Diagnostic Accuracy of Cone-Beam Computed Tomography at Different Tube Voltages for Vertical Root Fractures in Endodontically Treated Teeth with Metallic Posts

IF 0.6 Q4 DENTISTRY, ORAL SURGERY & MEDICINE
Hoorieh Bashizadehfakhar, B. Bolhari, A. Shamshiri, S. Amini, P. Ranji
{"title":"Diagnostic Accuracy of Cone-Beam Computed Tomography at Different Tube Voltages for Vertical Root Fractures in Endodontically Treated Teeth with Metallic Posts","authors":"Hoorieh Bashizadehfakhar, B. Bolhari, A. Shamshiri, S. Amini, P. Ranji","doi":"10.4103/denthyp.denthyp_107_21","DOIUrl":null,"url":null,"abstract":"Introduction: The study evaluated the effectiveness of different tube voltages of the cone-beam computed tomography (CBCT) device on the diagnosis of vertical root fractures in teeth with single root canals and posts. Materials and methods: Sixty single-canal and single-rooted extracted human mandibular premolars were selected, all of which were intact and contained no fractures. After disinfection, teeth crowns were sectioned and teeth canals were prepped and obturated by a method similar to the rotary technique. Root fractures were created in half of the teeth by applying controlled buccolingual force. Prefabricated metal posts were subsequently placed in all teeth using AH26 sealer. All teeth, both intact and fractured, were mounted in bovine ribs and X-rayed randomly using a CBCT device (Planmeca Romexis, Helsinki, Finland) with three tube voltages (min = 66 kVp, mid = 80 kVp, and max = 92 kVp). Three experienced maxillofacial radiologists observed and commented on the images. Research data were analyzed using SPSS-25 software. Results: Images taken by kVp min were removed from further statistical analysis due to a lack of diagnostic information. For mid (80) and max kVp (92), the sensitivity values were 86.6% and 60 % (P = 0.04), and the specificity values were 33.3% and 30%, respectively (P = 0.45). Overall, observers’ opinions were more alike concerning images taken with kVp mid than kVp max. Conclusion: Apparently, images taken by a tube voltage of 80 kVp are more efficient in the diagnosis of root fractures.","PeriodicalId":43354,"journal":{"name":"Dental Hypotheses","volume":"12 1","pages":"132 - 138"},"PeriodicalIF":0.6000,"publicationDate":"2021-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Dental Hypotheses","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4103/denthyp.denthyp_107_21","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"DENTISTRY, ORAL SURGERY & MEDICINE","Score":null,"Total":0}
引用次数: 1

Abstract

Introduction: The study evaluated the effectiveness of different tube voltages of the cone-beam computed tomography (CBCT) device on the diagnosis of vertical root fractures in teeth with single root canals and posts. Materials and methods: Sixty single-canal and single-rooted extracted human mandibular premolars were selected, all of which were intact and contained no fractures. After disinfection, teeth crowns were sectioned and teeth canals were prepped and obturated by a method similar to the rotary technique. Root fractures were created in half of the teeth by applying controlled buccolingual force. Prefabricated metal posts were subsequently placed in all teeth using AH26 sealer. All teeth, both intact and fractured, were mounted in bovine ribs and X-rayed randomly using a CBCT device (Planmeca Romexis, Helsinki, Finland) with three tube voltages (min = 66 kVp, mid = 80 kVp, and max = 92 kVp). Three experienced maxillofacial radiologists observed and commented on the images. Research data were analyzed using SPSS-25 software. Results: Images taken by kVp min were removed from further statistical analysis due to a lack of diagnostic information. For mid (80) and max kVp (92), the sensitivity values were 86.6% and 60 % (P = 0.04), and the specificity values were 33.3% and 30%, respectively (P = 0.45). Overall, observers’ opinions were more alike concerning images taken with kVp mid than kVp max. Conclusion: Apparently, images taken by a tube voltage of 80 kVp are more efficient in the diagnosis of root fractures.
不同管电压锥束ct对金属桩牙根管治疗牙根垂直骨折诊断的准确性
引言:本研究评估了锥束计算机断层扫描(CBCT)设备的不同管电压在诊断单根管和单桩牙齿垂直根骨折方面的有效性。材料与方法:选择60颗完整、无骨折的人下颌前磨牙单根、单管拔除。消毒后,用类似于旋转技术的方法对牙冠进行切片,并制备和堵塞牙管。通过施加可控的颊舌力,在一半的牙齿中产生了牙根骨折。随后使用AH26密封剂在所有牙齿上放置预制金属桩。将所有完整和断裂的牙齿安装在牛肋骨中,并使用CBCT设备(Planmeca Romexis,Helsinki,Finland)随机进行X光检查,该设备具有三个管电压(min = 66 kVp,中间 = 80 kVp和最大值 = 92kVp)。三位经验丰富的颌面放射科医生对图像进行了观察和评论。研究数据采用SPSS-25软件进行分析。结果:由于缺乏诊断信息,kVp-min拍摄的图像从进一步的统计分析中删除。对于中(80)和最大kVp(92),灵敏度分别为86.6%和60%(P = 特异性分别为33.3%和30%(P = 0.45)。总体而言,观察者对使用kVp mid拍摄的图像的看法比使用kVp max拍摄的图像更相似。结论:显然,使用80 kVp的管电压拍摄的图像在诊断根部骨折方面更有效。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Dental Hypotheses
Dental Hypotheses DENTISTRY, ORAL SURGERY & MEDICINE-
CiteScore
1.00
自引率
0.00%
发文量
16
审稿时长
21 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信