Evaluation of Six-Dimensional Cranial Target Positioning Accuracy in Two Different Immobilization Methods Using Exactrac System

Q3 Health Professions
T. Kasirajan, P. Ganesan, R. Kesavan, N. ArunaiNambiRaj, K. Senthilnathan, P. Rameshbabu
{"title":"Evaluation of Six-Dimensional Cranial Target Positioning Accuracy in Two Different Immobilization Methods Using Exactrac System","authors":"T. Kasirajan, P. Ganesan, R. Kesavan, N. ArunaiNambiRaj, K. Senthilnathan, P. Rameshbabu","doi":"10.22038/IJMP.2019.35652.1448","DOIUrl":null,"url":null,"abstract":"Introduction: The aim of this study was to determine the accuracy of two different immobilization methods in patient positioning in cranial radiotherapy. The six-dimensional (6D) target localization accuracy of using a dedicated stereotactic mask was compared with that of a conventional head mask by the ExacTrac system. \nMaterial and Methods: A total of 56 patients with cranial lesions were included in this study (26 patients with a dedicated stereotactic mask and 30 subjects with a conventional head mask). The ExacTrac image-guided positioning system was utilized to obtain daily translational and rotational patient positioning displacement from the intended position. The 6D setup data was analyzed to obtain population mean, systematic and random errors, and three-dimensional (3D) vector shifts in all the patients. \nResults: The population mean values of setup errors were comparable with both immobilization systems; however, the spread as indicated by population systematic and population random errors was more in the use of a conventional head mask. The mean values of the 3D vector shifts were 2.09±1.00 and 4.51±3.38 mm with the use of a dedicated stereotactic mask and conventional head mask, respectively. The frequency distribution of maximum rotational deviation and statistical analysis demonstrated a significant difference in immobilization accuracy between stereotactic immobilization and 3-clamp immobilization (P<0.05). \nConclusion: The results revealed that there was a significant reduction in target positioning errors with a dedicated stereotactic mask, compared to that with a conventional cranial mask. Furthermore, a dedicated stereotactic mask is required to keep rotational deviations within system correctable limits.","PeriodicalId":14613,"journal":{"name":"Iranian Journal of Medical Physics","volume":"17 1","pages":"308-315"},"PeriodicalIF":0.0000,"publicationDate":"2020-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Iranian Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.22038/IJMP.2019.35652.1448","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Health Professions","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: The aim of this study was to determine the accuracy of two different immobilization methods in patient positioning in cranial radiotherapy. The six-dimensional (6D) target localization accuracy of using a dedicated stereotactic mask was compared with that of a conventional head mask by the ExacTrac system. Material and Methods: A total of 56 patients with cranial lesions were included in this study (26 patients with a dedicated stereotactic mask and 30 subjects with a conventional head mask). The ExacTrac image-guided positioning system was utilized to obtain daily translational and rotational patient positioning displacement from the intended position. The 6D setup data was analyzed to obtain population mean, systematic and random errors, and three-dimensional (3D) vector shifts in all the patients. Results: The population mean values of setup errors were comparable with both immobilization systems; however, the spread as indicated by population systematic and population random errors was more in the use of a conventional head mask. The mean values of the 3D vector shifts were 2.09±1.00 and 4.51±3.38 mm with the use of a dedicated stereotactic mask and conventional head mask, respectively. The frequency distribution of maximum rotational deviation and statistical analysis demonstrated a significant difference in immobilization accuracy between stereotactic immobilization and 3-clamp immobilization (P<0.05). Conclusion: The results revealed that there was a significant reduction in target positioning errors with a dedicated stereotactic mask, compared to that with a conventional cranial mask. Furthermore, a dedicated stereotactic mask is required to keep rotational deviations within system correctable limits.
基于Exactrac系统的两种不同固定方法对六维颅骨目标定位精度的评价
引言:本研究的目的是确定两种不同固定方法在颅骨放疗中患者定位的准确性。通过ExacTrac系统将使用专用立体定向掩模的六维(6D)目标定位精度与传统头部掩模的精度进行了比较。材料和方法:本研究共纳入56名颅骨病变患者(26名患者使用专用立体定向面罩,30名受试者使用常规面罩)。ExacTrac图像引导定位系统用于获得患者从预期位置的每日平移和旋转定位位移。对6D设置数据进行分析,以获得所有患者的总体平均值、系统和随机误差以及三维(3D)矢量偏移。结果:两种固定化系统的设置误差的总体平均值具有可比性;然而,总体系统误差和总体随机误差表明,在使用传统口罩时传播更多。使用专用立体定向面罩和传统面罩时,3D矢量偏移的平均值分别为2.09±1.00和4.51±3.38mm。最大旋转偏差的频率分布和统计分析表明,立体定向固定和三灯固定在固定精度上有显著差异(P<0.05),与传统的颅骨面罩相比。此外,需要一个专用的立体定向掩模来将旋转偏差保持在系统可校正的限度内。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Iranian Journal of Medical Physics
Iranian Journal of Medical Physics Health Professions-Radiological and Ultrasound Technology
CiteScore
1.00
自引率
0.00%
发文量
0
审稿时长
8 weeks
期刊介绍: Iranian Journal of Medical Physics (IJMP) is the official scientific bimonthly publication of the Iranian Association of Medical Physicists. IJMP is an international and multidisciplinary journal, peer review, free of charge publication and open access. This journal devoted to publish Original Papers, Review Articles, Short Communications, Technical Notes, Editorial and Letters to the Editor in the field of “Medical Physics” involving both basic and clinical research. Submissions of manuscript from all countries are welcome and will be reviewed by at least two expert reviewers.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信