Reduction of Insertion Torque on Orthodontic Mini-screw Implant by means of Reduced Friction

Q3 Health Professions
C. Jongwannasiri, T. Charasseangpaisarn, S. Watanabe
{"title":"Reduction of Insertion Torque on Orthodontic Mini-screw Implant by means of Reduced Friction","authors":"C. Jongwannasiri, T. Charasseangpaisarn, S. Watanabe","doi":"10.12982/cmujns.2021.060","DOIUrl":null,"url":null,"abstract":"In this article, the reduction of insertion torque on the orthodontic mini-screw implants (OMI) was studied. Three types of diamond-like carbon (DLC) films (DLC, Si-DLC, and F-DLC) were deposited on the OMIs by plasma-based ion implantation technique. The maximum insertion torque (MIT) value was measured using a physiodispenser during insertion on a mandibular pig jaw. Ten OMIs of each condition were used to evaluate the MIT values compared to the Ti-6Al-4V typed OMI. The statistical analysis of data was analyzed by One-way analysis of variance (ANOVA) and Tukey HSD. The results indicated that the Ti-6Al-4V showed the highest MIT but not significantly different from the F-DLC (P >0.05). DLC and Si-DLC showed a significant reduction in MIT comparing to the Ti-6Al-4V (P <0.05). However, the Si-DLC showed the lowest MIT among groups (P <0.05). This phenomenon was due to the decreasing friction coefficient during OMI's insertion on a mandibular pig jaw. Therefore, the DLC coating, especially Si-DLC, could reduce the MIT value of OMI. It is thought that the lowering friction coefficient yields the lowering of MIT value.\n\n \n\nKeywords: Diamond-like carbon, Friction, Insertion torque, Mini-screw implant","PeriodicalId":10049,"journal":{"name":"Chiang Mai University journal of natural sciences","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2021-06-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Chiang Mai University journal of natural sciences","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.12982/cmujns.2021.060","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Health Professions","Score":null,"Total":0}
引用次数: 1

Abstract

In this article, the reduction of insertion torque on the orthodontic mini-screw implants (OMI) was studied. Three types of diamond-like carbon (DLC) films (DLC, Si-DLC, and F-DLC) were deposited on the OMIs by plasma-based ion implantation technique. The maximum insertion torque (MIT) value was measured using a physiodispenser during insertion on a mandibular pig jaw. Ten OMIs of each condition were used to evaluate the MIT values compared to the Ti-6Al-4V typed OMI. The statistical analysis of data was analyzed by One-way analysis of variance (ANOVA) and Tukey HSD. The results indicated that the Ti-6Al-4V showed the highest MIT but not significantly different from the F-DLC (P >0.05). DLC and Si-DLC showed a significant reduction in MIT comparing to the Ti-6Al-4V (P <0.05). However, the Si-DLC showed the lowest MIT among groups (P <0.05). This phenomenon was due to the decreasing friction coefficient during OMI's insertion on a mandibular pig jaw. Therefore, the DLC coating, especially Si-DLC, could reduce the MIT value of OMI. It is thought that the lowering friction coefficient yields the lowering of MIT value. Keywords: Diamond-like carbon, Friction, Insertion torque, Mini-screw implant
减小摩擦力降低正畸微型螺钉种植体的插入力矩
本文研究了微型正畸螺钉种植体(OMI)的插入力矩减小问题。采用等离子体离子注入技术在OMI上沉积了三种类型的类金刚石碳(DLC)膜(DLC、Si-DLC和F-DLC)。最大插入扭矩(MIT)值是在猪下颚插入过程中使用物理分配器测量的。使用每种条件的10个OMI来评估与Ti-6Al-4V型OMI相比的MIT值。数据的统计分析采用单因素方差分析(ANOVA)和Tukey HSD。结果表明,Ti-6Al-4V显示出最高的MIT,但与F-DLC没有显著差异(P>0.05)。DLC和Si-DLC显示出与Ti-6Al-IV相比显著降低的MIT(P<0.05)。然而,Si-DLC在各组中显示出最低的MIT(P<0.05)。这一现象是由于OMI插入猪下颌时摩擦系数降低所致。因此,DLC涂层,尤其是Si-DLC,可以降低OMI的MIT值。认为摩擦系数的降低导致MIT值的降低。关键词:类金刚石,摩擦,插入扭矩,微型螺钉植入
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Chiang Mai University journal of natural sciences
Chiang Mai University journal of natural sciences Health Professions-Health Professions (miscellaneous)
CiteScore
1.70
自引率
0.00%
发文量
67
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信