O. Nordström, P. Molander, T. Greibrokk, R. Blomhoff, E. Lundanes
{"title":"Evaluation of temperature programming for gradient elution in packed capillary liquid chromatography coupled to electrochemical detection","authors":"O. Nordström, P. Molander, T. Greibrokk, R. Blomhoff, E. Lundanes","doi":"10.1002/mcs.1039","DOIUrl":null,"url":null,"abstract":"<p>A miniaturized amperometric detector compatible with temperature-programmed packed capillary liquid chromatography has been constructed and evaluated. Temperature programming could successfully be used as a tool for retention control, providing gradient action that did not influence the detector performance. A 7-μm single carbon fiber was utilized as working electrode and detection was performed in a 50-μm i.d. fused silica detection capillary which also functioned as restriction capillary. Separation of tyrosine, 3-(3,4-dihydroxyphenyl)alanine, 3-chloro-tyrosine and 3-NO<sub>2</sub>-tyrosine was achieved in 6 min using a temperature program from 30°C to 100°C with a ramp of 8°C/min on a 25 cm 0.32 mm i.d. column packed with 3 μm Hypersil ODS. The mobile phase consisted of 50 mM citric acid (pH 3.10)-methanol (95+ 5, v/v), and the mass limit of detection was in the range 1–5 pg using 1.2 V vs the Ag/AgCl reference electrode. © 2001 John Wiley & Sons, Inc. J Micro Sep 13: 179–185, 2001</p>","PeriodicalId":83120,"journal":{"name":"The journal of microcolumn separations : JMS","volume":"13 5","pages":"179-185"},"PeriodicalIF":0.0000,"publicationDate":"2001-08-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1002/mcs.1039","citationCount":"10","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"The journal of microcolumn separations : JMS","FirstCategoryId":"1085","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/mcs.1039","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 10
Abstract
A miniaturized amperometric detector compatible with temperature-programmed packed capillary liquid chromatography has been constructed and evaluated. Temperature programming could successfully be used as a tool for retention control, providing gradient action that did not influence the detector performance. A 7-μm single carbon fiber was utilized as working electrode and detection was performed in a 50-μm i.d. fused silica detection capillary which also functioned as restriction capillary. Separation of tyrosine, 3-(3,4-dihydroxyphenyl)alanine, 3-chloro-tyrosine and 3-NO2-tyrosine was achieved in 6 min using a temperature program from 30°C to 100°C with a ramp of 8°C/min on a 25 cm 0.32 mm i.d. column packed with 3 μm Hypersil ODS. The mobile phase consisted of 50 mM citric acid (pH 3.10)-methanol (95+ 5, v/v), and the mass limit of detection was in the range 1–5 pg using 1.2 V vs the Ag/AgCl reference electrode. © 2001 John Wiley & Sons, Inc. J Micro Sep 13: 179–185, 2001
填充毛细管液相色谱-电化学耦合检测梯度洗脱温度规划的评价
构建并评价了与程序升温填充毛细管液相色谱兼容的小型安培检测器。温度编程可以成功地用作保留控制的工具,提供不影响检测器性能的梯度作用。采用7 μm的单碳纤维作为工作电极,在50 μm的熔融石英检测毛细管中进行检测。酪氨酸、3-(3,4-二羟基苯基)丙氨酸、3-氯酪氨酸和3- no2酪氨酸的分离在6分钟内完成,温度范围从30°C到100°C,坡度为8°C/min,色谱柱为25 cm 0.32 mm,填充3 μm Hypersil ODS。流动相为50 mM柠檬酸(pH 3.10)-甲醇(95+ 5,v/v),使用1.2 v vs Ag/AgCl参比电极,检测质量限在1-5 pg范围内。©2001 John Wiley &[J]微型电子,2001
本文章由计算机程序翻译,如有差异,请以英文原文为准。