Margin-of-safety Algorithm Used with EOS Imaging to Interpret MHRA Warning for 46-48mm MOM Arthroplasty

I. Clarke, J. Lazennec
{"title":"Margin-of-safety Algorithm Used with EOS Imaging to Interpret MHRA Warning for 46-48mm MOM Arthroplasty","authors":"I. Clarke, J. Lazennec","doi":"10.15438/RR.5.3.115","DOIUrl":null,"url":null,"abstract":"The Medical Healthcare Products Regulatory Agency (MHRA June-2015) warned of higher risks with 46-48mm sizes of BHR hip resurfacing arthroplasty (HRA). The most common condemnation of adverse results in MOM bearings has been termed edge loading. We originally developed a margin-of-safety (MOS) algorithm to define edge loading of cups in simulator studies. This method integrated simulator wear-patterns with respect to cup diameters and cup designs. The algorithm’s simplicity lay in the fact that with wear-patterns and rim-profile angles predetermined, the only input required was the cup inclination-angle. The algorithm demonstrated that the margin-of-safety decreased in smaller cups due to the tribo-mechanics of spherical CoCr bearings, a previously unrecognized feature. For the 46mm and 48mm cups highlighted in the MHRA alert, the critical cup inclinations where edge-wear became a risk occurred at 65-66°, revealing an insignificant difference with respect to diameters. The MOS-algorithm also indicated that lower lateral-inclination angles were particularly beneficial, i.e. a 46mm cup positioned at 50° inclination would exhibit a higher margin of safety than either 48mm or 50mm sizes positioned at 55° inclination. This evidence supported clinical studies that recommended BHR cup inclinations up to 50-55° and lower as optimal for reducing metal-ion concentrations. In a patient with normal spine mobility, our EOS imaging demonstrated that the inclination in the 46mm cup steepened by 9° from standing to the seated position while margin-of-safety was reduced by 50%. Our 2nd patient with a stiff spine sat with the same component orientations as in his standing posture. Thus MOM impingement and subluxation in different functional postures may also provoke rim-damage mechanisms. Here the combination of EOS imaging and the MOS-algorithm may aid understanding of such risks. Thus the margin-of-safety algorithm confirmed and helped explained the relative risks in the 46mm and 48mm cups highlighted by the MHRA. The algorithm’s stratification by cup rim-profile, inclination angle and cup diameter may assist the surgeon determine which patients may be more at risk for edge wear with the smaller BHR cups.","PeriodicalId":20884,"journal":{"name":"Reconstructive Review","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2015-10-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"7","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Reconstructive Review","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.15438/RR.5.3.115","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 7

Abstract

The Medical Healthcare Products Regulatory Agency (MHRA June-2015) warned of higher risks with 46-48mm sizes of BHR hip resurfacing arthroplasty (HRA). The most common condemnation of adverse results in MOM bearings has been termed edge loading. We originally developed a margin-of-safety (MOS) algorithm to define edge loading of cups in simulator studies. This method integrated simulator wear-patterns with respect to cup diameters and cup designs. The algorithm’s simplicity lay in the fact that with wear-patterns and rim-profile angles predetermined, the only input required was the cup inclination-angle. The algorithm demonstrated that the margin-of-safety decreased in smaller cups due to the tribo-mechanics of spherical CoCr bearings, a previously unrecognized feature. For the 46mm and 48mm cups highlighted in the MHRA alert, the critical cup inclinations where edge-wear became a risk occurred at 65-66°, revealing an insignificant difference with respect to diameters. The MOS-algorithm also indicated that lower lateral-inclination angles were particularly beneficial, i.e. a 46mm cup positioned at 50° inclination would exhibit a higher margin of safety than either 48mm or 50mm sizes positioned at 55° inclination. This evidence supported clinical studies that recommended BHR cup inclinations up to 50-55° and lower as optimal for reducing metal-ion concentrations. In a patient with normal spine mobility, our EOS imaging demonstrated that the inclination in the 46mm cup steepened by 9° from standing to the seated position while margin-of-safety was reduced by 50%. Our 2nd patient with a stiff spine sat with the same component orientations as in his standing posture. Thus MOM impingement and subluxation in different functional postures may also provoke rim-damage mechanisms. Here the combination of EOS imaging and the MOS-algorithm may aid understanding of such risks. Thus the margin-of-safety algorithm confirmed and helped explained the relative risks in the 46mm and 48mm cups highlighted by the MHRA. The algorithm’s stratification by cup rim-profile, inclination angle and cup diameter may assist the surgeon determine which patients may be more at risk for edge wear with the smaller BHR cups.
应用安全边际算法与EOS成像解释46-48mm MOM关节置换术的MHRA警告
医疗保健产品监管机构(MHRA)于2015年6月警告46-48mm尺寸的BHR髋关节置换术(HRA)风险较高。最常见的谴责不利结果在MOM轴承已被称为边缘加载。我们最初在模拟器研究中开发了一种安全边际(MOS)算法来定义杯子的边缘载荷。该方法将模拟器磨损模式与杯径和杯形设计相结合。该算法的简单之处在于,磨损模式和边缘轮廓角度都是预先确定的,唯一需要输入的是杯子的倾斜角。该算法表明,由于球形CoCr轴承的摩擦力学,小杯的安全边际减小,这是一个以前未被认识到的特征。对于MHRA警报中突出显示的46mm和48mm的杯形,在65-66°之间的杯形倾斜度会产生边缘磨损的风险,这表明它们的直径差异不显著。mos算法还表明,较低的横向倾角特别有利,即倾角为50°的46mm杯比倾角为55°的48mm或50mm杯具有更高的安全边际。这一证据支持临床研究建议BHR杯的倾斜度达到50-55°或更低是降低金属离子浓度的最佳方法。在一名脊柱活动正常的患者中,我们的EOS成像显示46mm杯的倾斜度从站立到坐姿增加了9°,而安全边缘减少了50%。我们的第二个脊柱僵硬的病人坐着的姿势和他站立时的姿势相同。因此,不同功能姿势的MOM撞击和半脱位也可能引起环损伤机制。在这里,EOS成像和mos算法的结合可能有助于了解此类风险。因此,安全边际算法确认并帮助解释了MHRA强调的46毫米和48毫米杯子的相对风险。该算法根据罩杯边缘轮廓、倾斜角和罩杯直径进行分层,可以帮助外科医生确定哪些患者使用较小的BHR罩杯更容易出现边缘磨损。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
2
审稿时长
24 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信