Compliance of the POCT method with the fully automated method for HbA1c determination

Q4 Pharmacology, Toxicology and Pharmaceutics
N. Milinković, M. Sarić-Matutinović, Gordana Dmitrašinović, S. Ignjatović
{"title":"Compliance of the POCT method with the fully automated method for HbA1c determination","authors":"N. Milinković, M. Sarić-Matutinović, Gordana Dmitrašinović, S. Ignjatović","doi":"10.5937/arhfarm72-35905","DOIUrl":null,"url":null,"abstract":"Previous research suggests that point-of-care (POCT) determination of glycated hemoglobin (HbA1c) is a diagnostic test that can be an adequate alternative to measuring HbA1c in the laboratory. The main goal of this study was to examine the analytical characteristics of the novel INCLIX POCT method for HbA1c determination in order to test its performance before introducing this method into routine use. HbA1c is measured in a duplicate in 44 EDTA blood samples parallel on INCLIX POCT device (Sugitech, Inc.) and using automated turbidimetric immunoinhibition test on Olympus AU400 (Beckman Coulter). The within run imprecision was 7.58%, between runs imprecision was 6.63% and 6.22%, and day-to-day imprecision was 8.80% and 7.51%. Total laboratory imprecision was in agreement with those stated by the manufacturer. A statistically significant Pearson correlation coefficient was calculated (r = 0.871, P < 0.01; linear R2 = 0.757). Using Deming regression analysis, the following equation was obtained: y = - 1.80 + 1.304x. Our results indicate statistically significant correlation, linear relationship, and a significant degree of compatibility between the two analyzed methods. However, the negative bias of the HbA1c values determined on the POCT analyzer compared to the Olympus AU400 was confirmed, highlighting the need to standardize the INCLIX method.","PeriodicalId":39173,"journal":{"name":"Arhiv za Farmaciju","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2022-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Arhiv za Farmaciju","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5937/arhfarm72-35905","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Pharmacology, Toxicology and Pharmaceutics","Score":null,"Total":0}
引用次数: 0

Abstract

Previous research suggests that point-of-care (POCT) determination of glycated hemoglobin (HbA1c) is a diagnostic test that can be an adequate alternative to measuring HbA1c in the laboratory. The main goal of this study was to examine the analytical characteristics of the novel INCLIX POCT method for HbA1c determination in order to test its performance before introducing this method into routine use. HbA1c is measured in a duplicate in 44 EDTA blood samples parallel on INCLIX POCT device (Sugitech, Inc.) and using automated turbidimetric immunoinhibition test on Olympus AU400 (Beckman Coulter). The within run imprecision was 7.58%, between runs imprecision was 6.63% and 6.22%, and day-to-day imprecision was 8.80% and 7.51%. Total laboratory imprecision was in agreement with those stated by the manufacturer. A statistically significant Pearson correlation coefficient was calculated (r = 0.871, P < 0.01; linear R2 = 0.757). Using Deming regression analysis, the following equation was obtained: y = - 1.80 + 1.304x. Our results indicate statistically significant correlation, linear relationship, and a significant degree of compatibility between the two analyzed methods. However, the negative bias of the HbA1c values determined on the POCT analyzer compared to the Olympus AU400 was confirmed, highlighting the need to standardize the INCLIX method.
POCT方法与全自动HbA1c测定方法的一致性
先前的研究表明,即时检测(POCT)糖化血红蛋白(HbA1c)是一种诊断测试,可以作为实验室测量HbA1c的适当替代方法。本研究的主要目的是研究新型的INCLIX POCT HbA1c检测方法的分析特点,以便在将该方法引入常规使用之前测试其性能。在INCLIX POCT设备(Sugitech, Inc.)上平行测量44份EDTA血样中的HbA1c,并使用Olympus AU400 (Beckman Coulter)上的自动浊度免疫抑制试验。运行内不精度为7.58%,运行间不精度为6.63%和6.22%,日常不精度为8.80%和7.51%。实验室总不精确度与制造商所述一致。Pearson相关系数有统计学意义(r = 0.871, P < 0.01;线性R2 = 0.757)。采用Deming回归分析得到:y = - 1.80 + 1.304x。我们的结果表明,两种分析方法之间存在统计学上显著的相关性、线性关系和显著程度的相容性。然而,与Olympus AU400相比,POCT分析仪测定的HbA1c值存在负偏倚,这凸显了对INCLIX方法进行标准化的必要性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Arhiv za Farmaciju
Arhiv za Farmaciju Pharmacology, Toxicology and Pharmaceutics-Pharmaceutical Science
自引率
0.00%
发文量
19
审稿时长
12 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信