The Normative Data for Sensorineural Acuity Level (SAL) Test among Young Adults: Comparisons Between B71 and B81 Bone Transducers

M. Awang, Ainul Hazwani Mohd Don, Rosdan Salim, M. D. Sul'ain, M. Zakaria
{"title":"The Normative Data for Sensorineural Acuity Level (SAL) Test among Young Adults: Comparisons Between B71 and B81 Bone Transducers","authors":"M. Awang, Ainul Hazwani Mohd Don, Rosdan Salim, M. D. Sul'ain, M. Zakaria","doi":"10.6000/1929-6029.2021.10.15","DOIUrl":null,"url":null,"abstract":"Introduction: The sensorineural acuity level (SAL) test is a valuable clinical assessment for hearing diagnosis, especially when the results of pure tone audiometry (PTA) are incomplete or questionable. Nevertheless, the literature on the SAL test is extremely limited and more research efforts are warranted to promote the application of this test in clinical settings. Essentially, the aim of the present study was to compare the normative data for the SAL test between two different bone vibrators (Radioear B71 and Radioear B81) among Malaysian participants. Materials and Methods: Forty-two healthy Malaysian adults (aged 19 to 27 years) participated in this repeated-measures study. The PTA and SAL test was carried out according to the recommended protocols. The SAL normative data produced by the two bone transducers were then compared at different frequencies. Paired t-test and Bayesian inference were used for data analyses. Results: The normative data for the SAL test gathered in the present study were descriptively higher than those reported among Caucasian adults. Relative to the Radioear B81 bone vibrator, the SAL normative data were significantly higher for the Radioear B71 bone transducer at all tested frequencies (p < 0.001, BF10 ≥ 48.67). Conclusions: The SAL normative data established by the two bone vibrators are statistically different. This information would be beneficial to provide clinicians with appropriate normative data when performing the SAL test on Malaysian patients.","PeriodicalId":73480,"journal":{"name":"International journal of statistics in medical research","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2021-11-29","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International journal of statistics in medical research","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6000/1929-6029.2021.10.15","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1

Abstract

Introduction: The sensorineural acuity level (SAL) test is a valuable clinical assessment for hearing diagnosis, especially when the results of pure tone audiometry (PTA) are incomplete or questionable. Nevertheless, the literature on the SAL test is extremely limited and more research efforts are warranted to promote the application of this test in clinical settings. Essentially, the aim of the present study was to compare the normative data for the SAL test between two different bone vibrators (Radioear B71 and Radioear B81) among Malaysian participants. Materials and Methods: Forty-two healthy Malaysian adults (aged 19 to 27 years) participated in this repeated-measures study. The PTA and SAL test was carried out according to the recommended protocols. The SAL normative data produced by the two bone transducers were then compared at different frequencies. Paired t-test and Bayesian inference were used for data analyses. Results: The normative data for the SAL test gathered in the present study were descriptively higher than those reported among Caucasian adults. Relative to the Radioear B81 bone vibrator, the SAL normative data were significantly higher for the Radioear B71 bone transducer at all tested frequencies (p < 0.001, BF10 ≥ 48.67). Conclusions: The SAL normative data established by the two bone vibrators are statistically different. This information would be beneficial to provide clinicians with appropriate normative data when performing the SAL test on Malaysian patients.
B71和B81骨传感器在青年人感觉神经敏锐度水平(SAL)测试中的规范性数据比较
引言:感音神经敏锐度水平(SAL)测试是一种有价值的听力诊断临床评估,尤其是当纯音听力计(PTA)的结果不完整或有问题时。然而,关于SAL测试的文献极其有限,需要更多的研究工作来促进该测试在临床环境中的应用。本质上,本研究的目的是比较马来西亚参与者中两种不同骨振子(Radioear B71和Radioear B81)之间SAL测试的标准数据。材料和方法:42名健康的马来西亚成年人(年龄19至27岁)参与了这项重复测量研究。PTA和SAL测试根据推荐方案进行。然后比较两个骨换能器在不同频率下产生的SAL标准数据。数据分析采用配对t检验和贝叶斯推理。结果:本研究中收集的SAL测试的标准数据在描述上高于在高加索成年人中报告的数据。与Radioear B81骨振子相比,Radioear B71骨振器在所有测试频率下的SAL标准数据均显著较高(p<0.001,BF10≥48.67)。在对马来西亚患者进行SAL测试时,这些信息将有助于为临床医生提供适当的规范性数据。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
0.40
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信