly numbered and read independently by three radiologists with regard to scoring of the quality of the echoes from the needle tip, needle shaft and the target. Results: US visualisation of the needle tip was significantly (P<0.005) improved as compared to normal settings (N) when the settings of ATO and CE were used. CHI resulted in the lowest score. A high agreement between observers was registered. Similar results were registered with regard to scorings from the needle shaft and target. Conclusion: Not only changes of needle designs and puncture techniques can enhance echoes from the needle but also changes in the settings of the US-scanner with the use of new technological applications can improve visualisation of the needle echo.

超声引导穿刺的最佳设置使用新的扫描仪应用:体外研究
目的:研究是否可以通过使用新的技术应用来改善US引导穿刺过程中针尖回声的US可视化。方法:将美国引导的18G Trocar针插入穿刺体模的目标中。使用美国扫描仪(GE Logic 700 Expert)的不同设置以10个为一系列进行穿刺。在穿刺深度为7cm时,使用四种不同的设置来测试回声的质量;正常(N)、N+自动组织优化(ATO)、编码谐波成像(CHI)、CHI+ATO和在13cm穿刺深度的六种不同设置;N、 N+ATO,编码励磁(CE),CE+ATO,CHI,CHI+ATO。总共有100(40+60)张图像被随机编号,并由三名放射科医生独立读取,以对来自针尖、针轴和目标的回波质量进行评分。结果:当使用ATO和CE的设置时,与正常设置(N)相比,针尖的US可视化显著改善(P<;0.005)。CHI得分最低。观察员之间达成了高度一致。关于针轴和目标的烧焦,记录了类似的结果。结论:不仅改变针头设计和穿刺技术可以增强针头的回声,而且随着新技术的应用,改变US扫描仪的设置可以提高针头回声的可视化。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信