M. Zabihzadeh, Mahbube Fadaei, S. Hoseini, Sholeh Arvandi, Mohamad Tahmasebi
{"title":"Characterization of Wedge Factors and Dose Distributions in Radiotherapy with Symmetric and Asymmetric Physical Wedged Beams of 6 MV Photon Beam","authors":"M. Zabihzadeh, Mahbube Fadaei, S. Hoseini, Sholeh Arvandi, Mohamad Tahmasebi","doi":"10.22038/IJMP.2019.41275.1595","DOIUrl":null,"url":null,"abstract":"Introduction: Physical wedge by modify photon beam shape and intensity has been utilized in radiotherapy to obtain uniformly dose distribution in tumor site with reduced hot spots. Calculation of dosimetric parameters for both symmetric and asymmetric wedged fields is proved necessary during linear accelerator (Linac) commissioning. The present study aimed to achieve output factors and dose profiles for symmetric and asymmetric wedged fields of 6 MV beams. \nMaterial and Methods: The Siemens PRIMUS Linac head for 6 MV beam was simulated by BEAMnrc and all dose calculations were performed by DOSXYZnrc code. Percentage depth dose (PDD) and profiles for open and wedged (15° and 45°) fields were compared with corresponding measurements. Wedge factors for 10 x 10 cm2 field were obtained as a function of lateral distance as well for half beam wedged fields. \nResults: Based on the results of the present study, the calculated doses were in agreement with the measured data. The output factors on the central axis of symmetric wedged beams decreased to 0.693 and 0.307 for 15˚, and 45˚ wedges. The total photon fluence of 15˚ and 45˚ physical wedged fields reduced to 71.6% and 27.7% of open field, respectively. \nConclusion: The output factor for asymmetric wedged fields was found to be lower than corresponding symmetric open and wedged fields, particularly at field edges. Lack of scattering photons near the half beam edges resulted in dose fall-off in these regions possible to be overestimated by treatment planning system and consequently caused cold spots at target volume.","PeriodicalId":14613,"journal":{"name":"Iranian Journal of Medical Physics","volume":"17 1","pages":"213-219"},"PeriodicalIF":0.0000,"publicationDate":"2020-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Iranian Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.22038/IJMP.2019.41275.1595","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Health Professions","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction: Physical wedge by modify photon beam shape and intensity has been utilized in radiotherapy to obtain uniformly dose distribution in tumor site with reduced hot spots. Calculation of dosimetric parameters for both symmetric and asymmetric wedged fields is proved necessary during linear accelerator (Linac) commissioning. The present study aimed to achieve output factors and dose profiles for symmetric and asymmetric wedged fields of 6 MV beams.
Material and Methods: The Siemens PRIMUS Linac head for 6 MV beam was simulated by BEAMnrc and all dose calculations were performed by DOSXYZnrc code. Percentage depth dose (PDD) and profiles for open and wedged (15° and 45°) fields were compared with corresponding measurements. Wedge factors for 10 x 10 cm2 field were obtained as a function of lateral distance as well for half beam wedged fields.
Results: Based on the results of the present study, the calculated doses were in agreement with the measured data. The output factors on the central axis of symmetric wedged beams decreased to 0.693 and 0.307 for 15˚, and 45˚ wedges. The total photon fluence of 15˚ and 45˚ physical wedged fields reduced to 71.6% and 27.7% of open field, respectively.
Conclusion: The output factor for asymmetric wedged fields was found to be lower than corresponding symmetric open and wedged fields, particularly at field edges. Lack of scattering photons near the half beam edges resulted in dose fall-off in these regions possible to be overestimated by treatment planning system and consequently caused cold spots at target volume.
通过改变光子束形状和强度的物理楔在放射治疗中得到均匀的剂量分布在肿瘤部位,减少了热点。在直线加速器调试过程中,计算对称和非对称楔形场的剂量学参数是必要的。本研究旨在获得对称和非对称6毫伏光束楔形场的输出因子和剂量分布。材料与方法:采用BEAMnrc软件模拟西门子PRIMUS直线磁头的6 MV束流,采用DOSXYZnrc程序进行剂量计算。对开放和楔形(15°和45°)场的百分比深度剂量(PDD)和剖面进行了比较。得到了10 x 10 cm2场的楔形因子作为侧向距离的函数,以及半束楔形场。结果:根据本研究的结果,计算出的剂量与实测数据基本一致。对称楔型梁在15°和45°楔型梁中轴线上的输出系数分别为0.693和0.307。15˚和45˚物理楔形场的总光子通量分别降低到开放场的71.6%和27.7%。结论:不对称楔形场的输出因子低于相应的对称开放和楔形场,特别是在场边缘。半束边缘附近缺乏散射光子导致这些区域的剂量衰减,可能被治疗计划系统高估,从而导致目标体积上的冷点。
期刊介绍:
Iranian Journal of Medical Physics (IJMP) is the official scientific bimonthly publication of the Iranian Association of Medical Physicists. IJMP is an international and multidisciplinary journal, peer review, free of charge publication and open access. This journal devoted to publish Original Papers, Review Articles, Short Communications, Technical Notes, Editorial and Letters to the Editor in the field of “Medical Physics” involving both basic and clinical research. Submissions of manuscript from all countries are welcome and will be reviewed by at least two expert reviewers.