Finite Element Predictions of Sutured and Coupled Microarterial Anastomoses

IF 0.8 Q4 ENGINEERING, BIOMEDICAL
R. Wain, Nicolas J Gaskell, A. Fsadni, J. Francis, J. Whitty
{"title":"Finite Element Predictions of Sutured and Coupled Microarterial Anastomoses","authors":"R. Wain, Nicolas J Gaskell, A. Fsadni, J. Francis, J. Whitty","doi":"10.14326/ABE.8.63","DOIUrl":null,"url":null,"abstract":"Simulation using computational methods is well-established for investigating mechanical and haemodynamic properties of blood vessels, however few groups have applied this technology to microvascular anastomoses. This study, for the first time, employs analytic and numeric models of sutured and coupled mi-croarterial anastomoses to evaluate the elastic and failure properties of these techniques in realistic geometries using measured arterial waveforms. Computational geometries were created of pristine microvessels and mi-croarterial anastomoses, performed using sutures and a coupling device. Vessel wall displacement, stress, and strain distributions were predicted for each anastomotic technique using finite element analysis (FEA) software in both static and transient simulations. This study focussed on mechanical properties of the anastomosis im-\r\nmediately after surgery, as failure is most likely in the early post-operative period. Comparisons were also drawn between stress distributions seen in analogous non-compliant simulations. The maximum principal strain in a sutured anastomosis was found to be 84% greater than in a pristine vessel, whereas a mechanically coupled anastomosis reduced arterial strain predictions by approximately 55%. Stress distributions in the su-tured anastomoses simulated here differed to those in reported literature. This result is attributed to the use of bonded connections in existing studies, to represent healed surgical sites. This has been confirmed by our study using FEA, and we believe this boundary condition significantly alters the stress distribution, and is less representative of the clinical picture following surgery. We have demonstrated that the inertial effects due to motion of the vessel during pulsatile flow are minimal, since the differences between the transient and static strain cal-\r\nculations range from around 0.6–7% dependent on the geometry. This implies that static structural analyses are likely sufficient to predict anastomotic strains in these simulations. Furthermore, approximations of the shear strain rate (SSR) were calculated and compared to analogous rigid-walled simulations, revealing that wall compliance had little influence on their overall magnitude. It is important to highlight, however, that SSR variations here are taken in isolation, and that changing pressure gradients are likely to produce much greater variation in vessel wall strain values than the influence of fluid flow alone. Hence, a formal fluid-structure interaction (FSI) study would be necessary to ascertain the true relationship.","PeriodicalId":54017,"journal":{"name":"Advanced Biomedical Engineering","volume":"1 1","pages":""},"PeriodicalIF":0.8000,"publicationDate":"2019-02-15","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.14326/ABE.8.63","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Advanced Biomedical Engineering","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.14326/ABE.8.63","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 1

Abstract

Simulation using computational methods is well-established for investigating mechanical and haemodynamic properties of blood vessels, however few groups have applied this technology to microvascular anastomoses. This study, for the first time, employs analytic and numeric models of sutured and coupled mi-croarterial anastomoses to evaluate the elastic and failure properties of these techniques in realistic geometries using measured arterial waveforms. Computational geometries were created of pristine microvessels and mi-croarterial anastomoses, performed using sutures and a coupling device. Vessel wall displacement, stress, and strain distributions were predicted for each anastomotic technique using finite element analysis (FEA) software in both static and transient simulations. This study focussed on mechanical properties of the anastomosis im- mediately after surgery, as failure is most likely in the early post-operative period. Comparisons were also drawn between stress distributions seen in analogous non-compliant simulations. The maximum principal strain in a sutured anastomosis was found to be 84% greater than in a pristine vessel, whereas a mechanically coupled anastomosis reduced arterial strain predictions by approximately 55%. Stress distributions in the su-tured anastomoses simulated here differed to those in reported literature. This result is attributed to the use of bonded connections in existing studies, to represent healed surgical sites. This has been confirmed by our study using FEA, and we believe this boundary condition significantly alters the stress distribution, and is less representative of the clinical picture following surgery. We have demonstrated that the inertial effects due to motion of the vessel during pulsatile flow are minimal, since the differences between the transient and static strain cal- culations range from around 0.6–7% dependent on the geometry. This implies that static structural analyses are likely sufficient to predict anastomotic strains in these simulations. Furthermore, approximations of the shear strain rate (SSR) were calculated and compared to analogous rigid-walled simulations, revealing that wall compliance had little influence on their overall magnitude. It is important to highlight, however, that SSR variations here are taken in isolation, and that changing pressure gradients are likely to produce much greater variation in vessel wall strain values than the influence of fluid flow alone. Hence, a formal fluid-structure interaction (FSI) study would be necessary to ascertain the true relationship.
缝合和耦合微动脉吻合的有限元预测
使用计算方法进行模拟是研究血管力学和血液动力学特性的公认方法,但很少有研究小组将这项技术应用于微血管吻合。这项研究首次采用缝合和耦合小动脉吻合的分析和数值模型,利用测量的动脉波形,评估这些技术在现实几何形状中的弹性和失效特性。使用缝合线和耦合装置创建原始微血管和小动脉吻合的计算几何结构。在静态和瞬态模拟中,使用有限元分析(FEA)软件预测每种吻合技术的血管壁位移、应力和应变分布。这项研究的重点是术后即刻吻合的力学性能,因为失败最有可能发生在术后早期。还对类似的非顺应性模拟中的应力分布进行了比较。缝合吻合中的最大主应变比原始血管中的大84%,而机械耦合吻合将动脉应变预测降低了约55%。此处模拟的网状吻合中的应力分布与文献报道中的不同。这一结果归因于现有研究中使用的粘结连接,以代表愈合的手术部位。我们使用有限元分析的研究已经证实了这一点,我们认为这种边界条件显著改变了应力分布,并且不太能代表手术后的临床情况。我们已经证明,在脉动流过程中,由于瞬态和静态应变计算之间的差异在0.6-7%左右,取决于几何形状,因此由血管运动引起的惯性效应是最小的。这意味着静态结构分析可能足以预测这些模拟中的吻合应变。此外,还计算了剪切应变速率(SSR)的近似值,并将其与类似的刚性壁模拟进行了比较,表明壁的柔度对其总体大小几乎没有影响。然而,重要的是要强调,这里的SSR变化是孤立的,并且变化的压力梯度可能比单独的流体流动的影响产生更大的容器壁应变值变化。因此,有必要进行正式的流体-结构相互作用(FSI)研究,以确定真正的关系。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Advanced Biomedical Engineering
Advanced Biomedical Engineering ENGINEERING, BIOMEDICAL-
CiteScore
1.40
自引率
10.00%
发文量
15
审稿时长
15 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信