Jeremy Genter, G. Rauter, M. Müller-Gerbl, A. Mündermann, D. Baumgartner
{"title":"Musculoskeletal model-based control strategy of an over-actuated glenohumeral simulator to assess joint biomechanics","authors":"Jeremy Genter, G. Rauter, M. Müller-Gerbl, A. Mündermann, D. Baumgartner","doi":"10.1515/auto-2023-0064","DOIUrl":null,"url":null,"abstract":"Abstract Determining the acting shoulder and muscle forces in vivo is very complex. In this study, we developed a control strategy for a glenohumeral simulator for ex vivo experiments that can mimic physiological glenohumeral motion and overcome the problem of over-actuation. The system includes ten muscle portions actuated via cables to induce upper arm motion in three degrees of freedom, including scapula rotation. A real-time optimizer was implemented to handle the over-actuation of the glenohumeral joint while ensuring a minimum of muscle tension. The functionality of the real-time optimizer was also used to simulate different extents of rotator cuff tears. Joint reaction forces were consistent with in vivo measurements. These results demonstrate the feasibility and added value of implementing a real-time optimizer for using in vivo data to drive a shoulder simulator.","PeriodicalId":55437,"journal":{"name":"At-Automatisierungstechnik","volume":"71 1","pages":"505 - 514"},"PeriodicalIF":0.9000,"publicationDate":"2023-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"At-Automatisierungstechnik","FirstCategoryId":"94","ListUrlMain":"https://doi.org/10.1515/auto-2023-0064","RegionNum":4,"RegionCategory":"计算机科学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"AUTOMATION & CONTROL SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Abstract Determining the acting shoulder and muscle forces in vivo is very complex. In this study, we developed a control strategy for a glenohumeral simulator for ex vivo experiments that can mimic physiological glenohumeral motion and overcome the problem of over-actuation. The system includes ten muscle portions actuated via cables to induce upper arm motion in three degrees of freedom, including scapula rotation. A real-time optimizer was implemented to handle the over-actuation of the glenohumeral joint while ensuring a minimum of muscle tension. The functionality of the real-time optimizer was also used to simulate different extents of rotator cuff tears. Joint reaction forces were consistent with in vivo measurements. These results demonstrate the feasibility and added value of implementing a real-time optimizer for using in vivo data to drive a shoulder simulator.
期刊介绍:
Automatisierungstechnik (AUTO) publishes articles covering the entire range of automation technology: development and application of methods, the operating principles, characteristics, and applications of tools and the interrelationships between automation technology and societal developments. The journal includes a tutorial series on "Theory for Users," and a forum for the exchange of viewpoints concerning past, present, and future developments. Automatisierungstechnik is the official organ of GMA (The VDI/VDE Society for Measurement and Automatic Control) and NAMUR (The Process-Industry Interest Group for Automation Technology).
Topics
control engineering
digital measurement systems
cybernetics
robotics
process automation / process engineering
control design
modelling
information processing
man-machine interfaces
networked control systems
complexity management
machine learning
ambient assisted living
automated driving
bio-analysis technology
building automation
factory automation / smart factories
flexible manufacturing systems
functional safety
mechatronic systems.