{"title":"Precision and Bias of Graphite Furnace Analysis of Environmental Samples","authors":"G. L. Mckinney","doi":"10.6028/jres.093.056","DOIUrl":null,"url":null,"abstract":"* Determination of Cr(lIl)+Cr(VI). To 25 mL of sample, add 0.3 mL of 8% PHP, adjust pH to 3.5, and add 3 mL of 10% APDC and 5 mL of PHP-saturated MIBK. Extract for 20 min. Determine Cr(lll)+Cr(VI) in the MIBK layer. * Determination of Cr(VI). Proceed as above except for the addition of 3 mL of 8% PHP and 3 mL of 2% APDC and for the extraction time of 10 mm. * Determination of Cr(III). The value of Cr(III) is obtained as the difference between Cr(Ill) +Cr(VI) and Cr(VI). The concentration of Cr(111) is also determined by the SM-7 ion exchange procedure [1]. In all the above cases, the Cr atomic absorption signal was measured by using the dry/atomize program of 900 'C-30 s (ramp)-30 s (hold)/2500 0 C-O s (ramp)-6 s (hold).","PeriodicalId":17082,"journal":{"name":"Journal of research of the National Bureau of Standards","volume":"93 1","pages":"307 - 310"},"PeriodicalIF":0.0000,"publicationDate":"1988-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of research of the National Bureau of Standards","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6028/jres.093.056","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
* Determination of Cr(lIl)+Cr(VI). To 25 mL of sample, add 0.3 mL of 8% PHP, adjust pH to 3.5, and add 3 mL of 10% APDC and 5 mL of PHP-saturated MIBK. Extract for 20 min. Determine Cr(lll)+Cr(VI) in the MIBK layer. * Determination of Cr(VI). Proceed as above except for the addition of 3 mL of 8% PHP and 3 mL of 2% APDC and for the extraction time of 10 mm. * Determination of Cr(III). The value of Cr(III) is obtained as the difference between Cr(Ill) +Cr(VI) and Cr(VI). The concentration of Cr(111) is also determined by the SM-7 ion exchange procedure [1]. In all the above cases, the Cr atomic absorption signal was measured by using the dry/atomize program of 900 'C-30 s (ramp)-30 s (hold)/2500 0 C-O s (ramp)-6 s (hold).