{"title":"A Method of Iodine Determination by Characteristic X-Ray Absorption","authors":"O. Z. Roy, R. Béique, G. L. D'ombrain","doi":"10.1109/TBMEL.1962.4322949","DOIUrl":null,"url":null,"abstract":"This paper describes a technique for the quantitative analysis of nonradioactive iodine. The method uses the X-ray absorption discontinuities for the detection of iodine in the presence of any other absorbing substance. A stationary lanthanum radiator converts an X-ray beam having a continuous spectrum into two monochromatic beams which straddle the iodine absorption edge. The energy spread is such as to make the technique relatively independent of other foreign elements. A rotating filter alternately absorbs one of the two beams while leaving the other on continuously. The information in the beams is detected and synchronously switched into two counting circuits. The intensity of the beams is measured by two counting-rate meters and the ratio of these intensities is obtained by a conventional self-balancing potentiometer. Since this ratio is determined every 1/60 of a second the iodine determination is insensitive to X-ray output, amplifier and detector drift.","PeriodicalId":86470,"journal":{"name":"IRE transactions on bio-medical electronics","volume":"9 1","pages":"50-53"},"PeriodicalIF":0.0000,"publicationDate":"1962-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1109/TBMEL.1962.4322949","citationCount":"4","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"IRE transactions on bio-medical electronics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/TBMEL.1962.4322949","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 4

Abstract

This paper describes a technique for the quantitative analysis of nonradioactive iodine. The method uses the X-ray absorption discontinuities for the detection of iodine in the presence of any other absorbing substance. A stationary lanthanum radiator converts an X-ray beam having a continuous spectrum into two monochromatic beams which straddle the iodine absorption edge. The energy spread is such as to make the technique relatively independent of other foreign elements. A rotating filter alternately absorbs one of the two beams while leaving the other on continuously. The information in the beams is detected and synchronously switched into two counting circuits. The intensity of the beams is measured by two counting-rate meters and the ratio of these intensities is obtained by a conventional self-balancing potentiometer. Since this ratio is determined every 1/60 of a second the iodine determination is insensitive to X-ray output, amplifier and detector drift.
特征x射线吸收法测定碘
本文介绍了一种非放射性碘的定量分析方法。该方法使用x射线吸收不连续来检测存在任何其他吸收物质的碘。固定的镧辐射器将具有连续光谱的x射线束转换成跨越碘吸收边缘的两个单色光束。能量的扩散使得该技术相对独立于其他外来元素。一个旋转的滤光器交替地吸收两束光中的一束,同时使另一束继续打开。光束中的信息被检测并同步切换到两个计数电路中。光束的强度由两个计数速率计测量,这些强度的比值由传统的自平衡电位器获得。由于该比值每1/60秒测定一次,所以碘的测定对x射线输出、放大器和探测器漂移不敏感。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信