Modernization of the flash point tester PE-TVZ for determining fire safety of electrolyte systems of energy-intensive batteries

V. Y. Mishinkin, G. Kamalova, E. Kuzmina, V. Kolosnitsyn
{"title":"Modernization of the flash point tester PE-TVZ for determining fire safety of electrolyte systems of energy-intensive batteries","authors":"V. Y. Mishinkin, G. Kamalova, E. Kuzmina, V. Kolosnitsyn","doi":"10.18500/1608-4039-2023-23-2-80-86","DOIUrl":null,"url":null,"abstract":"The flash point tester PE-TVZ was modernized. The sample volume was reduced from 70 to 5 ml. Mixing of the condensed and gas-vapor phases was done simultaneously. The mercury thermometer was replaced by an electronic one. The correctness of the flash point measurement by the modernized device was tested on the samples with the flash point in the range of 25–170°C. The correlation coefficient between the measured flash points of the samples using the flash point tester PE-TVZ and the modernized one was 99.6%, the discrepancy between the measured values not exceeding 1°C. Modernization of the tester made it possible to reduce the amount of samples used by a factor of 14 while maintaining the accuracy of measuring the flash point.","PeriodicalId":34569,"journal":{"name":"Elektrokhimicheskaia energetika","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2023-06-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Elektrokhimicheskaia energetika","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.18500/1608-4039-2023-23-2-80-86","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The flash point tester PE-TVZ was modernized. The sample volume was reduced from 70 to 5 ml. Mixing of the condensed and gas-vapor phases was done simultaneously. The mercury thermometer was replaced by an electronic one. The correctness of the flash point measurement by the modernized device was tested on the samples with the flash point in the range of 25–170°C. The correlation coefficient between the measured flash points of the samples using the flash point tester PE-TVZ and the modernized one was 99.6%, the discrepancy between the measured values not exceeding 1°C. Modernization of the tester made it possible to reduce the amount of samples used by a factor of 14 while maintaining the accuracy of measuring the flash point.
用于测定能量密集型电池电解质系统防火安全性的闪点测试仪PE-TVZ的现代化
对闪点测试仪PE-TVZ进行了现代化改造。样品体积从70毫升减少到5毫升。同时进行冷凝相和气相的混合。水银温度计被电子温度计取代了。对闪点在25–170°C范围内的样品进行了现代化设备闪点测量的正确性测试。使用闪点测试仪PE-TVZ测量的样品闪点与现代化的闪点之间的相关系数为99.6%,测量值之间的差异不超过1°C。测试仪的现代化使得可以将使用的样本量减少14倍,同时保持闪点测量的准确性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
6
审稿时长
12 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信