{"title":"A Novel Tungsten Coil Electrothermal Vaporizer With Composite Structure Coupled With Dielectric Barrier Discharge Optical Emission Spectrometer For Direct Determination Of Trace Mercury","authors":"Jixin Liu, Xuefei Mao","doi":"10.46770/as.2023.089","DOIUrl":null,"url":null,"abstract":": In this study, a small platinized tungsten coil (TC) with a composite structure was utilized for the first time in a liquid sampling electrothermal vaporizer (ETV) setup. A novel direct sampling mercury (Hg) analyzer coupled with a dielectric barrier discharge (DBD) and a miniature fiber optic spectrometer (FOS) as an optical emission spectrometer was fabricated for the sensitive determination of Hg in environmental water. This ETV setup displayed a maximum sample volume of 100 μL and fulfilled the task of sequestering Hg during the dehydration and ashing processes under high temperature due to the formation of a platinum (Pt) and Hg amalgam. Under optimized conditions, the limit of detection (LOD) of Hg was 0.1 μg L -1 for a 100 μL liquid sample and the relative standard deviation (RSD) of 11 repeated measurements of Hg standard solution was 3.2%; the linearity ( R 2 ) was > 0.999 in the range of 0.4-100 μg L -1 . The results for the GBW08063 sample (9.89 μg L -1 ) agreed with the certified values of certified reference materials (CRMs) (10 μg L -1 ), and the spiked recoveries were 94-100%, which proved favorable analytical accuracy and precision. The total analysis time was observed to be less than 3 min, and this small-sized ETV-DBD-OES system consumes only ~100 W of power while weighting less than 12 kg. Therefore, the proposed ETV-DBD-OES method demonstrates simplicity, suitable sensitivity, precision, and robustness, with promising application in the field for rapid water analysis for Hg.","PeriodicalId":8642,"journal":{"name":"Atomic Spectroscopy","volume":" ","pages":""},"PeriodicalIF":3.4000,"publicationDate":"2023-04-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Atomic Spectroscopy","FirstCategoryId":"92","ListUrlMain":"https://doi.org/10.46770/as.2023.089","RegionNum":2,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"SPECTROSCOPY","Score":null,"Total":0}
引用次数: 0
Abstract
: In this study, a small platinized tungsten coil (TC) with a composite structure was utilized for the first time in a liquid sampling electrothermal vaporizer (ETV) setup. A novel direct sampling mercury (Hg) analyzer coupled with a dielectric barrier discharge (DBD) and a miniature fiber optic spectrometer (FOS) as an optical emission spectrometer was fabricated for the sensitive determination of Hg in environmental water. This ETV setup displayed a maximum sample volume of 100 μL and fulfilled the task of sequestering Hg during the dehydration and ashing processes under high temperature due to the formation of a platinum (Pt) and Hg amalgam. Under optimized conditions, the limit of detection (LOD) of Hg was 0.1 μg L -1 for a 100 μL liquid sample and the relative standard deviation (RSD) of 11 repeated measurements of Hg standard solution was 3.2%; the linearity ( R 2 ) was > 0.999 in the range of 0.4-100 μg L -1 . The results for the GBW08063 sample (9.89 μg L -1 ) agreed with the certified values of certified reference materials (CRMs) (10 μg L -1 ), and the spiked recoveries were 94-100%, which proved favorable analytical accuracy and precision. The total analysis time was observed to be less than 3 min, and this small-sized ETV-DBD-OES system consumes only ~100 W of power while weighting less than 12 kg. Therefore, the proposed ETV-DBD-OES method demonstrates simplicity, suitable sensitivity, precision, and robustness, with promising application in the field for rapid water analysis for Hg.
期刊介绍:
The ATOMIC SPECTROSCOPY is a peer-reviewed international journal started in 1962 by Dr. Walter Slavin and now is published by Atomic Spectroscopy Press Limited (ASPL). It is intended for the rapid publication of both original articles and review articles in the fields of AAS, AFS, ICP-OES, ICP-MS, GD-MS, TIMS, SIMS, AMS, LIBS, XRF and related techniques. Manuscripts dealing with (i) instrumentation & fundamentals, (ii) methodology development & applications, and (iii) standard reference materials (SRMs) development can be submitted for publication.