Effect of Phantom Size and Tube Voltage on the Size-Conversion Factor for Patient Dose Estimation in Computed Tomography Examinations

Q3 Health Professions
Dwi Adhianto, C. Anam, H. Sutanto, Mohd Hanafi Ali
{"title":"Effect of Phantom Size and Tube Voltage on the Size-Conversion Factor for Patient Dose Estimation in Computed Tomography Examinations","authors":"Dwi Adhianto, C. Anam, H. Sutanto, Mohd Hanafi Ali","doi":"10.22038/IJMP.2019.43059.1649","DOIUrl":null,"url":null,"abstract":"Introduction: This study aimed to establish the conversion factors to normalize the output dose of volumetric computed tomography dose index (CTDIvol) to the patient dose (i.e. size-specific dose estimate (SSDE)) for various phantom diameters and tube voltages. \nMaterial and Methods: In-house cylindrical acrylic phantoms with physical diameters ranging from 8 to 40 cm were developed in this study. Each phantom had a hole in the center and four holes in the peripheral areas. The phantoms were scanned by a Siemens Somatom Definition AS CT Scanner using different tube voltages (i.e. 80, 100, 120, and 140 kVps) and with 200 mAs and 10 mm slice thickness. In addition, the doses in every hole and phantom were measured using a Raysafe X2 CT Sensor. The weighted SSDE (SSDEw) values were calculated using the five holes in every measurement. The size-conversion factors for the body and the head CTDI phantoms were established by dividing the SSDEw for various sizes with the SSDEw at the water-equivalent diameter of 33.90 cm and 16.95 cm, respectively. \nResults: The results revealed that the size-conversion factor exponentially decreased with an increase in the phantom size. It was also found that the size-conversion factor was affected by the tube voltages. Furthermore, the different size-conversion factor between 80 and 140 kVp was more than 15% in very thin and obese patients. \nConclusion: Higher accuracy of the size-specific dose estimation can be achieved considering the impact of the tube voltages beside the size of the patient.","PeriodicalId":14613,"journal":{"name":"Iranian Journal of Medical Physics","volume":"17 1","pages":"282-288"},"PeriodicalIF":0.0000,"publicationDate":"2020-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"4","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Iranian Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.22038/IJMP.2019.43059.1649","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Health Professions","Score":null,"Total":0}
引用次数: 4

Abstract

Introduction: This study aimed to establish the conversion factors to normalize the output dose of volumetric computed tomography dose index (CTDIvol) to the patient dose (i.e. size-specific dose estimate (SSDE)) for various phantom diameters and tube voltages. Material and Methods: In-house cylindrical acrylic phantoms with physical diameters ranging from 8 to 40 cm were developed in this study. Each phantom had a hole in the center and four holes in the peripheral areas. The phantoms were scanned by a Siemens Somatom Definition AS CT Scanner using different tube voltages (i.e. 80, 100, 120, and 140 kVps) and with 200 mAs and 10 mm slice thickness. In addition, the doses in every hole and phantom were measured using a Raysafe X2 CT Sensor. The weighted SSDE (SSDEw) values were calculated using the five holes in every measurement. The size-conversion factors for the body and the head CTDI phantoms were established by dividing the SSDEw for various sizes with the SSDEw at the water-equivalent diameter of 33.90 cm and 16.95 cm, respectively. Results: The results revealed that the size-conversion factor exponentially decreased with an increase in the phantom size. It was also found that the size-conversion factor was affected by the tube voltages. Furthermore, the different size-conversion factor between 80 and 140 kVp was more than 15% in very thin and obese patients. Conclusion: Higher accuracy of the size-specific dose estimation can be achieved considering the impact of the tube voltages beside the size of the patient.
体模尺寸和管电压对计算机断层扫描中患者剂量估计的尺寸转换因子的影响
引言:本研究旨在建立转换因子,以将体积计算机断层扫描剂量指数(CTDIvol)的输出剂量标准化为不同体模直径和管电压的患者剂量(即尺寸特异性剂量估计(SSDE))。材料和方法:本研究开发了物理直径为8至40厘米的室内圆柱形丙烯酸模型。每个模型的中心有一个洞,外围区域有四个洞。通过Siemens Somatom Definition AS CT扫描仪使用不同的管电压(即80、100、120和140kVps)和200mAs和10mm的切片厚度扫描体模。此外,使用Raysafe X2 CT传感器测量每个孔和体模中的剂量。在每次测量中使用五个孔来计算加权的SSDE(SSDEw)值。通过将不同尺寸的SSDEw与水当量直径分别为33.90厘米和16.95厘米的SSDEw分开,建立了身体和头部CTDI模型的尺寸转换因子。结果:结果显示,随着体模尺寸的增加,尺寸转换因子呈指数下降。还发现尺寸转换因子受管电压的影响。此外,在非常瘦和肥胖的患者中,80和140kVp之间的不同尺寸转换因子超过15%。结论:考虑到管电压对患者体型的影响,可以获得更高的尺寸特异性剂量估计精度。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Iranian Journal of Medical Physics
Iranian Journal of Medical Physics Health Professions-Radiological and Ultrasound Technology
CiteScore
1.00
自引率
0.00%
发文量
0
审稿时长
8 weeks
期刊介绍: Iranian Journal of Medical Physics (IJMP) is the official scientific bimonthly publication of the Iranian Association of Medical Physicists. IJMP is an international and multidisciplinary journal, peer review, free of charge publication and open access. This journal devoted to publish Original Papers, Review Articles, Short Communications, Technical Notes, Editorial and Letters to the Editor in the field of “Medical Physics” involving both basic and clinical research. Submissions of manuscript from all countries are welcome and will be reviewed by at least two expert reviewers.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信