Establishing local Diagnostic Reference Level for Adult Patients in Computed Tomography Examination in Kohgiluyeh and Boyer-Ahmad province

Q3 Health Professions
Majid Abiar, Mazyar Mahdavi, G. Haddadi
{"title":"Establishing local Diagnostic Reference Level for Adult Patients in Computed Tomography Examination in Kohgiluyeh and Boyer-Ahmad province","authors":"Majid Abiar, Mazyar Mahdavi, G. Haddadi","doi":"10.22038/IJMP.2020.46309.1724","DOIUrl":null,"url":null,"abstract":"Introduction: Nowadays, the absorbed dose of patients is on the rise due to the widespread use of computed tomography (CT) during the diagnosis process. Patients' doses for similar procedures are very different due to diversity in scanners and protocols. Hence, the purpose of this study was to determine the diagnostic reference levels for routine CT scan procedures in Kohgiluyeh and Boyer-Ahmad province, Iran.Material and methods: In this study, four common brain, sinus, chest and abdominopelvic procedures in spiral mode were selected. Next, the doses were measured in head and body phantom. Then, the third quartile of CTDIw was considered as the diagnostic reference dose. Finally, considering the pitch factor and the mean scan length in each protocol, the diagnostic reference level values based on the third quartile of the dose length product (DLP) and volume CTDI (CTDIvol) were determined.Results:The dose reference level values according to CTDIw third quartile in the brain, sinus, chest and abdominopelvic procedures were 39.82, 20.88, 14.10 and 17.07 mGy, respectively. In terms of dose length product, the diagnostic reference level values in the above procedures were determined to be 702.75, 243.90, 422.02, 865.62 mGy.cm, respectively.Conclusion:The DRLs of the CTDIW, DLP and CTDIvol of brain and sinus scans calculated in this study, were comparable to other provinces, national DRL and eight other countries. However, the same quantities for the chest and abdominopelvic scans showed higher values compared to the mentioned studies, suggesting lowering mAs and increasing pitch number for patient’s dose optimization. In some centers to preserve image quality, it is necessary to optimize radiation conditions, especially for chest and abdominopelvic scans.","PeriodicalId":14613,"journal":{"name":"Iranian Journal of Medical Physics","volume":" ","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2020-07-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Iranian Journal of Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.22038/IJMP.2020.46309.1724","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Health Professions","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: Nowadays, the absorbed dose of patients is on the rise due to the widespread use of computed tomography (CT) during the diagnosis process. Patients' doses for similar procedures are very different due to diversity in scanners and protocols. Hence, the purpose of this study was to determine the diagnostic reference levels for routine CT scan procedures in Kohgiluyeh and Boyer-Ahmad province, Iran.Material and methods: In this study, four common brain, sinus, chest and abdominopelvic procedures in spiral mode were selected. Next, the doses were measured in head and body phantom. Then, the third quartile of CTDIw was considered as the diagnostic reference dose. Finally, considering the pitch factor and the mean scan length in each protocol, the diagnostic reference level values based on the third quartile of the dose length product (DLP) and volume CTDI (CTDIvol) were determined.Results:The dose reference level values according to CTDIw third quartile in the brain, sinus, chest and abdominopelvic procedures were 39.82, 20.88, 14.10 and 17.07 mGy, respectively. In terms of dose length product, the diagnostic reference level values in the above procedures were determined to be 702.75, 243.90, 422.02, 865.62 mGy.cm, respectively.Conclusion:The DRLs of the CTDIW, DLP and CTDIvol of brain and sinus scans calculated in this study, were comparable to other provinces, national DRL and eight other countries. However, the same quantities for the chest and abdominopelvic scans showed higher values compared to the mentioned studies, suggesting lowering mAs and increasing pitch number for patient’s dose optimization. In some centers to preserve image quality, it is necessary to optimize radiation conditions, especially for chest and abdominopelvic scans.
在Kohgiluyeh和Boyer-Ahmad省建立成人ct检查的本地诊断参考水平
引言:如今,由于计算机断层扫描(CT)在诊断过程中的广泛使用,患者的吸收剂量正在上升。由于扫描仪和协议的多样性,类似手术的患者剂量差异很大。因此,本研究的目的是确定伊朗Kohkiluyeh省和Boyer Ahmad省常规CT扫描程序的诊断参考水平。材料和方法:在本研究中,选择了四种螺旋模式的常见脑、窦、胸和腹盆腔程序。接下来,在头部和身体模型中测量剂量。然后,CTDIw的第三个四分位数被认为是诊断参考剂量。最后,考虑到每个方案中的间距因子和平均扫描长度,确定了基于剂量-长度乘积(DLP)的第三个四分位数和体积CTDI(CTDIvol)的诊断参考水平值。结果:在脑、窦、胸和腹盆腔手术中,根据CTDIw第三个四分位数的剂量参考水平值分别为39.82、20.88、14.10和17.07mGy。就剂量-长度乘积而言,上述程序中的诊断参考水平值分别为702.75243.90422.02865.62mGy.cm。结论:本研究计算的脑和鼻窦扫描的CTDIW、DLP和CTDIvol的DRL与其他省份、国家和其他八个国家的DRL具有可比性。然而,与上述研究相比,胸部和腹骨盆扫描的相同数量显示出更高的值,这表明降低mAs和增加间距数可以优化患者的剂量。在一些中心,为了保持图像质量,有必要优化辐射条件,尤其是胸部和腹部骨盆扫描。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Iranian Journal of Medical Physics
Iranian Journal of Medical Physics Health Professions-Radiological and Ultrasound Technology
CiteScore
1.00
自引率
0.00%
发文量
0
审稿时长
8 weeks
期刊介绍: Iranian Journal of Medical Physics (IJMP) is the official scientific bimonthly publication of the Iranian Association of Medical Physicists. IJMP is an international and multidisciplinary journal, peer review, free of charge publication and open access. This journal devoted to publish Original Papers, Review Articles, Short Communications, Technical Notes, Editorial and Letters to the Editor in the field of “Medical Physics” involving both basic and clinical research. Submissions of manuscript from all countries are welcome and will be reviewed by at least two expert reviewers.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信