Patient-Specific Pre-Treatment VMAT Plan Verification Using Gamma Passing Rates

IF 0.6 Q4 NUCLEAR SCIENCE & TECHNOLOGY
M. Z. Arsalan, M. Kakakhel, M. Shamshad, T. Afridi
{"title":"Patient-Specific Pre-Treatment VMAT Plan Verification Using Gamma Passing Rates","authors":"M. Z. Arsalan, M. Kakakhel, M. Shamshad, T. Afridi","doi":"10.55981/aij.2023.1261","DOIUrl":null,"url":null,"abstract":"Continuous gantry motion, continuous beam modulation, and variable dose rate are used in volumetric modulated arc therapy (VMAT) to obtain highly conformal radiation therapy dose distributions. Several errors during daily radiation therapy treatment can be sources of uncertainties in dose delivery. These errors include monitor unit calculation errors and other human mistakes. Due to the uncertainties in the excessively modulated VMAT plan, the intended dose distribution is not delivered perfectly, leading to a mismatch between the measured and planned dose distributions. This necessitates an extensive and effective quality assurance (QA) program for both machine and patient. In this study, VMAT QA plan verification of 62 head and neck (HN) and 19 prostate cases was done using Octavius 4D setup with its associating VeriSoft gamma analysis software. The plans showed a maximum 3D gamma passing rate with 4 mm/3 % gamma acceptance criteria, i.e., 99.7 % for the HN cancer cases and 99.5 % for the prostate cancer cases. Local gamma analysis was also performed for both regions. Furthermore, 2D and volumetric gamma analyses were also carried out. Gamma analysis with respect to different axis was also carried out. It was known that the transversal axis showed    the highest gamma passing rate in both HN and prostate cases, i.e., 99.17 % and 98.3 %, respectively. The transverse axis came to be a better fit for the planned dose distribution.","PeriodicalId":8647,"journal":{"name":"Atom Indonesia","volume":" ","pages":""},"PeriodicalIF":0.6000,"publicationDate":"2023-04-14","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Atom Indonesia","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.55981/aij.2023.1261","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"NUCLEAR SCIENCE & TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Continuous gantry motion, continuous beam modulation, and variable dose rate are used in volumetric modulated arc therapy (VMAT) to obtain highly conformal radiation therapy dose distributions. Several errors during daily radiation therapy treatment can be sources of uncertainties in dose delivery. These errors include monitor unit calculation errors and other human mistakes. Due to the uncertainties in the excessively modulated VMAT plan, the intended dose distribution is not delivered perfectly, leading to a mismatch between the measured and planned dose distributions. This necessitates an extensive and effective quality assurance (QA) program for both machine and patient. In this study, VMAT QA plan verification of 62 head and neck (HN) and 19 prostate cases was done using Octavius 4D setup with its associating VeriSoft gamma analysis software. The plans showed a maximum 3D gamma passing rate with 4 mm/3 % gamma acceptance criteria, i.e., 99.7 % for the HN cancer cases and 99.5 % for the prostate cancer cases. Local gamma analysis was also performed for both regions. Furthermore, 2D and volumetric gamma analyses were also carried out. Gamma analysis with respect to different axis was also carried out. It was known that the transversal axis showed    the highest gamma passing rate in both HN and prostate cases, i.e., 99.17 % and 98.3 %, respectively. The transverse axis came to be a better fit for the planned dose distribution.
使用伽玛通过率的患者特定治疗前VMAT计划验证
连续龙门架运动、连续光束调制和可变剂量率用于体积调制电弧治疗(VMAT),以获得高度共形的放射治疗剂量分布。日常放射治疗过程中的几个错误可能是剂量传递不确定性的来源。这些错误包括监控器单元计算错误和其他人为错误。由于过度调制的VMAT计划中的不确定性,预期的剂量分布没有完美地传递,导致测量的剂量分布和计划的剂量分布不匹配。这就需要为机器和患者制定广泛有效的质量保证(QA)计划。在本研究中,使用Octavius 4D装置及其相关的VeriSoft伽马分析软件对62例头颈部(HN)和19例前列腺病例进行了VMAT QA计划验证。该计划显示了最大3D伽马通过率,伽马接受标准为4mm/3%,即HN癌症病例为99.7%,癌症病例为99.5%。还对这两个区域进行了局部伽马分析。此外,还进行了二维和体积伽马分析。还对不同轴进行了伽马分析。众所周知,横轴在HN和前列腺病例中显示出最高的γ通过率,分别为99.17%和98.3%。横轴变得更适合计划的剂量分布。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Atom Indonesia
Atom Indonesia NUCLEAR SCIENCE & TECHNOLOGY-
CiteScore
1.00
自引率
0.00%
发文量
20
审稿时长
16 weeks
期刊介绍: The focus of Atom Indonesia is research and development in nuclear science and technology. The scope of this journal covers experimental and analytical research in nuclear science and technology. The topics include nuclear physics, reactor physics, radioactive waste, fuel element, radioisotopes, radiopharmacy, radiation, and neutron scattering, as well as their utilization in agriculture, industry, health, environment, energy, material science and technology, and related fields.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信