Studying of Common Factors Affecting Reading of Lithium Fluoride Thermo-Luminescence Dosimeter Crystal

M. Omer, Abdulrahman A. Alsayyari, J. Tamboul, R. B. Ali, A. A. Ahmed, A. Abukonna
{"title":"Studying of Common Factors Affecting Reading of Lithium Fluoride Thermo-Luminescence Dosimeter Crystal","authors":"M. Omer, Abdulrahman A. Alsayyari, J. Tamboul, R. B. Ali, A. A. Ahmed, A. Abukonna","doi":"10.4236/OJBIPHY.2017.71002","DOIUrl":null,"url":null,"abstract":"The aim of this study was to study the technical factors affecting the reading of TLD (LiF:Mg:Ti) for radiation dosimetry. The method adapted was an exposure of LiF:Mg:Ti crystal with γ-ray with the following filters (Hp 0.07 and Hp 10), before and after usage of filter Hp (0.07), annealing, cooling by nitrogen flux and crystal morphology by scanning electron microscope. The collected data analyzed by EXCELL software showed that: TLD average count/second (C/s) with filter Hp (10) was greater than that obtained with filter Hp (0.07) with an amount of 1191 c/s (19%). Also the analysis showed that: the average C/s without filter Hp (0.07) was greater than that with filter by an amount of 994.5 C/s (16.4%), and there was proportional linear relationship; as the applied dose increased the TLD C/s increased significantly (R2 = 0.9) based on the equations: y = 885.1x ? 286.5 and y = 860.5x + 856.8 for filtered and without filtered beam respectively, where x refers to applied dose in mGy and y refers to C/s. Also the average C/s without nitrogen flux (cooling) was greater than with nitrogen count with an amount of 52.4 C/s (11%). The annealing (without filter) causes a reduction in dose by 82.2 μGy (19.2%), and in case of annealing with filter, the dose reduced by 66.7 μGy (21.9%) relative to annealing without filter. The morphology of TLD crystal shown by scanning electron microscope SEM was a darkening surface after irradiation.","PeriodicalId":59528,"journal":{"name":"生物物理学期刊(英文)","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2017-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"生物物理学期刊(英文)","FirstCategoryId":"1089","ListUrlMain":"https://doi.org/10.4236/OJBIPHY.2017.71002","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

The aim of this study was to study the technical factors affecting the reading of TLD (LiF:Mg:Ti) for radiation dosimetry. The method adapted was an exposure of LiF:Mg:Ti crystal with γ-ray with the following filters (Hp 0.07 and Hp 10), before and after usage of filter Hp (0.07), annealing, cooling by nitrogen flux and crystal morphology by scanning electron microscope. The collected data analyzed by EXCELL software showed that: TLD average count/second (C/s) with filter Hp (10) was greater than that obtained with filter Hp (0.07) with an amount of 1191 c/s (19%). Also the analysis showed that: the average C/s without filter Hp (0.07) was greater than that with filter by an amount of 994.5 C/s (16.4%), and there was proportional linear relationship; as the applied dose increased the TLD C/s increased significantly (R2 = 0.9) based on the equations: y = 885.1x ? 286.5 and y = 860.5x + 856.8 for filtered and without filtered beam respectively, where x refers to applied dose in mGy and y refers to C/s. Also the average C/s without nitrogen flux (cooling) was greater than with nitrogen count with an amount of 52.4 C/s (11%). The annealing (without filter) causes a reduction in dose by 82.2 μGy (19.2%), and in case of annealing with filter, the dose reduced by 66.7 μGy (21.9%) relative to annealing without filter. The morphology of TLD crystal shown by scanning electron microscope SEM was a darkening surface after irradiation.
影响氟化锂热发光剂量计晶体读数的常见因素研究
本研究的目的是研究影响TLD (LiF:Mg:Ti)读数的技术因素。采用的方法是用γ射线对LiF:Mg:Ti晶体用Hp(0.07)和Hp(10)滤光片曝光,使用Hp(0.07)滤光片前后进行退火、氮通量冷却和扫描电镜观察晶体形貌。用EXCELL软件对采集的数据进行分析,结果表明:过滤器Hp(10)的TLD平均计数/秒(C/s)大于过滤器Hp(0.07)的TLD平均计数/秒(1191 C/s)(19%)。分析还表明:未加过滤器Hp(0.07)的平均C/s比加过滤器的平均C/s大994.5 C/s(16.4%),两者呈正比线性关系;随着剂量的增加,TLD C/s显著增加(R2 = 0.9),根据公式:y = 885.1x ?滤光束和未滤光束分别为286.5和y = 860.5x + 856.8,其中x为辐照剂量,单位为mGy, y为C/s。无氮通量(冷却)的平均C/s大于有氮计数的平均C/s,为52.4 C/s(11%)。退火(不带滤波器)使辐照剂量降低82.2 μGy(19.2%),带滤波器退火时,辐照剂量比不带滤波器退火时降低66.7 μGy(21.9%)。扫描电镜显示,辐照后的TLD晶体形貌为暗化表面。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
128
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信