Changes in corneal thickness after daily and extended wear of hydrogel lenses: a comparison of optical and ultrasonic pachometry

Eva Yebra-Pimentel MS, PhD (OD), María Jesús Giráldez (OD), Javier González (OD), María Victoría Pérez MS, PhD, Manuel Parafita MS, MD, PhD (Dipl Ophthalmol)
{"title":"Changes in corneal thickness after daily and extended wear of hydrogel lenses: a comparison of optical and ultrasonic pachometry","authors":"Eva Yebra-Pimentel MS, PhD (OD),&nbsp;María Jesús Giráldez (OD),&nbsp;Javier González (OD),&nbsp;María Victoría Pérez MS, PhD,&nbsp;Manuel Parafita MS, MD, PhD (Dipl Ophthalmol)","doi":"10.1016/S0892-8967(98)00021-2","DOIUrl":null,"url":null,"abstract":"<div><p>Central corneal thickness of both eyes of 18 normal subjects was measured in a study comparing a Haag-Streit optical pachometer and an ultrasonic pachometer (Teknar Ophthasonic, preset velocity 1630 m/sec). All subjects were awake for 3 hours before measurements to reduce swelling of the cornea from overnight eye closure<span>. Reliability was assessed by a three-way analysis of variance between repeated measurements for the optical and ultrasonic pachometers. The former revealed a systematic right–left difference, the left cornea being significantly thicker. The correlation coefficient indicates that the link between the two variables is 0.92 and 0.88 for right and left eyes, respectively. It is concluded that, for clinical purposes, optical and ultrasonic pachometry techniques are comparable. The effect of the daily and extended-wear contact lenses on central corneal thickness also was studied. The degree of change (morning vs evening) in corneal thickness of the polymacon group was lower and decreased with the wear, whereas in the etafilcon A group the change was greater and almost constant with the time of wear.</span></p></div>","PeriodicalId":80286,"journal":{"name":"International contact lens clinic (New York, N.Y.)","volume":"25 4","pages":"Pages 103-108"},"PeriodicalIF":0.0000,"publicationDate":"1998-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1016/S0892-8967(98)00021-2","citationCount":"11","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International contact lens clinic (New York, N.Y.)","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0892896798000212","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 11

Abstract

Central corneal thickness of both eyes of 18 normal subjects was measured in a study comparing a Haag-Streit optical pachometer and an ultrasonic pachometer (Teknar Ophthasonic, preset velocity 1630 m/sec). All subjects were awake for 3 hours before measurements to reduce swelling of the cornea from overnight eye closure. Reliability was assessed by a three-way analysis of variance between repeated measurements for the optical and ultrasonic pachometers. The former revealed a systematic right–left difference, the left cornea being significantly thicker. The correlation coefficient indicates that the link between the two variables is 0.92 and 0.88 for right and left eyes, respectively. It is concluded that, for clinical purposes, optical and ultrasonic pachometry techniques are comparable. The effect of the daily and extended-wear contact lenses on central corneal thickness also was studied. The degree of change (morning vs evening) in corneal thickness of the polymacon group was lower and decreased with the wear, whereas in the etafilcon A group the change was greater and almost constant with the time of wear.

水凝胶隐形眼镜日常佩戴和长时间佩戴后角膜厚度的变化:光学和超声厚度测量的比较
采用Haag-Streit光学测角膜仪与超声波测角膜仪(Teknar ophthultrasonic,预设速度1630 m/sec)对18名正常人双眼角膜中央厚度进行了测量。所有受试者在测量前保持清醒3小时,以减少夜间闭眼造成的角膜肿胀。通过对光学和超声测厚仪重复测量之间的三向方差分析来评估可靠性。前者显示出系统性的左右差异,左侧角膜明显变厚。相关系数表明,右眼和左眼两个变量之间的联系分别为0.92和0.88。结论是,对于临床目的,光学和超声测厚技术是可比较的。研究了日常佩戴和长时间佩戴的隐形眼镜对角膜中央厚度的影响。polymacon组的角膜厚度变化程度(早晨和晚上)较低,随磨损而下降,而依他非康A组的变化程度较大,随磨损时间几乎不变。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信