Aminotransferase Activity Measurement on The Atellica® CH 930 and The Architect® ci4100

Soumia Nachate, Hind Zrikem, Fayrouz Debbagh, S. Chellak, A. Boukhira
{"title":"Aminotransferase Activity Measurement on The Atellica® CH 930 and The Architect® ci4100","authors":"Soumia Nachate, Hind Zrikem, Fayrouz Debbagh, S. Chellak, A. Boukhira","doi":"10.24018/ejbio.2023.4.4.472","DOIUrl":null,"url":null,"abstract":"\n\n\n\nThe Atellica® CH 930 is a new biochemistry analyzer recently introduced to the world market. This study aims to compare the Atellica® CH 930 analyzer to the Architect® ci4100 analyzer for measurement of serum aspartate (AST) and alanine (ALT) aminotransferase concentrations. A total of 112 sera were tested on the Architect® ci4100 and the Atellica® CH 930 analyzers for ALT and AST activities, and the results were compared using the paired Student's t-test, Pearson’s correlation analysis, Passing-Bablok regression analysis, and Bland–Altman plots. There was no significant difference between the means of the ALT and AST concentrations measured using the two instruments (p = 0.659 and p = 0.506, respectively). For ALT, the Passing-Bablok equation was Atellica=1.11 ×Architect -0.96, with a correlation coefficient (r) of 0.999. For AST, it was Atellica=1.08 ×Architect -0.50, with r = 0.998. According to Bland-Altman plots, the mean difference between both methods was 2.3 IU/L (4.3%) for ALT and 2.1 IU/L (5.8%) for AST, with 95.53% and 96.43% of points within the interval [0 ± 1.96 × SD], respectively. These findings demonstrated that the Atellica® CH 930 and Architect® ci4100 methods had an overall good agreement in aminotransferase measurement.\n\n\n\n\n ","PeriodicalId":72969,"journal":{"name":"European journal of biology and biotechnology","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2023-08-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"European journal of biology and biotechnology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.24018/ejbio.2023.4.4.472","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The Atellica® CH 930 is a new biochemistry analyzer recently introduced to the world market. This study aims to compare the Atellica® CH 930 analyzer to the Architect® ci4100 analyzer for measurement of serum aspartate (AST) and alanine (ALT) aminotransferase concentrations. A total of 112 sera were tested on the Architect® ci4100 and the Atellica® CH 930 analyzers for ALT and AST activities, and the results were compared using the paired Student's t-test, Pearson’s correlation analysis, Passing-Bablok regression analysis, and Bland–Altman plots. There was no significant difference between the means of the ALT and AST concentrations measured using the two instruments (p = 0.659 and p = 0.506, respectively). For ALT, the Passing-Bablok equation was Atellica=1.11 ×Architect -0.96, with a correlation coefficient (r) of 0.999. For AST, it was Atellica=1.08 ×Architect -0.50, with r = 0.998. According to Bland-Altman plots, the mean difference between both methods was 2.3 IU/L (4.3%) for ALT and 2.1 IU/L (5.8%) for AST, with 95.53% and 96.43% of points within the interval [0 ± 1.96 × SD], respectively. These findings demonstrated that the Atellica® CH 930 and Architect® ci4100 methods had an overall good agreement in aminotransferase measurement.  
Atellica®CH 930和The Architect®ci4100上的氨基转移酶活性测量
Atellica®CH 930是最近引入世界市场的一种新型生物化学分析仪。本研究旨在将Atellica®CH 930分析仪与Architect®ci4100分析仪进行比较,以测量血清天冬氨酸(AST)和丙氨酸(ALT)转氨酶浓度。在Architect®ci4100和Atellica®CH 930分析仪上对总共112份血清进行ALT和AST活性测试,并使用配对Student t检验、Pearson相关分析、Passing Bablok回归分析和Bland–Altman图对结果进行比较。使用两种仪器测量的ALT和AST浓度的平均值之间没有显著差异(分别为p=0.659和p=0.506)。对于ALT,Passing Bablok方程为Atellica=1.11×Architect-0.96,相关系数(r)为0.999。AST为Atellica=1.08×Architect-0.50,r=0.998。根据Bland-Altman图,两种方法之间的ALT平均差异为2.3 IU/L(4.3%),AST平均差异为2.1 IU/L(5.8%),区间[0±1.96×SD]内的点分别为95.53%和96.43%。这些发现表明,Atellica®CH 930和Architect®ci4100方法在转氨酶测量方面总体一致。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信