Why Does Vascular Access Dysfunction Occur despite Brachial Artery Blood Flow Being Higher than Preset Blood Flow?

Jun-ichi Ono, Takushi Oiwa, Y. Ogasawara, S. Mochizuki
{"title":"Why Does Vascular Access Dysfunction Occur despite Brachial Artery Blood Flow Being Higher than Preset Blood Flow?","authors":"Jun-ichi Ono, Takushi Oiwa, Y. Ogasawara, S. Mochizuki","doi":"10.3390/kidneydial2010005","DOIUrl":null,"url":null,"abstract":"Background: In recent years, many reports have investigated the usefulness of brachial artery blood flow (BAF) measured by ultrasonography as an evaluation index for the vascular access (VA) stenosis of hemodialysis patients. However, the mechanism of VA dysfunction, despite BAF being higher than the preset blood flow, has not been clarified to date. Methods: The relationship between actual blood-removal flow and recirculation rate with decreasing VA flow was examined using a VA flow path model and pure water as a model fluid. The blood-flow rate was set at 180 mL/min, and the set VA flow rate was lowered stepwise from 350 to 50 mL/min. VA flow rate, blood-removal flow rate, and flow waveform measured between two needle-puncture sites were recorded, and then the actual blood-removal flow rate and recirculation rate were calculated. Results: Recirculation was observed at a VA flow rate < 300 mL/min. The recirculation was due to the VA flow rate, which was transiently reduced to the level below the blood-removal flow rate, resulting in backflow. In contrast, no decrease in the actual blood-removal flow rate was observed. Conclusion: It is suggested that the mechanism of the VA dysfunction, despite the BAF being higher than the preset blood-flow rate, was due to the diastolic BAF being lower than the blood-removal flow rate.","PeriodicalId":74038,"journal":{"name":"Kidney and dialysis","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2022-01-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Kidney and dialysis","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.3390/kidneydial2010005","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

Background: In recent years, many reports have investigated the usefulness of brachial artery blood flow (BAF) measured by ultrasonography as an evaluation index for the vascular access (VA) stenosis of hemodialysis patients. However, the mechanism of VA dysfunction, despite BAF being higher than the preset blood flow, has not been clarified to date. Methods: The relationship between actual blood-removal flow and recirculation rate with decreasing VA flow was examined using a VA flow path model and pure water as a model fluid. The blood-flow rate was set at 180 mL/min, and the set VA flow rate was lowered stepwise from 350 to 50 mL/min. VA flow rate, blood-removal flow rate, and flow waveform measured between two needle-puncture sites were recorded, and then the actual blood-removal flow rate and recirculation rate were calculated. Results: Recirculation was observed at a VA flow rate < 300 mL/min. The recirculation was due to the VA flow rate, which was transiently reduced to the level below the blood-removal flow rate, resulting in backflow. In contrast, no decrease in the actual blood-removal flow rate was observed. Conclusion: It is suggested that the mechanism of the VA dysfunction, despite the BAF being higher than the preset blood-flow rate, was due to the diastolic BAF being lower than the blood-removal flow rate.
为什么尽管肱动脉血流量高于预设血流量,血管通路仍会出现功能障碍?
背景:近年来,许多报道探讨了超声测量肱动脉血流(BAF)作为血液透析患者血管通路(VA)狭窄的评价指标的有效性。然而,尽管BAF高于预设血流量,但VA功能障碍的机制迄今尚未明确。方法:采用VA流道模型,以纯水为模型流体,考察实际血流量和再循环速率随VA流量减小的关系。血流量设定为180 mL/min,设定的VA流量由350 mL/min逐步降低至50 mL/min。记录VA流量、取血流量、两针间测量的血流波形,计算实际取血流量和再循环率。结果:在VA流速< 300 mL/min时出现再循环。再循环是由于VA流量,它暂时降低到低于除血流量的水平,导致回流。相比之下,没有观察到实际血液清除流量的减少。结论:提示尽管BAF高于预设血流量,但导致VA功能障碍的机制是由于舒张BAF低于血流量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信