An Enumeration Survey on Diagnostic X-Ray Generators and Essential Safety Parameters in Mizoram, India

Q3 Health Professions
R. Tiwari, J. Lalrinmawia, K. S. Pau
{"title":"An Enumeration Survey on Diagnostic X-Ray Generators and Essential Safety Parameters in Mizoram, India","authors":"R. Tiwari, J. Lalrinmawia, K. S. Pau","doi":"10.22038/IJMP.2019.39307.1519","DOIUrl":null,"url":null,"abstract":"Introduction: Best radiography practice involves operational optimal machine performance, delivering cost-effective healthcare services under appropriate safety conditions for workers and the public. The present study aimed to investigate the safety status of diagnostic X-ray installations in Mizoram, India. Material and Methods: Linearity of time (sec), linearity of current (mA), output reproducibility, table dose (μGy/mAs), peak voltage (kVp) accuracy, and 16 essential safety parameters of 135 X-ray machines were considered in this study. A battery-operated dosimeter and wide-range digital kVp meter were used to measure output radiation and effective peak potential of X-ray tube. Data analysis was performed using SPSS software to obtain the mean, standard deviation, and coefficient of variation. Results: Among different electronic parameters, 59.2% linearity of time, 82.6% linearity of current, 89.7% kVp accuracy, 35.1% output reproducibility, and 92.8% table dose were beyond the acceptable limits. Based on 16 essential safety parameters, it was observed that 98.7% of X-ray machines did not receive proper quality assurance test, 1.9% of the installations employed lead-line patient entrance doors, 46.8% of the machines were operated without any protective barriers and 83.1% of the units were operated without personnel monitoring service. Conclusion: The present study had concluded with more problems than the previous studies in different parts of the world in this regard. Due to the absence of proper quality control (QC) programs, many installations did not follow standard installation guidelines. The authors recommended that proper QC should be implemented by the frequent monitoring of each and every diagnostic X-ray installation.","PeriodicalId":14613,"journal":{"name":"Iranian Journal of Medical Physics","volume":"17 1","pages":"273-281"},"PeriodicalIF":0.0000,"publicationDate":"2020-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Iranian Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.22038/IJMP.2019.39307.1519","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Health Professions","Score":null,"Total":0}
引用次数: 1

Abstract

Introduction: Best radiography practice involves operational optimal machine performance, delivering cost-effective healthcare services under appropriate safety conditions for workers and the public. The present study aimed to investigate the safety status of diagnostic X-ray installations in Mizoram, India. Material and Methods: Linearity of time (sec), linearity of current (mA), output reproducibility, table dose (μGy/mAs), peak voltage (kVp) accuracy, and 16 essential safety parameters of 135 X-ray machines were considered in this study. A battery-operated dosimeter and wide-range digital kVp meter were used to measure output radiation and effective peak potential of X-ray tube. Data analysis was performed using SPSS software to obtain the mean, standard deviation, and coefficient of variation. Results: Among different electronic parameters, 59.2% linearity of time, 82.6% linearity of current, 89.7% kVp accuracy, 35.1% output reproducibility, and 92.8% table dose were beyond the acceptable limits. Based on 16 essential safety parameters, it was observed that 98.7% of X-ray machines did not receive proper quality assurance test, 1.9% of the installations employed lead-line patient entrance doors, 46.8% of the machines were operated without any protective barriers and 83.1% of the units were operated without personnel monitoring service. Conclusion: The present study had concluded with more problems than the previous studies in different parts of the world in this regard. Due to the absence of proper quality control (QC) programs, many installations did not follow standard installation guidelines. The authors recommended that proper QC should be implemented by the frequent monitoring of each and every diagnostic X-ray installation.
印度米佐拉姆邦诊断X射线发生器和基本安全参数的计数调查
简介:最佳射线照相实践涉及操作最佳机器性能,在适当的安全条件下为工人和公众提供具有成本效益的医疗服务。本研究旨在调查印度米佐拉姆诊断X射线装置的安全状况。材料和方法:本研究考虑了135台X光机的时间线性(秒)、电流线性(mA)、输出再现性、表剂量(μGy/mAs)、峰值电压(kVp)精度和16个基本安全参数。用电池式剂量计和宽量程数字kVp计测量了X射线管的输出辐射和有效峰值电位。使用SPSS软件进行数据分析,获得平均值、标准差和变异系数。结果:在不同的电子参数中,59.2%的时间线性度、82.6%的电流线性度、89.7%的kVp准确度、35.1%的输出再现性和92.8%的表剂量超出了可接受的范围。根据16个基本安全参数,观察到98.7%的X光机没有接受适当的质量保证测试,1.9%的设备使用了铅线患者入口门,46.8%的机器在没有任何防护屏障的情况下运行,83.1%的设备在没有人员监控服务的情况下运行。结论:本研究在这方面比以往世界各地的研究存在更多的问题。由于缺乏适当的质量控制(QC)程序,许多安装没有遵循标准安装指南。作者建议,应通过频繁监测每个诊断X射线装置来实施适当的QC。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Iranian Journal of Medical Physics
Iranian Journal of Medical Physics Health Professions-Radiological and Ultrasound Technology
CiteScore
1.00
自引率
0.00%
发文量
0
审稿时长
8 weeks
期刊介绍: Iranian Journal of Medical Physics (IJMP) is the official scientific bimonthly publication of the Iranian Association of Medical Physicists. IJMP is an international and multidisciplinary journal, peer review, free of charge publication and open access. This journal devoted to publish Original Papers, Review Articles, Short Communications, Technical Notes, Editorial and Letters to the Editor in the field of “Medical Physics” involving both basic and clinical research. Submissions of manuscript from all countries are welcome and will be reviewed by at least two expert reviewers.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信